Thermal conductivity detector
Thermal conductivity detector
from Wikipedia

The thermal conductivity detector (TCD), also known as a katharometer, is a bulk property detector and a chemical specific detector commonly used in gas chromatography.[1] This detector senses changes in the thermal conductivity of the column eluent and compares it to a reference flow of carrier gas. Since most compounds have a thermal conductivity much less than that of the common carrier gases of helium or hydrogen, when an analyte elutes from the column the effluent thermal conductivity is reduced, and a detectable signal is produced.

Operation

[edit]

The TCD consists of an electrically heated filament in a temperature-controlled cell. Under normal conditions there is a stable heat flow from the filament to the detector body. When an analyte elutes and the thermal conductivity of the column effluent is reduced, the filament heats up and changes resistance. This resistance change is often sensed by a Wheatstone bridge circuit which produces a measurable voltage change. The column effluent flows over one of the resistors while the reference flow is over a second resistor in the four-resistor circuit.

TCD schematic

A schematic of a classic thermal conductivity detector design utilizing a Wheatstone bridge circuit is shown. The reference flow across resistor 4 of the circuit compensates for drift due to flow or temperature fluctuations. Changes in the thermal conductivity of the column effluent flow across resistor 3 will result in a temperature change of the resistor and therefore a resistance change which can be measured as a signal.

Since all compounds, organic and inorganic, have a thermal conductivity different from helium or hydrogen, virtually all compounds can be detected. That's why the TCD is often called a universal detector.

Used after a separation column (in a chromatograph), a TCD measures the concentrations of each compound contained in the sample. Indeed, the TCD signal changes when a compound passes through it, shaping a peak on a baseline. The peak position on the baseline reflects the compound type. The peak area (computed by integrating the TCD signal over time) is representative of the compound concentration. A sample whose compounds concentrations are known is used to calibrate the TCD: concentrations are affected to peak areas through a calibration curve.

The TCD is a good general purpose detector for initial investigations with an unknown sample compared to the FID that will react only to combustible compounds (Ex: hydrocarbons). Moreover, the TCD is a non-specific and non-destructive technique. The TCD is also used in the analysis of permanent gases (argon, oxygen, nitrogen, carbon dioxide) because it responds to all these substances unlike the FID which cannot detect compounds which do not contain carbon-hydrogen bonds.

Considering detection limit, both TCD and FID reach low concentration levels (inferior to ppm or ppb).[2]

Both of them require pressurized carrier gas (Typically: H2 for FID, He for TCD) but due to the risk associated with storing H2 (high flammability, see Hydrogen safety), TCD with He should be considered in locations where safety is crucial.

Considerations

[edit]

One thing to be aware of when operating a TCD is that gas flow must never be interrupted when the filament is hot, as doing so may cause the filament to burn out. While the filament of a TCD is generally chemically passivated to prevent it from reacting with oxygen, the passivation layer can be attacked by halogenated compounds, so these should be avoided wherever possible.[3]

If analyzing for hydrogen, the peak will appear as negative when helium is used as the reference gas. This problem can be avoided if another reference gas is used, for example argon or nitrogen, although this will significantly reduce the detector's sensitivity towards any compounds other than hydrogen.

Process description

[edit]

It functions by having two parallel tubes both containing gas and heating coils. The gases are examined by comparing the rate of loss of heat from the heating coils into the gas. The coils are arranged in a bridge circuit so that resistance changes due to unequal cooling can be measured. One channel normally holds a reference gas and the mixture to be tested is passed through the other channel.

Applications

[edit]

Katharometers are used medically in lung function testing equipment and in gas chromatography. The results are slower to obtain compared to a mass spectrometer, but the device is inexpensive, and has good accuracy when the gases in question are known, and it is only the proportion that must be determined.

Monitoring of hydrogen purity in hydrogen-cooled turbogenerators.

Detection of helium loss from the helium vessel of an MRI superconducting magnet.

Also used within the brewing industry to quantify the amount of carbon dioxide within beer samples.

Used within the energy industry to quantify the amount (calorific value) of methane within biogas samples.

Used within the food and drink industry to quantify and/or validate food packaging gases.

Used within the oil&gas industry to quantify the percentage of HCs when drilling into a formation.

References

[edit]
Revisions and contributorsEdit on WikipediaRead on Wikipedia
from Grokipedia
A thermal conductivity detector (TCD), also known as a hot-wire detector or katharometer, is a device used primarily in to identify and quantify analytes by measuring differences in the thermal conductivity of the carrier gas versus the carrier gas mixed with sample components. The device, originally known as the katharometer, was invented in 1915 by G. A. Shakespear for detecting in air mixtures. It was first applied to in the early 1950s, becoming a standard detector in the field's initial commercial instruments. The TCD operates on the principle that each gas has a unique conductivity, which affects the rate of heat dissipation from electrically heated filaments or thermistors within the detector. In a typical setup, the detector employs a circuit containing four sensing elements—two in a reference cell flushed with pure carrier gas (such as , chosen for its high conductivity) and two in a sample cell where the eluent from the chromatographic column enters. When analytes pass through the sample cell, they alter the gas mixture's conductivity, changing the and thus the electrical resistance of the filaments; this imbalance in the bridge circuit generates a voltage signal proportional to the concentration. Key performance characteristics include a linear dynamic range of approximately 10^4, operation at temperatures 20–25°C above the column temperature (typically up to 350–400°C for conditioning), and a universal response that detects a broad range of compounds, particularly permanent gases like hydrogen, oxygen, nitrogen, carbon monoxide, and carbon dioxide. Unlike selective detectors such as the flame ionization detector (FID), the TCD is non-destructive and does not require combustible gases, making it robust and suitable for environments where flames pose risks. However, its sensitivity is generally lower than that of ionization-based detectors, and it can be influenced by variations in flow rate, temperature, or contaminants that corrode the filaments. Applications of the TCD span gas analysis in industries including for purity testing, for air quality and composition (e.g., ), and process control for high-accuracy measurements. Modern advancements, such as integration into micro-gas chromatographs using micro-electro-mechanical systems (), have enhanced its portability for real-time field analysis while maintaining reliability.

Introduction

Definition and overview

A thermal conductivity detector (TCD), also known as a katharometer, is a bulk property detector that measures differences in thermal conductivity between a pure carrier gas and the mixture of carrier gas with sample eluent emerging from a (GC) column. This detection principle relies on the variation in heat dissipation caused by the presence of analytes, producing a signal proportional to their concentration without regard to . In , the TCD functions as a universal detector suitable for both organic and inorganic compounds, including permanent gases such as , , and that are challenging for other detectors. Its broad applicability stems from responsiveness to any species with thermal conductivity differing from the carrier gas, making it indispensable for analyses involving fixed gases or when compound-specific detection is unnecessary. The TCD operates in a non-destructive manner, preserving the integrity of eluted components for potential downstream processing or additional detection. Optimal performance requires carrier gases like helium or hydrogen, which exhibit high thermal conductivity to maximize sensitivity against most analytes. Detection limits typically fall in the parts per million (ppm) to parts per billion (ppb) range, enabling reliable quantification of trace-level components in diverse samples.

Historical development

The thermal conductivity detector, originally known as the katharometer, was invented in 1915 by H.A. Daynes at the request of the British Admiralty's Board of Invention and Research. Daynes developed the device to automatically detect small quantities of (1-2%) in air by exploiting differences in thermal conductivity between gas mixtures, using a circuit with heated platinum wire filaments in sealed and exposed cells. This innovation, detailed in his 1920 publication, marked the foundation for non-destructive gas analysis tools. Prior to its integration with chromatography, the katharometer found use in industrial settings for gas purity monitoring from the onward, such as assessing air quality in confined spaces or verifying gas compositions in manufacturing processes. Its ability to measure thermal conductivity variations enabled reliable detection of impurities without chemical alteration of samples, establishing it as a staple in early analysis. The device's application to emerged in the 1950s, with demonstrating its use as a universal detector in 1954 by monitoring resistance changes in a heated filament as column eluent altered the environment relative to pure carrier gas. This adaptation leveraged the katharometer's bulk property sensitivity for separating and quantifying diverse analytes in GC effluents. Commercial gas chromatographs incorporating became available in the mid-1950s, serving as the sole detection method in early instruments like those introduced in 1955-1956. Notable designs included W. Stuve's katharometer around 1957, which employed a filament for enhanced stability in chromatographic setups. Over the late 20th and early 21st centuries, thermal conductivity detectors evolved through , enabling integration into portable systems via microelectromechanical systems () technology. A key milestone was the 1979 development of the first silicon-based thermal conductivity detector for GC by S.C. Terry and colleagues at , which reduced size and power needs for field-deployable analyzers. Subsequent advances in the , such as microfabricated detectors in systems like microChemLab, further supported handheld applications for detection in . Recent advancements as of 2025 include novel silicon-based TCD designs that improve sensitivity and integration in portable systems.

Operating principle

Fundamentals of thermal conductivity

Thermal conductivity, denoted as κ\kappa, is a measure of a material's ability to conduct through the transfer of without bulk motion of the material itself. It is quantified as the amount of transmitted per unit time through a unit thickness and unit area under a unit , with standard units of watts per meter-kelvin (W/(m·K)). The governing equation for steady-state heat conduction is Fourier's law, which expresses the q\mathbf{q} (in W/m²) as proportional to the negative temperature T\nabla T: q=κT\mathbf{q} = -\kappa \nabla T This vector equation indicates that heat flows from regions of higher temperature to lower temperature, with κ\kappa determining the conduction efficiency; higher κ\kappa values imply better for a given . In gases, thermal conductivity primarily results from molecular diffusion of kinetic energy via collisions, as predicted by kinetic theory of gases. Key factors influencing κ\kappa include molecular weight, molecular structure (e.g., monatomic vs. polyatomic), and temperature; lighter, simpler molecules generally yield higher κ\kappa due to greater molecular velocities and mean free paths that facilitate energy transport. For instance, thermal conductivity increases with temperature because molecular speeds rise, enhancing collision-based energy exchange, while it remains largely independent of pressure at moderate levels where the mean free path is unaffected. Representative values at 300 K and atmospheric pressure illustrate these trends: helium (monatomic, low molecular weight) has κ0.156\kappa \approx 0.156 W/(m·K), hydrogen (diatomic, very low molecular weight) 0.187\approx 0.187 W/(m·K), nitrogen (diatomic, higher molecular weight) 0.026\approx 0.026 W/(m·K), and carbon dioxide (triatomic, heavier and more complex) 0.017\approx 0.017 W/(m·K). Such variations in κ\kappa between gases cause differential heat dissipation from a heated element immersed in them, enabling sensitive detection of compositional changes.

Detection mechanism

In a thermal conductivity detector (TCD), the core detection mechanism relies on a heated sensing element, such as a filament or , positioned within a gas stream. This element is electrically heated to maintain an elevated above that of the surrounding gas, and its heat loss occurs primarily through conduction to the gas, which is directly proportional to the gas's thermal conductivity (κ). When the sample gas mixture enters the stream, any difference in κ between the and the carrier gas alters the rate of heat dissipation from the sensing element. The underlying physics is governed by a heat balance equation, where the electrical power input to the sensing element balances the heat lost to the gas. This is approximated as P=I2R=hA(TfTg)P = I^2 R = h A (T_f - T_g), with PP representing the power, II the current, RR the resistance, hh the heat transfer coefficient (which depends on κ), AA the surface area of the element, TfT_f the filament temperature, and TgT_g the gas temperature. A change in κ modifies hh, thereby shifting TfT_f to restore equilibrium; for instance, a gas with higher κ enhances convective cooling, lowering TfT_f, while a lower-κ gas reduces cooling, raising TfT_f. This temperature variation causes a corresponding change in the resistance of the sensing element, as most materials exhibit a positive temperature coefficient of resistance. The resistance change (ΔR) is transduced into a measurable electrical signal using a circuit, typically configured with multiple sensing elements: some exposed to the reference carrier gas and others to the sample stream. The bridge imbalance produces a voltage output (ΔV) proportional to ΔR, and thus to the difference in κ (ΔV ∝ ΔR ∝ Δκ), which serves as the detector signal. For analytes with lower κ than the carrier (e.g., organic compounds in ), the sensing element in the sample cell heats up relative to the reference, yielding a positive peak; conversely, analytes with higher κ (e.g., hydrogen in carrier) cool the element more, producing a negative peak. The TCD response is linear with concentration for small differences in κ, typically spanning a of about 10^4 from parts per million to several percent.

Design and components

Main components

The thermal conductivity detector (TCD) consists of several key hardware elements that enable the of thermal conductivity differences in gas streams. Central to the is the sensing element, typically an electrically heated filament constructed from materials such as tungsten-rhenium or nickel-iron, which is maintained at typically ranging from 150°C to 350°C to facilitate sensitive to gas composition. These filaments function by altering their electrical resistance in response to changes in surrounding gas thermal conductivity, with the power supply adjusted to keep the temperature constant. To establish a stable baseline, the TCD incorporates a reference cell that mirrors the sensing element's configuration but exposes it exclusively to pure carrier gas, such as , allowing direct comparison with the sample stream and compensating for environmental fluctuations. This dual-channel setup ensures that any detected signal reflects only the thermal conductivity variation introduced by analytes in the sample gas. The electrical measurement relies on a Wheatstone bridge circuit, comprising four resistors: two active filaments (one for the sample path and one for the reference) and two fixed resistors, which detects imbalances in resistance as a differential voltage output indicative of thermal conductivity differences. This circuit configuration provides high sensitivity to small resistance changes, with the output voltage directly proportional to the power required to maintain filament temperature. Housing these elements is the detector cell, a compact, temperature-controlled chamber—often regulated via an integrated or Peltier device to temperatures above 150°C—that isolates the system from ambient influences and directs separate flow paths for sample effluent and reference gas over the respective filaments. The cell's design minimizes dead volume and ensures uniform gas distribution, typically supporting carrier flow rates aligned with chromatographic needs. Supporting electronics, including amplifiers and signal processors, condition the Wheatstone bridge's voltage output into a stable analog signal suitable for integration with systems, often providing ranges from -10 V to +10 V. In modern TCD designs, filaments and cell components frequently feature chemical passivation layers to safeguard against degradation by corrosive or oxidizing analytes, such as acids and .

Types of thermal conductivity detectors

Thermal conductivity detectors (TCDs) are categorized based on their sensing elements, which determine their design, operational characteristics, and suitability for various applications in . The primary types include hot-wire, , , and modern micro-machined variants, each employing different materials and configurations to measure differences in gas thermal conductivity. The hot-wire TCD, the most common type, utilizes a resistive filament typically made of tungsten or tungsten-rhenium wire suspended in a gas cell. This filament serves as both heater and sensor, with its resistance changing based on the surrounding gas's thermal conductivity, which affects cooling rates. It offers high sensitivity but is fragile due to the thin wire, prone to breakage or burnout, and typically dissipates power in the range of 0.5–1 W during operation. Hot-wire TCDs dominate in traditional gas chromatography systems for their established reliability in detecting a wide range of compounds. Thermistor-based TCDs employ beads, such as metal oxides, with a , placed in sample and reference gas streams. These provide greater robustness compared to hot-wire designs, as the solid beads resist mechanical damage, and operate at lower currents (8–15 mA), making them suitable for portable systems. Their temperature range is limited to about 150°C, with optimal performance below 100°C where sensitivity is enhanced for fixed-gas analysis. Thermopile TCDs use arrays of multiple thermocouples to measure heat flux differences without relying on resistive heating elements, offering improved high-temperature stability and reduced power consumption. Constructed from MEMS-fabricated thermocouples, they detect thermal gradients induced by gas conductivity variations, providing a non-resistive alternative that enhances longevity in demanding environments. These are particularly noted for low-power operation in sensing applications. Thick-film or micro-machined TCDs represent modern advancements using technology, featuring thin-film heaters and sensors on substrates for compact integration. Designs often include suspended thermistors on beams made from materials like and , minimizing dead volume and heat loss while achieving response times under 1 second. These emerged in the to enable for fast and portable devices, with improved sensitivity (down to 10 ppm) for detecting small-molecule gases. Recent advancements as of 2024 include ultralow-power suspended nanoheater designs and finite element simulations for optimized performance in gas sensing.

Operation

Measurement process

In the measurement process of a thermal conductivity detector (TCD) used in , the carrier gas, typically flowing at 20-50 mL/min, is split into two parallel channels: a reference channel containing pure carrier gas and a sample channel through which the column eluent passes. As analytes elute from the column, they mix with the carrier gas in the sample channel, altering the overall gas composition entering the detector. The filaments or thermistors in both channels are heated to a constant , typically 150-300°C, using an applied current of 50-300 mA for filaments, to establish . The circuit is then balanced with pure reference gas flowing through both channels, resulting in a zero baseline signal before sample introduction. This dual-channel configuration inherently cancels out common fluctuations in flow rate or ambient , ensuring stable operation. During analysis, when an elutes into the sample channel, it changes the thermal conductivity (κ) of the gas mixture compared to the reference channel, leading to differential loss from the heated elements. This imbalance in the bridge circuit produces an output voltage peak proportional to the analyte concentration, as the circuit supplies additional current to restore equilibrium in the sample-side element. The resulting electrical signal is amplified, filtered with a time constant typically between 50-220 ms for a response time of 100-500 ms, and digitized for . Peak height or area integration then quantifies the amount relative to standards. To conclude the measurement, the filament current is reduced to cool the elements gradually before stopping the gas flow, preventing or damage to the detector components.

Calibration and performance factors

Calibration of thermal conductivity detectors (TCDs) typically involves the use of standard gas mixtures containing known concentrations of the in the carrier gas to establish a . By injecting these standards into the gas chromatograph and measuring the resulting peak areas, a plot of peak area versus analyte concentration is generated, allowing determination of the for each analyte. This relationship is linear over a wide (approximately 10^4) for most compounds, enabling quantitative analysis based on the slope of the curve. The sensitivity of a TCD, defined as the minimum detectable quantity (MDQ), is approximately 10^{-7} to 10^{-8} g/s for typical organic analytes using as the carrier gas, though it can reach lower limits like 0.6 ppm v/v for under optimized conditions. Sensitivity strongly depends on the difference in thermal conductivity (Δκ) between the analyte and carrier gas; for example, detecting with as the carrier provides significantly higher sensitivity (up to ~28-fold) compared to due to the larger relative change in thermal conductivity (Δκ/κ_carrier). Representative examples include MDQs around 100 ppm v/v for in , highlighting the detector's suitability for analysis when Δκ is favorable. Key performance factors influencing TCD accuracy and reliability include flow rate stability, , and carrier gas purity. Flow rate variations exceeding 5-10% between reference and sample sides can cause baseline drift by altering heat dissipation, necessitating electronic flow controllers to maintain constancy within 10% for optimal operation. is critical, with the detector block typically maintained at high stability and 20-25°C above the maximum column temperature to prevent and ensure consistent filament or performance; fluctuations lead to signal instability. Carrier gas purity must be at least 99.999% to avoid impurities that mimic signals or degrade sensitivity, as even trace contaminants like oxygen or can alter thermal conductivity and cause erroneous peaks. Noise sources in TCDs primarily arise from in the sensing elements and electronic drift in the bridge circuit, which degrade the and limit detectability. These can be mitigated by operating at the lowest feasible filament current (e.g., 50-300 mA for filaments) and using time constants (e.g., 220 ms) to filter random , potentially reducing it by a factor of 2. Modulation techniques, such as excitation of the bridge, further improve signal-to-noise by suppressing DC drift from temperature or flow variations, enhancing overall reliability. Response factors vary significantly by , typically near 1 for compounds with thermal conductivities similar to the carrier but reaching up to 7-10 for extremes like in (relative conductivity ~7.17), where the large Δκ amplifies the signal.

Applications

In gas chromatography

In gas chromatography (GC) systems, the thermal conductivity detector (TCD) is positioned at the outlet of the separation column to analyze the eluent gas stream by comparing its thermal conductivity to that of the pure carrier gas. This placement allows for direct detection of separated components as they exit the column, making TCDs especially suitable for analyzing fixed and permanent gases such as O₂, N₂, CO, CO₂, and H₂. TCDs exhibit broad compatibility with various column types, including both packed columns and columns, particularly wide-bore variants (≥0.32 mm i.d.) that support flow rates of 3–5 mL/min. is the preferred carrier gas in these setups, as its high thermal conductivity enhances sensitivity for most analytes by amplifying differences in conductivity between the carrier and sample gases. For quantitative analysis, TCDs excel in measuring mixtures, supporting applications in —such as assessing air quality through detection of components like O₂ and CO—and in the petrochemical sector for evaluating gas compositions. An illustrative example is the analysis of light hydrocarbons in streams, where TCDs enable accurate quantification of like CH₄ (e.g., at 5740 ppmv) and C₂H₆ (e.g., at 1720 ppmv). TCDs rank as the third most commonly used GC detectors overall, with particular prevalence in laboratories specializing in inorganic and gas-phase analyses.

Other applications

Thermal conductivity detectors (TCDs), also known as katharometers, find extensive use in analysis for monitoring gas purity and composition in critical processes. In , TCDs measure hydrogen purity in synthesis gas mixtures to ensure optimal reaction conditions in the Haber-Bosch process, where impurities can reduce catalyst efficiency. Similarly, portable TCD-based helium sniffers detect leaks in superconducting magnets of MRI systems, where coolant losses must be minimized to maintain operational integrity and prevent costly refills. In medical applications, TCDs serve as katharometers in equipment, particularly for the closed-circuit helium dilution method to measure (FRC) by tracking concentrations during rebreathing until equilibrium. This inert tracer gas technique relies on the detector's ability to precisely quantify levels, aiding in the of volume abnormalities without invasive procedures. TCDs are integral to process control across various industries for real-time gas monitoring. In , thermal conductivity-based CO2 sensors assess carbonation levels in and carbonated beverages, ensuring consistent product quality by measuring dissolved CO2 during production and . For food , they analyze binary mixtures such as CO2 in air or to validate modified atmosphere , which extends by controlling oxygen and levels. In the oil and gas sector, TCDs detect leaks and quantify percentages in streams, supporting safety protocols and . Standalone TCD analyzers, often portable katharometers, enable on-site and assessments. These devices measure gases like , , and CO2 in binary mixtures for applications such as combustible gas detection in industrial settings or air quality checks in confined spaces, featuring weatherproof designs for field use. TCDs have been employed since the 1920s for gas analysis, with early methods applied to mixtures and mine gases like in air to evaluate composition and . In modern contexts, they provide real-time purity monitoring in systems, achieving sensitivities down to parts per million (ppm) for key impurities to safeguard performance and longevity.

Advantages and limitations

Advantages

Thermal conductivity detectors (TCDs) offer broad universality, capable of detecting a wide array of compounds including organics, inorganics, and permanent gases, as they respond to any solute with thermal conductivity differing from the carrier gas. This non-specific response enables analysis without destroying the sample, distinguishing TCDs from element-specific detectors and allowing post-detection sample collection for additional studies. TCDs are prized for their simplicity and reliability, requiring no reagents or chemical additives, which makes them robust for routine laboratory and industrial use. With proper maintenance, such as avoiding exposure to corrosive compounds, TCD filaments exhibit long lifespans, contributing to low operational costs and high stability over time. The detectors provide a linear response over a wide , typically spanning 10^5 to 10^6 (5 to 6 orders of magnitude), facilitating quantitative analysis without intricate for analytes with similar thermal conductivities. This ensures accurate measurements across varying concentrations, enhancing their utility in diverse applications. TCDs excel in detecting trace levels of permanent gases, such as and CO2, achieving sensitivities down to ppm levels where flame ionization detectors (FIDs) perform poorly due to low response to non-hydrocarbon . In , this capability supports precise monitoring of inert and reactive gases in complex mixtures.

Limitations

Thermal conductivity detectors (TCDs) exhibit relatively low sensitivity compared to other gas chromatography detectors, such as the (FID), particularly for organic compounds. The sensitivity of TCDs is typically 10-100 times lower than that of FIDs, with minimum detectable concentrations around 1-100 ppm (0.0001-0.01% v/v) for many analytes under standard conditions, though this can reach 0.1-1% v/v without optimization. TCDs are highly sensitive to fluctuations in flow rate and , leading to baseline drift that can compromise measurement accuracy. Changes in column or reference flow rates, as well as temperature-programmed runs that alter gas , contribute to this instability, even when mass-flow controllers are employed. Performance is strongly dependent on the carrier gas, with poorer sensitivity observed when using air or compared to or . For instance, hydrogen analytes produce negative peaks when is the carrier due to its higher thermal conductivity relative to , requiring specialized carrier blends like 10% hydrogen in to mitigate this issue. Additionally, carriers limit detection of gases like oxygen or because of their similar thermal conductivities. Interference from column bleed, gas leaks, or impurities in the carrier gas can significantly affect TCD signals, as even ppm-level contaminants like in produce noticeable responses. Passivation layers on TCD filaments, designed to protect against reactive , can be damaged by exposure to or acids, leading to permanent changes in detector response. The response time of TCDs is relatively slow due to their reliance on bulk thermal conductivity measurements and larger cell volumes, making them unsuitable for fast applications with peak widths under 1 second. This inherent limitation reduces specificity for complex mixtures. To prevent filament burnout, flow through the TCD must not be interrupted while the detector is hot, as this can cause overheating and permanent damage. While TCDs offer universality as a counterbalance to these drawbacks, their limitations often necessitate careful experimental design.

References