Marsh funnel
View on Wikipedia
The Marsh funnel is a simple device for measuring viscosity by observing the time it takes a known volume of liquid to flow from a cone through a short tube. It is standardized for use by mud engineers to check the quality of drilling mud. Other cones with different geometries and orifice arrangements are called flow cones, but have the same operating principle.
In use, the funnel is held vertically with the end of the tube closed by a finger. The liquid to be measured is poured through the mesh to remove any particles which might block the tube. When the fluid level reaches the mesh, the amount inside is equal to the rated volume. To take the measurement, the finger is released as a stopclock is started, and the liquid is allowed to run into a measuring container. The time in seconds is recorded as a measure of the viscosity.

The Marsh Funnel
[edit]Based on a method published in 1931 by H.N.Marsh,[1] a Marsh cone is a flow cone with an aspect ratio of 2:1 and a working volume of at least a litre. A Marsh funnel is a Marsh cone with a particular orifice and a working volume of 1.5 litres. It consists of a cone 6 inches (152 mm) across and 12 inches in height (305 mm) to the apex of which is fixed a tube 2 inches (50.8 mm) long and 3/16 inch (4.76 mm) internal diameter. A 10-mesh screen is fixed near the top across half the cone.[2]
In American practice (and most of the oil industry) the volume collected is a quart. If water is used, the time should be 26 +/- 0.5 seconds.[2] If the time is less than this the tube is probably enlarged by erosion, if more it may be blocked or damaged, and the funnel should be replaced. In some companies, and Europe in particular, the volume collected is a litre, for which the water funnel time should be 28 seconds. Marsh himself collected 0.50 litre, for which the time was 18.5 seconds.[1]
The Marsh funnel time is often referred to as the Marsh funnel viscosity, and represented by the abbreviation FV. The unit (seconds) is often omitted. Formally, the volume should also be stated. The (quart) Marsh funnel time for typical drilling muds is 34 to 50 seconds, though mud mixtures to cope with some geological conditions may have a time of 100 or more seconds.[3]
While the most common use is for drilling muds, which are non-Newtonian fluids, the Marsh funnel is not a rheometer, because it only provides one measurement under one flow condition. However the effective viscosity can be determined from following simple formula.[4]
μ = ρ (t - 25)
where μ = effective viscosity in centipoise
ρ = density in g/cm3
t = quart funnel time in seconds
For example, a mud of funnel time 40 seconds and density 1.1 g/cm3 has an effective viscosity of about 16.5 cP. For the range of times of typical muds above, the shear rate in the Marsh funnel is about 2000 s−1.[4]
Other Flow Cones
[edit]The term Marsh cone is also used in the concrete and oil industries. European standard EN445[5] and American standard C939[6] for measuring the flow properties of cement grout mixtures specify a funnel similar to the Marsh cone. Some manufacturers supply devices which they call Marsh cones, with removable tubes with size ranges from 5 to 15 mm. These can be used for quality control by selecting a tube which gives a convenient time, say 30 to 60 seconds.
References
[edit]- ^ a b H. N. Marsh (1931)Trans AIME vol 92 pp 234-251 "Properties and Treatment of Rotary Mud"
- ^ a b 13B-1 Recommended Standard Procedure for Field Testing Water-Based Drilling Fluids, Dallas: American Petroleum Institute, 1990
- ^ A. W. McCray & F. W. Cole (1979) Oil Well Drilling Technology University of Oklahoma Press ISBN 0-8061-0423-6
- ^ a b M. J. Pitt (2000) SPE Drilling and Completion vol 15 no1 pages 3-6 “The Marsh Funnel and Drilling Fluid Viscosity: A New Equation for Field Use”
- ^ BS EN445:1997 Grout for prestressing tendons - test methods, British Standards Institute, 1997
- ^ ASTM C939-10 Standard Test Method for Flow of Grout, ASTM International, 2010
Further reading
[edit]- ASTM D6910-04 Standard Test Method for Marsh Funnel Viscosity of Clay Construction Slurries
- N. Roussel & R. Le Roy (2005) Cement and Concrete Research vol 35 823-830 “The Marsh Cone: a test or a rheological apparatus?”
Marsh funnel
View on GrokipediaHistory
Invention
The Marsh funnel was invented by Hallan N. Marsh, an engineer affiliated with the General Petroleum Corporation of California and based in Los Angeles.[5] In 1931, Marsh detailed the design and application of his funnel viscometer in the paper "Properties and Treatment of Rotary Mud," published in the Transactions of the AIME.[5] The publication provided a comprehensive overview of rotary drilling mud properties, including humorous anecdotes about the funnel's development, such as Marsh's wry observation that he would be remembered more for the "d*** funnel" than for his other technical contributions.[5][6] Marsh created the device as a straightforward, portable instrument to assess drilling mud viscosity directly in the field, enabling quick evaluations without laboratory setups.[5] This addressed the need for reliable, on-site measurements of mud flow characteristics during rotary drilling operations.[5] The invention arose amid the California oil boom of the 1920s and early 1930s, a period of explosive growth in oil exploration and production in fields like Signal Hill and Huntington Beach, where efficient drilling fluid management became critical to operations.[7]Adoption and standardization
Following its invention in 1931, the Marsh funnel was rapidly adopted in oilfield operations during the 1930s as a practical tool for routine assessment of drilling fluid viscosity, enabling field engineers to monitor fluid properties on-site without complex laboratory equipment.[8] By the late 1930s, it had become integral to standard practices, with references to its use in conjunction with American Petroleum Institute (API) methods for controlling water loss in producing zones during drilling in fields like Turner Valley.[9] By the mid-20th century, the Marsh funnel was formally incorporated into API recommended practices for field testing of drilling fluids, specifically outlined in API RP 13B (later evolving into API RP 13B-1 for water-based muds), which standardized procedures for viscosity measurements to ensure consistency across the industry.[10] This standardization solidified its role as a benchmark tool in the oil and gas sector, with the device conforming to API specifications for dimensions and testing protocols that remain in use today.[11] The Marsh funnel's adoption extended globally, becoming a staple in drilling operations worldwide due to its simplicity and reliability, while adaptations appeared in international standards such as EN 445 (first published in 1996), which specifies a similar cone for measuring the fluidity of grout in prestressing tendons.[12] Over more than 90 years, the original design has endured without significant modifications, continuing to serve as the primary field viscometer in petroleum engineering and related applications.[4]Design and specifications
Components
The Marsh funnel consists of several key physical components designed for reliable viscosity measurement in drilling fluids. The primary element is the main body, a cone-shaped funnel constructed from impact-resistant plastic, measuring approximately 305 mm (12 inches) in height and 152 mm (6 inches) in diameter at the top, with a total capacity of 1500 ml up to the bottom of the screen.[11][13] At the base of the funnel is the orifice, a tubular outlet with an inside diameter of 4.7 mm (3/16 inch) and a length of 50.8 mm (2 inches), which controls the flow rate of the fluid during testing.[11][13] Over the top opening sits a fixed metal screen with 1.6 mm (1/16 inch) openings equivalent to 12 mesh, covering half the area and positioned 19 mm below the funnel's top edge to prevent large particles from entering and clogging the device.[11][13] The outflow is collected in a graduated measuring cup made of impact-resistant plastic, with a minimum capacity of 946 ml (1 quart) and markings in milliliters and US ounces for precise volume measurement.[11][13] An optional stand or support frame may be used to position the funnel stably during operation, ensuring consistent test conditions.[11]Dimensions and materials
The Marsh funnel features a conical design with a top diameter of 152 mm and a total height of 305 mm, providing a total volume capacity of 1.5 liters up to the screen before outflow measurement.[14] These dimensions ensure consistent flow dynamics for viscosity testing of drilling fluids, as specified in industry standards for field portability and accuracy.[15] The discharge tube attached to the funnel's apex has an internal diameter of 4.7 mm (3/16 inch) and a length of 50.8 mm (2 inches), optimized to facilitate laminar flow primarily for low-viscosity fluids like water-based muds.[14] This configuration allows the device to measure the time for 946 ml of fluid to efflux, serving as a proxy for relative viscosity.[16] The body of the Marsh funnel is typically constructed from molded, impact-resistant plastic to provide durability, corrosion resistance, and resistance to temperature variations encountered in drilling operations.[15] The tube and the 12-mesh screen fitted at the top are usually made of brass or stainless steel, enabling them to withstand exposure to aggressive drilling fluid chemicals without degradation. While plastic-bodied funnels represent the standard for routine field use due to their lightweight nature and cost-effectiveness, metal variants—often constructed from aluminum or full stainless steel—are available for harsh environments requiring enhanced mechanical strength and chemical resistance.[17]Principle of operation
Viscosity measurement basics
The Marsh funnel quantifies apparent viscosity by measuring the time required for exactly 946 ml (one quart) of fluid to efflux through its standardized orifice under gravity alone, without applied pressure. This efflux time, denoted as in seconds, serves as the primary output, reflecting the fluid's resistance to flow in a simple, empirical manner suitable for field conditions. The device is calibrated according to American Petroleum Institute (API) Recommended Practice 13B-1 such that fresh water at a temperature of 21 ± 3°C (70 ± 5°F) discharges in 26 ± 0.5 seconds, establishing a baseline for low-viscosity Newtonian fluids. Although designed primarily for Newtonian fluids—where viscosity remains constant independent of shear rate—the Marsh funnel is routinely employed for non-Newtonian drilling muds, which display variable viscosity under shear, such as shear-thinning behavior common in bentonite-based suspensions. For these complex fluids, the measurement yields an apparent viscosity that indicates overall consistency and stability rather than true rheological parameters like plastic viscosity or yield point. This application relies on the funnel's ability to simulate coarse flow conditions akin to those in drilling operations, providing a practical, though approximate, indicator of fluid performance.[18] The resulting viscosity is reported in units of "Marsh funnel seconds" or "funnel viscosity," which differ from absolute measures like centipoise (cP) and emphasize relative comparisons across samples rather than precise thermodynamic values. Conceptually, apparent viscosity scales directly with efflux time, such that higher corresponds to greater flow resistance: . To estimate effective viscosity in centipoise for field use, particularly with non-Newtonian muds, the relation incorporates fluid density: where is effective viscosity (cP), is density (g/cm³), and 25 seconds approximates the adjusted baseline for water's flow. This equation, derived from numerical simulations of power-law fluids validated against experiments, enables quick conversions while highlighting density's role in gravitational flow.[19]Flow characteristics
The Marsh funnel operates on a gravity-driven flow principle, where the drilling fluid is poured into the conical reservoir and flows downward through a narrow exit tube solely under the influence of gravitational force. As the fluid traverses the narrowing cone and constricts at the 4.7 mm diameter tube, viscous forces create resistance that slows the flow rate, with higher viscosity leading to greater opposition and longer efflux times.[20][21] The device assumes predominantly laminar flow conditions during measurement, characterized by low Reynolds numbers typically below 2000, where viscous effects dominate over inertial forces to ensure smooth, predictable fluid motion without turbulence. This regime is maintained in calibration with water and for most drilling fluids exhibiting satisfactory flow properties, as transitional or turbulent behavior near the orifice can introduce inaccuracies if Reynolds numbers exceed this threshold.[22] A 12-mesh screen (approximately 1.6 mm openings) positioned across half the funnel's top inlet filters out larger solid particles from the drilling fluid, preventing immediate clogging and allowing only finer suspensions to influence the flow. However, elevated concentrations of smaller solids can still accumulate, unpredictably increasing flow resistance or causing partial blockages that deviate from ideal viscous behavior.[23][24] Fluid temperature significantly impacts flow rate, as higher temperatures reduce viscosity and accelerate efflux, while lower temperatures do the opposite; API standards specify calibration and testing at 21 ± 3 °C to standardize measurements and minimize thermal variations. The efflux time measured serves as a practical proxy for relative viscosity, correlating inversely with flow ease under these controlled conditions.[2][11]Usage procedure
Step-by-step testing
To conduct the Marsh funnel test, begin by ensuring the device is clean and properly assembled, with the receiving cup positioned to collect exactly 946 ml (1 quart) of fluid. The test measures the time required for a specified volume of drilling fluid to flow through the funnel under gravity, providing a simple field assessment of its consistency.[2] The procedure follows these steps:- Preparation: Cover the funnel's discharge orifice with a finger or plug to prevent flow. Pour freshly sampled and well-mixed drilling fluid through the built-in screen into the upright funnel until it reaches the bottom of the screen (filling to the brim, approximately 1.5 liters). This ensures large particles are filtered out and the sample is representative.[11][2]
- Timing: Position the funnel over the empty graduated receiving cup. Remove the finger or plug to start the flow, simultaneously activating a stopwatch. Continue timing until the fluid level in the cup reaches the 946 ml mark, recording the elapsed time to the nearest second. Measure the fluid temperature immediately after, as it influences flow properties.[11][2]
- Repetition: While a single test is standard, repeat with fresh samples from the same batch if inconsistencies such as air bubbles in the fluid column or clogs in the orifice are observed, to ensure reliability. Discard any test showing anomalies and average consistent results as needed.[11]