Recent from talks
Nothing was collected or created yet.
Activator technique
View on Wikipedia
The Activator Method Chiropractic Technique is a chiropractic treatment method that uses a device created by Arlan Fuhr as an alternative to manual manipulation of the spine or extremity joints. The device is categorized as a mechanical force manual assisted (MFMA) instrument which is generally regarded as a softer chiropractic treatment technique.[1] The method purports to use the device to identify and remove vertebral subluxations[2] and correct "pelvic deficiency", defined as an "'apparent' difference in leg length, not an anatomical difference".[3] These claims have been criticized.[2][3]
Activator Adjusting Instrument
[edit]
The traditional Activator Adjusting Instrument (AAI), or more simply, Activator, is a small handheld spring-loaded instrument which delivers a controlled and reproducible tap to the spine or other body part.[4] The aim is to produce enough force to move the vertebrae, but not enough to cause injury.[5] The tool's design is based on a dental impactor,[6] a device that taps dental amalgam into cavities in teeth.[4]
Activator I was patented by Activator Methods International on September 26, 1978.[4][7][8] Activator II was released, with an added "impedance head", in 1994.[4] Activator V is a cordless electric version that gives off no more than 0.3 J of kinetic energy in a 3-millisecond pulse.[4] It is the first FDA registered and approved cordless electronic chiropractic adjustment instrument.[9]
Leg length test reliability
[edit]The method involves using the device to identify and remove vertebral subluxations and correct "pelvic deficiency", defined as an "'apparent' difference in leg length, not an anatomical difference". The chiropractor checks the patient's "functional leg length" in the prone position, then uses the device on various parts of the body, and then rechecks the leg length to see if the Activator produced a change. This is repeated until the legs are apparently of an equal length and the treatment is considered to have resolved any vertebral subluxations.[3]
Although prone "functional leg length" is a widely used chiropractic tool, it is not a recognized anthropometric technique, since legs are often naturally of unequal length, and measurements in the prone position are not entirely valid estimates of standing X-ray differences.[10] Measurements in the standing position are far more reliable.[11] Another confounding factor is that simply moving the two legs held together and leaning them imperceptibly to one side or the other produces different results.[12][13] The Activator Methods technique uses leg length checks while prone (Position 1) and with the knees bent to 90 degrees (Position 2). Research shows good intraexaminer reliability and moderate interexaminer reliability with leg length checks in position 1, however no consensus has been met on the accuracy of leg length checks in position 1.[5]
Utilization and criticism
[edit]In 2003, the National Board of Chiropractic Examiners found that 69.9% of chiropractors used the technique, and 23.9% of patients received it.[14] The majority of US chiropractic schools and some schools in other countries teach the AMCT method, and an estimated 45,000 chiropractors worldwide use AMCT or some part of the technique.[5]
In 2001, the Chiropractors' Association of Saskatchewan considered the device "useless" and prohibited its use, a ban that was disputed in court.[15] The ban was lifted in 2003.[2] Its use has been criticized: "Activator Methods thus piles one dubious concept upon another. Its leg-length tests have not been demonstrated to be reliable or to yield significant data. Nor is there any reason to believe that 'pelvic deficiency' or its associated 'subluxations' are pathologic conditions."[3] "Activator Methods Chiropractic Technique is a nonsensical diagnostic and treatment system centered on the notion that leg-length analysis can locate subluxations and determine when to adjust the spine."[2]
References
[edit]- ^ Chiropractic: An Illustrated History. Mosby. 1995. ISBN 978-0-8016-7735-9.
- ^ a b c d "Chiropractic News Digest #03-04". Quackwatch. June 25, 2003. Retrieved June 19, 2024.
- ^ a b c d Homola, Samuel (March 25, 2006). "Questionable Claims Made for Activator Methods (AM)". Quackwatch. Retrieved June 19, 2024.
- ^ a b c d e Fuhr, Arlan W.; Colloca, Christopher J.; Green, John R.; Keller, Tony S. (1997). Activator Methods Chiropractic Technique (1st ed.). St. Louis, MO: Mosby. ISBN 0-8151-3684-6.
- ^ a b c Fuhr, Arlan W.; J. Michael Menke (February 2005). "Status of Activator Methods Chiropractic Technique, Theory, and Practice". Journal of Manipulative and Physiological Therapeutics. 28 (2): e1 – e20. doi:10.1016/j.jmpt.2005.01.001. PMID 15800504.
- ^ Vintage Dudley Research Corp Impactor Dental Instrument In Wooden Case & Parts
- ^ "United States Patent - Chiropractic Adjusting Instrument". United States Patent and Trademark Office. Retrieved 3 September 2017.
- ^ "Patent Images". United States Patent and Trademark Office. Retrieved 3 September 2017.
- ^ "510(k) Summary of Safety and Effectiveness" (PDF). Department of Health and Human Services. Archived from the original (PDF) on September 24, 2014. Retrieved 3 September 2017.
- ^ D W Rhodes, E R Mansfield, P A Bishop, J F Smith. The validity of the prone leg check as an estimate of standing leg length inequality measured by X-ray. Archived 2011-05-22 at the Wayback Machine J Manipulative Physiol Ther.; 18 (6):343-6
- ^ E Hanada, R L Kirby, M Mitchell, J M Swuste. Measuring leg-length discrepancy by the "iliac crest palpation and book correction" method: reliability and validity. Archived 2011-05-22 at the Wayback Machine Arch Phys Med Rehabil. 2001 Jul; 82 (7):938-42
- ^ Mitchell, Jacqueline S. (June 4, 2002). "Keeping Your Spine In Line". PBS. Archived from the original on June 28, 2002.
- ^ "Adjusting the Joints". PBS. Archived from the original on June 28, 2002.
- ^ Christensen MG, Kollasch MW (2005). "Professional functions and treatment procedures" (PDF). Job Analysis of Chiropractic. Greeley, CO: National Board of Chiropractic Examiners. pp. 121–38. ISBN 1-884457-05-3. Archived from the original (PDF) on 2008-09-10. Retrieved 2008-08-25.
- ^ "Chiropractors argument arrives in court". CBC. January 17, 2001. Retrieved June 19, 2024.
External links
[edit]Activator technique
View on GrokipediaHistory and Development
Origins and Key Inventors
The Activator technique was developed in 1967 by chiropractors Arlan W. Fuhr and Warren C. Lee in rural Minnesota as an instrument-assisted approach to spinal manipulation, aiming to provide a low-force alternative to traditional manual adjustments. Fuhr, who earned his Doctor of Chiropractic degree from Logan College in 1961, experienced physical fatigue from repetitive high-velocity, low-amplitude manual thrusts on numerous patients daily, prompting the search for a mechanical aid that could replicate precise, controlled impulses without excessive practitioner strain.[1][11][12] The initial prototype emerged serendipitously when Fuhr modified a dental surgical mallet—originally used for splitting impacted wisdom teeth—by substituting its scalpel head with a brake shoe rivet and capping it with a rubber tip from a doorstop, inspired by a dentist patient's tool. This rudimentary device delivered a quick percussive force, but its lack of durability limited clinical use until refinements produced a more reliable version. Fuhr and Lee formalized the technique through iterative testing, incorporating biomechanical principles to target vertebral subluxations with impulses mimicking manual adjustments but at reduced risk of practitioner injury or patient discomfort.[1] The first Activator Adjusting Instrument received a U.S. federal patent in 1976, establishing it as the predicate device for subsequent FDA approvals of chiropractic instruments. Fuhr, recognized as the co-inventor and primary architect, co-founded Activator Methods International to standardize and disseminate the technique, which drew on prior chiropractic concepts like those from early percussive devices while emphasizing empirical refinement through clinical observation. Lee contributed to the foundational protocol development, though Fuhr's leadership drove its evolution into a widely adopted method.[1][12][11]Evolution of the Technique and Instrument
The Activator technique emerged in the late 1960s as a response to limitations in traditional high-velocity, low-amplitude (HVLA) chiropractic adjustments, emphasizing controlled, low-force impulses to address vertebral subluxations with reduced risk of patient discomfort or injury. Dr. Arlan W. Fuhr, a chiropractor in rural Minnesota, collaborated with Dr. Warren C. Lee to pioneer the method around 1967, drawing on biomechanical observations of spinal motion and early diagnostic protocols like leg length inequality assessments to identify fixations.[13][14] This foundational approach prioritized precision over manual thrust, evolving through clinical refinement to incorporate supine and prone positioning for segmental analysis and correction.[1] The accompanying Activator Adjusting Instrument (AAI) originated from modifications to a dental mallet used for impacted tooth extraction, adapted in the early 1970s to deliver a rapid, mechanical impulse mimicking but gentler than manual adjustments. The first version received a U.S. federal patent in 1976, establishing it as a predicate device for subsequent FDA clearances, with Activator I formally patented on September 26, 1978, featuring a spring-loaded mechanism for consistent force application.[1] Activator Methods International, Ltd., founded by Fuhr, drove further iterations; Activator II, released in 1994, introduced an "impedance head" for real-time tissue resistance feedback, enhancing diagnostic accuracy during impulses.[15] Subsequent models advanced ergonomics and control: Activator III refined force-frequency profiles for broader clinical applicability, while Activator IV, developed under Fuhr's oversight, incorporated selectable predetermined force settings to tailor impulses to patient size and condition, marking a shift toward user-customizable precision.[16][17] The current Activator V represents the third generation overall, integrating cordless operation and optimized impulse delivery, with over 35 years of iterative research supporting its evolution from prototype to standardized tool in chiropractic practice.[18][9]Theoretical Foundations
Vertebral Subluxation Concept
The vertebral subluxation concept forms a core theoretical pillar of the Activator Technique, positing that misalignments or dysfunctions in the spinal vertebrae disrupt neural integrity and biomechanical function, thereby impairing the body's self-regulatory mechanisms. In chiropractic parlance, it is defined as a complex of functional, structural, or pathological articular changes that compromise spinal motion and nerve signaling, potentially leading to distant health disturbances beyond localized pain.[19][20] This notion traces to early 20th-century chiropractic founders like D.D. Palmer, who in 1910 described subluxation as nerve pressure causing abnormal organ function, a view retained in subluxation-centered practices including Activator Methods.[21] Proponents argue that such subluxations arise from trauma, posture, or repetitive stress, manifesting as altered segmental motion, muscle imbalance, or neuropathologic reflexes that cascade into systemic effects via viscerosomatic or somatovisceral pathways.[22] In the Activator approach, detection relies on protocols assessing leg length inequality and joint stiffness to infer subluxation presence, with the instrument delivering targeted thrusts to restore alignment and purportedly normalize neural flow without high-velocity manipulation.[23] Theoretical models emphasize kinesiologic, neurologic, and histologic components, suggesting subluxation induces peripheral nerve irritation or central sensitization.[24] Empirical validation remains scant, however; systematic reviews identify vertebral subluxation as a theoretical construct lacking robust experimental support for its detection, causality in non-musculoskeletal conditions, or correction via manipulation yielding broad health benefits.[25] Mainstream biomedical consensus deems the concept implausible, attributing any symptomatic relief from adjustments to placebo, endorphin release, or mechanical pain modulation rather than subluxation resolution, with no radiographic or physiologic markers consistently verifying its existence.[26][27] Chiropractic institutions often affirm its centrality despite these evidentiary gaps, reflecting a paradigm prioritizing clinical observation over randomized controlled trials.[28]Neurological and Biomechanical Rationale
The Activator technique posits that vertebral subluxations—functional spinal joint lesions characterized by restricted motion and aberrant sensory input—contribute to biomechanical dysfunction by altering segmental kinematics and load distribution. The adjusting instrument delivers a precise, high-velocity low-amplitude (HVLA) impulse, typically 116–140 N peak force over <0.1 ms followed by a sustained lower force phase (30–100 N for 1–5 ms), which induces small vertebral displacements (e.g., 1.62 mm axial and 0.48 mm shear at L2–L3 levels) without excessive preload or patient discomfort.[29] This mechanical input is theorized to restore joint play and intersegmental mobility, particularly in hypo-mobile segments, by overcoming articular fixations through resonant frequency excitation around 20 Hz. Biomechanical studies using cadaveric models demonstrate that repeated impulses (3–7 applications) at varying force settings (133–380 N) significantly enhance multiaxial lumbar motion by 3–26% (P < 0.005), with greater effects at lower forces on adjacent segments, supporting the technique's aim to normalize spinal kinematics and reduce compensatory overload.[30] However, these effects are primarily observed in controlled ex vivo settings, and in vivo translation to subluxation correction remains inferential, as direct causal links to long-term biomechanical homeostasis lack large-scale randomized validation beyond chiropractic-focused research.[30] Neurologically, the rationale centers on subluxations impairing somatosensory processing via mechanoreceptor hypo- or hyper-excitation, leading to altered reflex arcs and central sensitization. The instrument's impulse stimulates primary afferent fibers, including muscle spindles (Group Ia/II) and Golgi tendon organs, eliciting paraspinal muscle reflexes with EMG latencies of 2–3 ms and peaks at 50–100 ms, which transiently modulate motoneuron pool excitability—evidenced by H-reflex inhibition lasting up to 15 minutes post-thrust.[31] Proponents argue this resets aberrant proprioceptive feedback loops, reduces nociceptive barrage from trapped synovial folds or inflamed tissues, and diminishes peripheral drive to central pain pathways, potentially via descending inhibitory mechanisms. Supporting data from spinal manipulation literature indicate post-adjustment silencing of spindle afferents (average 1.3 s) and overall sensory inflow normalization, which could mitigate subluxation-induced neurological interference.[31] Critically, while these neurophysiological responses are documented in peer-reviewed studies, they derive largely from general spinal manipulation paradigms rather than Activator-specific trials, and causal attribution to subluxation resolution is contested outside chiropractic paradigms due to limited neuroimaging or longitudinal evidence linking impulses to sustained neural integrity improvements.[31] Empirical outcomes, such as pain reduction in musculoskeletal trials, indirectly align with these mechanisms but do not conclusively verify the subluxation model amid potential placebo or non-specific effects.[6]The Activator Adjusting Instrument
Design Features and Mechanism of Action
The Activator Adjusting Instrument (AAI) is a handheld, spring-loaded mechanical device engineered to deliver controlled, low-force impulses to spinal and extremity joints during chiropractic adjustments. It consists of a stylus tip—typically rubber or plastic, available in various sizes for targeting specific anatomical sites—attached to a preload mechanism that compresses against the patient's tissue prior to activation, ensuring reproducible force delivery independent of initial contact pressure variations. The instrument features ergonomic handles for practitioner comfort, weighs approximately 10 ounces in models like the EZ-Grip variant, and includes multiple predetermined force settings (commonly four levels) to accommodate different body regions, such as lighter impulses for cervical areas and stronger for lumbar segments.[16][32] Upon manual activation, the AAI releases stored spring energy to propel the tip forward, generating a high-velocity, low-amplitude thrust with a total duration of 2–5 milliseconds. Force profiles exhibit an initial spike of 116–140 N lasting less than 0.1 ms, followed by sustained lower forces of 30–100 N for 1–5 ms, with peak forces varying by setting: 115–123 N for cervical/thoracic applications and up to 211 N for lumbopelvic regions. The impulse spectrum peaks at approximately 20 Hz, minimizing variability (under 8% across trials) and enabling precise mechanical excitation suitable for dynamic assessment of tissue impedance.[29][32] This mechanism aims to induce localized joint motion by matching the impulse to the biomechanical properties of the target segment, potentially eliciting subtle displacements or acoustic emissions akin to cavitation without requiring patient relaxation or high manual forces. Validation studies using load cells and accelerometers on simulated spinal models confirm the instrument's consistency in force application and frequency output, supporting its role in reproducible, non-invasive manipulation protocols. Later electronic variants, such as the Activator V, incorporate solenoid-driven thrusts with similar force ranges but button-activated delivery for enhanced control.[29][6]Technical Specifications and Variations
The Activator Adjusting Instrument delivers a high-velocity, low-amplitude (HVLA) thrust characterized by an initial peak force ranging from 116 to 140 Newtons, followed by a short-duration pulse under 0.1 seconds, producing a complex dynamic impact profile.[29] This force-time characteristic enables precise mechanical impedance measurement and adjustment application without requiring patient twisting or excessive force from the practitioner.[33] Earlier mechanical models, such as the Activator I, output forces between approximately 47 to 123 Newtons (10.6 to 27.6 pounds) across adjustable settings, prioritizing high speed over high amplitude to minimize tissue resistance.[34] Variations across models reflect iterative enhancements in force-frequency spectra, ergonomics, and power delivery. The Activator I, introduced as a basic spring-loaded device made of stainless steel with a two-step mechanism, serves entry-level use with limited force customization.[35] The Activator II improves upon this by enhancing force profiles in the 10 to 100 Hz frequency range, incorporating an impedance head for better tissue interaction, and offering an EZ Grip variant for practitioners with smaller hands featuring reduced reach between palm and finger pads.[36][37] The Activator III further refines these force-frequency characteristics for more uniform thrust delivery compared to predecessors.[38] Subsequent models emphasize usability and precision. The Activator IV, constructed from lightweight, durable materials, includes predetermined force settings, preloaded tips for consistent depth control, and ergonomic handles eliminating the need for separate palm and finger pads, with an EZ Grip option for enhanced comfort.[16][39] The Activator V represents a shift to cordless electronic operation, FDA-registered for chiropractic use, with four adjustable thrust settings generating deeper force waves, a lithium-ion battery supporting all-day operation from one charge, and a non-slip ergonomic handle for reduced fatigue during one-handed delivery.[40][41] These electronic features provide more consistent and measurable impulses relative to mechanical spring-loaded variants, though peak forces remain calibrated for low-amplitude adjustments across all models.[42]Diagnostic Methods
Leg Length Inequality Assessment
In the Activator Methods Chiropractic Technique (AMCT), leg length inequality (LLI) assessment functions as the primary diagnostic procedure for identifying functional asymmetries linked to vertebral subluxations, pelvic distortions, or sacral unleveling. The patient is positioned prone on an adjusting table with hips extended and legs relaxed, enabling the practitioner to observe relative differences in heel or medial malleolar heights from a cephalad viewpoint.[43][32] An apparent short leg in this prone extended position (often termed Position 1) signals potential biomechanical dysfunction, with the contralateral leg typically appearing longer due to rotational or translational pelvic misalignment.[43] To localize the affected spinal segment, the assessment incorporates isolation tests—provocative maneuvers designed to challenge specific anatomical regions. These include contralateral heel compression to stress the sacroiliac joint, prone knee flexion (Position 2) to evaluate lumbar facets, cephalic perturbations like head turning for cervical influences, and caudal tests such as pelvic toggling. Observable changes in leg length following a test indicate the involved level; for instance, a shortening reversal after lumbar isolation suggests ipsilateral facet fixation.[32][44] Studies using optoelectric devices have measured these dynamic LLI shifts, reporting mean changes of 2-5 mm during isolation protocols in asymptomatic subjects, supporting the procedure's sensitivity to induced perturbations.[44] The method builds on foundational leg-checking approaches, such as the Derifield-Thompson prone analysis, integrated with directional non-force principles to emphasize reproducible, low-velocity assessments over direct measurement tools like tape or blocks.[32] Interexaminer reliability for the baseline prone extended check demonstrates moderate to good agreement, with one study of proficient practitioners achieving 85% concordance (kappa = 0.66) across 34 subjects categorized as left short, equal, or right short legs.[43] This reproducibility holds primarily for trained examiners, though findings often favor right short leg detections, potentially reflecting population asymmetries or procedural biases.[43] Post-adjustment rechecks verify symmetry restoration, with persistent LLI prompting segmental retargeting. While the assessment prioritizes functional over structural inequality—distinguishing apparent disparities from true bony discrepancies via response to challenges—its clinical utility relies on consistent application within the AMCT protocol.[32][44]Supine and Prone Protocols
The supine protocol for leg length inequality (LLI) assessment in Activator Methods Chiropractic Technique positions the patient on their back with legs extended and relaxed, head in midline. The examiner stands at the foot of the table, places index fingers on the distal medial malleoli, and gently elevates both legs until the heels clear the table surface, comparing malleolar heights for apparent discrepancies indicative of pelvic or lower extremity dysfunction.[45] This method aims to minimize pelvic torsion influences from prone positioning, though empirical comparisons show inconsistent agreement with prone evaluations, with supine often yielding smaller LLI magnitudes.[45] The prone protocol, foundational to Activator diagnostics, commences in Position 1 with the patient face down, legs extended, and head turned laterally. The practitioner stabilizes the pelvis if needed and compares heel or medial malleolar positions to detect a short leg, hypothesized to reflect sacral base posteriority or innominate rotation from subluxation. Transitioning to Position 2 involves knee flexion to 90 degrees while maintaining prone posture; observed leg length changes (e.g., short leg lengthening or contralateral shortening) guide subluxation localization to lumbar, pelvic, or sacral segments via predefined directional patterns.[18] Interexaminer reliability for prone extended Position 1 has been reported as moderate to substantial (kappa values 0.47–0.72) in standardized training contexts, though validity relative to imaging remains debated. These protocols integrate sequential challenges—such as heel flexion or cervical flexion in prone—to refine analysis, prioritizing functional over anatomic LLI for treatment targeting. Discrepancies between supine and prone methods, where prone detects larger inequalities in 68% of cases per one study of 50 asymptomatic subjects, underscore positioning's role in unloading gravitational vectors on sacroiliac mechanics.[45] Proponents attribute protocol specificity to biomechanical causality, yet causal inference requires radiographic correlation absent in routine application.[18]Treatment Procedures
Step-by-Step Adjustment Process
The Activator adjustment process begins after subluxation identification via leg length analysis and provocative tests, with the patient positioned prone on the adjustment table.[4] The practitioner selects the appropriate force setting on the Activator Adjusting Instrument, typically starting at level 2 or 3 for adults to deliver a controlled, high-velocity low-amplitude (HVLA) impulse calibrated between 0.3 and 1.0 Joules depending on the model and patient factors.[36] Contact is established precisely at the targeted anatomical site, such as the mammillary process for lumbar segments or the crest of the ilium for pelvic adjustments, with the instrument's tip perpendicular to the skin.[36] The line of drive (LOD) is oriented according to protocol-specific vectors—for instance, anterior-superior for L4 subluxations indicated by leg shortening in flexed knee position, or inferior-medial for anterosuperior ilium dysfunctions—to impart corrective force along the path of joint restriction.[36] Activation of the spring-loaded mechanism releases the percussive thrust in approximately 1/150th of a second, minimizing patient guarding and enabling adjustments on patients contraindicated for manual techniques.[4] Post-adjustment verification involves repeating the leg check protocol to confirm equalization of leg lengths, indicating biomechanical restoration; if imbalance persists, additional contacts may be applied sequentially up the spine or to extremities as per the basic scan hierarchy.[4] Adjustments are delivered in a cephalad-to-caudad direction for spinal protocols, prioritizing lower segments first to avoid compensatory changes.[36] The process emphasizes minimal force to enhance safety, with studies reporting peak forces under 400 Newtons, far below manual HVLA methods.[36]| Subluxation Example | Contact Point | Line of Drive (LOD) |
|---|---|---|
| L4 vertebra | Mammillary process | Anterior-superior |
| Anterosuperior ilium | Crest of ilium | Inferior-medial |
| Medial knee joint | Medial knee joint | Lateral-inferior |
