Recent from talks
Nothing was collected or created yet.
Chemical shift
View on Wikipedia
In nuclear magnetic resonance (NMR) spectroscopy, the chemical shift is the resonant frequency of an atomic nucleus relative to a standard in a magnetic field. Often the position and number of chemical shifts are diagnostic of the structure of a molecule.[1][2][3] Chemical shifts are also used to describe signals in other forms of spectroscopy such as photoemission spectroscopy.
Some atomic nuclei possess a magnetic moment (nuclear spin), which gives rise to different energy levels and resonance frequencies in a magnetic field. The total magnetic field experienced by a nucleus includes local magnetic fields induced by currents of electrons in the molecular orbitals (electrons have a magnetic moment themselves). The electron distribution of the same type of nucleus (e.g. 1H, 13C, 15N) usually varies according to the local geometry (binding partners, bond lengths, angles between bonds, and so on), and with it the local magnetic field at each nucleus. This is reflected in the spin energy levels (and resonance frequencies). The variations of nuclear magnetic resonance frequencies of the same kind of nucleus, due to variations in the electron distribution, is called the chemical shift. The size of the chemical shift is given with respect to a reference frequency or reference sample (see also chemical shift referencing), usually a molecule with a barely distorted electron distribution.
Operating frequency
[edit]The operating (or Larmor) frequency of a magnet (usually quoted as absolute value in MHz) is calculated from the Larmor equation[4]
where B0 is the induction of the magnet (SI units of tesla), and is the magnetogyric ratio of the nucleus — an empirically measured fundamental constant determined by the details of the structure of each nucleus. For example, the proton operating frequency for a 1-tesla magnet is calculated as
MRI scanners are often referred to by their field strengths B0 (e.g. "a 7 T scanner"), whereas NMR spectrometers are commonly referred to by the corresponding proton Larmor frequency (e.g. "a 300 MHz spectrometer", which has a B0 of 7 T). While chemical shift is referenced in order that the units are equivalent across different field strengths, the actual frequency separation in hertz scales with field strength (B0). As a result, the difference of chemical shift between two signals (ppm) represents a larger number of hertz on machines that have larger B0, and therefore the signals are less likely to be overlapping in the resulting spectrum. This increased resolution is a significant advantage for analysis. (Larger-field machines are also favoured on account of having intrinsically higher signal arising from the Boltzmann distribution of magnetic spin states.)
Chemical shift referencing
[edit]Chemical shift δ is usually expressed in parts per million (ppm) by frequency, because it is calculated from[5]
where νsample is the absolute resonance frequency of the sample, and νref is the absolute resonance frequency of a standard reference compound, measured in the same applied magnetic field B0. Since the numerator is usually expressed in hertz, and the denominator in megahertz, δ is expressed in ppm.
The detected frequencies (in Hz) for 1H, 13C, and 29Si nuclei are usually referenced against TMS (tetramethylsilane), TSP (trimethylsilylpropanoic acid), or DSS, which by the definition above have a chemical shift of zero if chosen as the reference. Other standard materials are used for setting the chemical shift for other nuclei.
Thus an NMR signal observed at a frequency 300 Hz higher than the signal from TMS, where the TMS resonance frequency is 300 MHz, has a chemical shift of
Although the absolute resonance frequency depends on the applied magnetic field, the chemical shift is independent of external magnetic field strength. On the other hand, the resolution of NMR will increase with applied magnetic field.
Referencing methods
[edit]Practically speaking, diverse methods may be used to reference chemical shifts in an NMR experiment, which can be subdivided into indirect and direct referencing methods.[5] Indirect referencing uses a channel other than the one of interest to adjust chemical shift scale correctly, i.e. the solvent signal in the deuterium (lock) channel can be used to reference the a 1H NMR spectrum.[5] Both indirect and direct referencing can be done as three different procedures:
- Internal referencing, where the reference compound is added directly to the system under study."[5] In this common practice, users adjust residual solvent signals of 1H or 13C NMR spectra with calibrated spectral tables.[6][7] If substances other than the solvent itself are used for internal referencing, the sample has to be combined with the reference compound, which may affect the chemical shifts.
- External referencing, involving sample and reference contained separately in coaxial cylindrical tubes.[5] With this procedure, the reference signal is still visible in the spectrum of interest, although the reference and the sample are physically separated by a glass wall. Magnetic susceptibility differences between the sample and the reference phase need to be corrected theoretically,[5] which lowers the practicality of this procedure.
- Substitution method: The use of separate cylindrical tubes for the sample and the reference compound, with (in principle) spectra recorded individually for each.[5] Similar to external referencing, this method allows referencing without sample contamination. If field/frequency locking via the 2H signal of the deuterated solvent is used and the solvents of reference and analyte are the same, the use of this methods is straightforward. Problems may arise if different solvents are used for the reference compound and the sample as (just like for external referencing) magnetic susceptibility differences need to be corrected theoretically.[5][8] If this method is used without field/frequency locking, shimming procedures between the sample and the reference need to be avoided as they change the applied magnetic field (and thereby influence the chemical shift).[5]
Modern NMR spectrometers commonly make use of the absolute scale,[8][5] which defines the 1H signal of TMS as 0 ppm in proton NMR and the center frequencies of all other nuclei as percentage of the TMS resonance frequency:[5][8]
The use of the deuterium (lock) channel, so the 2H signal of the deuterated solvent, and the Ξ value of the absolute scale is a form of internal referencing and is particularly useful in heteronuclear NMR spectroscopy as local reference compounds may not be always be available or easily used (i.e. liquid NH3 for 15N NMR spectroscopy). This system, however, relies on accurately determined 2H NMR chemical shifts enlisted in the spectrometer software and correctly determined Ξ values by IUPAC.[5][8] A recent study for 19F NMR spectroscopy revealed that the use of the absolute scale and lock-based internal referencing led to errors in chemical shifts.[9][10] These may be negated by inclusion of calibrated reference compounds.[9][10]
The induced magnetic field
[edit]The electrons around a nucleus will circulate in a magnetic field and create a secondary induced magnetic field. This field opposes the applied field as stipulated by Lenz's law and atoms with higher induced fields (i.e., higher electron density) are therefore called shielded, relative to those with lower electron density. Electron-donating alkyl groups, for example, lead to increased shielding whereas electron-withdrawing substituents such as nitro groups lead to deshielding of the nucleus. Not only substituents cause local induced fields. Bonding electrons can also lead to shielding and deshielding effects. A striking example of this is the pi bonds in benzene. Circular current through the hyperconjugated system causes a shielding effect at the molecule's center and a deshielding effect at its edges. Trends in chemical shift are explained based on the degree of shielding or deshielding.
Nuclei are found to resonate in a wide range to the left (or more rare to the right) of the internal standard. When a signal is found with a higher chemical shift:
- The applied effective magnetic field is lower, if the resonance frequency is fixed (as in old traditional CW spectrometers)
- The frequency is higher, when the applied magnetic field is static (normal case in FT spectrometers)
- The nucleus is more deshielded
- The signal or shift is downfield or at low field or paramagnetic.
Conversely a lower chemical shift is called a diamagnetic shift, and is upfield and more shielded.
Diamagnetic shielding
[edit]In real molecules protons are surrounded by a cloud of charge due to adjacent bonds and atoms. In an applied magnetic field (B0) electrons circulate and produce an induced field (Bi) which opposes the applied field. The effective field at the nucleus will be B = B0 − Bi. The nucleus is said to be experiencing a diamagnetic shielding.
Factors causing chemical shifts
[edit]Important factors influencing chemical shift are electron density, electronegativity of neighboring groups and anisotropic induced magnetic field effects.
Electron density shields a nucleus from the external field. For example, in proton NMR the electron-poor tropylium ion has its protons downfield at 9.17 ppm, those of the electron-rich cyclooctatetraenyl anion move upfield to 6.75 ppm and its dianion even more upfield to 5.56 ppm.
A nucleus in the vicinity of an electronegative atom experiences reduced electron density and the nucleus is therefore deshielded. In proton NMR of methyl halides (CH3X) the chemical shift of the methyl protons increase in the order I < Br < Cl < F from 2.16 ppm to 4.26 ppm reflecting this trend. In carbon NMR the chemical shift of the carbon nuclei increase in the same order from around −10 ppm to 70 ppm. Also when the electronegative atom is removed further away the effect diminishes until it can be observed no longer.
Anisotropic induced magnetic field effects are the result of a local induced magnetic field experienced by a nucleus resulting from circulating electrons that can either be paramagnetic when it is parallel to the applied field or diamagnetic when it is opposed to it. It is observed in alkenes where the double bond is oriented perpendicular to the external field with pi electrons likewise circulating at right angles. The induced magnetic field lines are parallel to the external field at the location of the alkene protons which therefore shift downfield to a 4.5 ppm to 7.5 ppm range. The three-dimensional space where a diamagnetic shift is called the shielding zone with a cone-like shape aligned with the external field.
The protons in aromatic compounds are shifted downfield even further with a signal for benzene at 7.73 ppm as a consequence of a diamagnetic ring current.
Alkyne protons by contrast resonate at high field in a 2–3 ppm range. For alkynes the most effective orientation is the external field in parallel with electrons circulation around the triple bond. In this way the acetylenic protons are located in the cone-shaped shielding zone hence the upfield shift.
Magnetic properties of most common nuclei
[edit]1H and 13C are not the only nuclei susceptible to NMR experiments. A number of different nuclei can also be detected, although the use of such techniques is generally rare due to small relative sensitivities in NMR experiments (compared to 1H) of the nuclei in question, the other factor for rare use being their slender representation in nature and organic compounds.
| Isotope | Occurrence in nature (%) |
Spin number I | Magnetic moment μ (μN) |
Electric quadrupole moment (e × 10−24 cm2) |
Operating frequency at 7 T (MHz) |
Relative sensitivity |
|---|---|---|---|---|---|---|
| 1H | 99.984 | 1/2 | 2.79628 | 0 | 300.13 | 1 |
| 2H | 0.016 | 1 | 0.85739 | 0.0028 | 46.07 | 0.0964 |
| 10B | 18.8 | 3 | 1.8005 | 0.074 | 32.25 | 0.0199 |
| 11B | 81.2 | 3/2 | 2.6880 | 0.026 | 96.29 | 0.165 |
| 12C | 98.9 | 0 | 0 | 0 | 0 | 0 |
| 13C | 1.1 | 1/2 | 0.70220 | 0 | 75.47 | 0.0159 |
| 14N | 99.64 | 1 | 0.40358 | 0.071 | 21.68 | 0.00101 |
| 15N | 0.37 | 1/2 | −0.28304 | 0 | 30.41 | 0.00104 |
| 16O | 99.76 | 0 | 0 | 0 | 0 | 0 |
| 17O | 0.0317 | 5/2 | −1.8930 | −0.0040 | 40.69 | 0.0291 |
| 19F | 100 | 1/2 | 2.6273 | 0 | 282.40 | 0.834 |
| 28Si | 92.28 | 0 | 0 | 0 | 0 | 0 |
| 29Si | 4.70 | 1/2 | −0.5548 | 0 | 59.63 | 0.0785 |
| 31P | 100 | 1/2 | 1.1205 | 0 | 121.49 | 0.0664 |
| 35Cl | 75.4 | 3/2 | 0.92091 | −0.079 | 29.41 | 0.0047 |
| 37Cl | 24.6 | 3/2 | 0.68330 | −0.062 | 24.48 | 0.0027 |
1H, 13C, 15N, 19F and 31P are the five nuclei that have the greatest importance in NMR experiments:
- 1H because of high sensitivity and vast occurrence in organic compounds
- 13C because of being the key component of all organic compounds despite occurring at a low abundance (1.1%) compared to the major isotope of carbon 12C, which has a spin of 0 and therefore is NMR-inactive.
- 15N because of being a key component of important biomolecules such as proteins and DNA
- 19F because of high relative sensitivity
- 31P because of frequent occurrence in organic compounds and moderate relative sensitivity
Chemical shift manipulation
[edit]In general, the associated increased signal-to-noise and resolution has driven a move towards increasingly high field strengths. In limited cases, however, lower fields are preferred; examples are for systems in chemical exchange, where the speed of the exchange relative to the NMR experiment can cause additional and confounding linewidth broadening. Similarly, while avoidance of second order coupling is generally preferred, this information can be useful for elucidation of chemical structures. Using refocussing pulses placed between recording of successive points of the free induction decay, in an analogous fashion to the spin echo technique in MRI, the chemical shift evolution can be scaled to provide apparent low-field spectra on a high-field spectrometer.[12] In a similar fashion, it is possible to upscale the effect of J-coupling relative to the chemical shift using pulse sequences that include additional J-coupling evolution periods interspersed with conventional spin evolutions.[13]
Other chemical shifts
[edit]The Knight shift (first reported in 1949) and Shoolery's rule are observed with pure metals and methylene groups, respectively. The NMR chemical shift in its present-day meaning first appeared in journals in 1950. Chemical shifts with a different meaning appear in X-ray photoelectron spectroscopy as the shift in atomic core-level energy due to a specific chemical environment. The term is also used in Mössbauer spectroscopy, where similarly to NMR it refers to a shift in peak position due to the local chemical bonding environment. As is the case for NMR the chemical shift reflects the electron density at the atomic nucleus.[14]
See also
[edit]- EuFOD, a shift agent
- MRI
- Nuclear magnetic resonance
- Nuclear magnetic resonance spectroscopy of carbohydrates
- Nuclear magnetic resonance spectroscopy of nucleic acids
- Nuclear magnetic resonance spectroscopy of proteins
- Random coil index
- Relaxation (NMR)
- Solid-state NMR
- TRISPHAT, a chiral shift reagent for cations
- Zeeman effect
References
[edit]- ^ Silverstein; Bassler; Morrill (1981). Spectrometric Identification of organic Compounds (4th ed.). ISBN 978-0-471-09070-0.
- ^ Kemp, William (1987). Organic Spectroscopy (3rd ed.). ISBN 978-0-333-41767-6.
- ^ Balei, Metin. Basic 1H and 13C-NMR spectroscopy. ISBN 978-0-444-51811-8.
- ^ "Chemical Shift". NMRCentral. Archived from the original on 26 September 2011.
- ^ a b c d e f g h i j k l Harris, R. K.; Becker, E. D.; Cabral de Menezes, S. M.; Goodfellow, R.; Granger, P. (2001). "NMR nomenclature. Nuclear spin properties and conventions for chemical shifts (IUPAC Recommendations 2001)". Pure Appl. Chem. 73 (11): 1795–1818. doi:10.1351/pac200173111795.
- ^ Gottlieb, Hugo E.; Kotlyar, Vadim; Nudelman, Abraham (1997). "NMR Chemical Shifts of Common Laboratory Solvents as Trace Impurities". The Journal of Organic Chemistry. 62 (21): 7512–7515. doi:10.1021/jo971176v. ISSN 0022-3263. PMID 11671879.
- ^ Fulmer, Gregory R.; Miller, Alexander J. M.; Sherden, Nathaniel H.; Gottlieb, Hugo E.; Nudelman, Abraham; Stoltz, Brian M.; Bercaw, John E.; Goldberg, Karen I. (10 May 2010). "NMR Chemical Shifts of Trace Impurities: Common Laboratory Solvents, Organics, and Gases in Deuterated Solvents Relevant to the Organometallic Chemist" (PDF). Organometallics. 29 (9): 2176–2179. doi:10.1021/om100106e. ISSN 0276-7333.
- ^ a b c d Harris, Robin K.; Becker, Edwin D.; Cabral de Menezes, Sonia M.; Granger, Pierre; Hoffman, Roy E.; Zilm, Kurt W. (2008). "Further conventions for NMR shielding and chemical shifts (IUPAC Recommendations 2008)". Pure and Applied Chemistry. 80 (1): 59–84. doi:10.1351/pac200880010059. ISSN 0033-4545.
- ^ a b Rosenau, Carl Philipp; Jelier, Benson J.; Gossert, Alvar D.; Togni, Antonio (16 May 2018). "Exposing the Origins of Irreproducibility in Fluorine NMR Spectroscopy". Angewandte Chemie International Edition. 51 (30): 9528–9533. doi:10.1002/anie.201802620. ISSN 1433-7851. PMID 29663671.
- ^ a b Rosenau, Carl Philipp; Jelier, Benson J.; Gossert, Alvar D.; Togni, Antonio (16 May 2018). "Fluor-NMR-Spektroskopie rekalibriert". Angewandte Chemie (in German). 130 (30): 9672–9677. Bibcode:2018AngCh.130.9672R. doi:10.1002/ange.201802620. ISSN 0044-8249.
- ^ CRC Handbook of Chemistry and Physics (65th ed.).
- ^ Morris, Gareth A.; Jerome, Neil P.; Lian, Lu-Yun (17 February 2003). "Real-Time Chemical-Shift Scaling in High-Resolution NMR Spectroscopy". Angewandte Chemie (in German). 115 (7): 847–849. doi:10.1002/ange.200390189.
- ^ Glanzer, Simon; Zangger, Klaus (13 April 2015). "Visualizing Unresolved Scalar Couplings by Real-Time-Upscaled NMR". Journal of the American Chemical Society. 137 (15): 5163–5169. Bibcode:2015JAChS.137.5163G. doi:10.1021/jacs.5b01687. PMC 4415032. PMID 25837306.
- ^ Nagaoka, Shin'ichi (May 2007). "A Short History of Three Chemical Shifts". Journal of Chemical Education. 84 (5): 801. Bibcode:2007JChEd..84..801N. doi:10.1021/ed084p801.
External links
[edit]- chem.wisc.edu
- BioMagResBank
- NMR Table
- Proton chemical shifts
- Carbon chemical shifts
- Online tutorials (these generally involve combined use of IR, 1H NMR, 13C NMR and mass spectrometry)
- Problem set 1 (see also this link for more background information on spin-spin coupling)
- Problem set 2
- Problem set 4
- Problem set 5
- Combined solutions to problem set 5 (Problems 1–32) and (Problems 33–64)
Chemical shift
View on GrokipediaFundamentals of Chemical Shift
Definition and Basic Principles
In nuclear magnetic resonance (NMR) spectroscopy, the chemical shift (δ) is defined as the difference in the resonance frequency of a nucleus relative to that of a standard reference compound, expressed in parts per million (ppm) to render it independent of the spectrometer's magnetic field strength.[3] This dimensionless quantity arises because the resonance frequency of a nucleus is proportional to the applied magnetic field, so normalizing the frequency difference by the operating frequency ensures comparability across different instruments. The basic principle underlying chemical shift stems from the local electronic environment surrounding the nucleus, which induces variations in the effective magnetic field experienced by the nucleus. In NMR, nuclei with the same gyromagnetic ratio but placed in different molecular contexts—such as varying bonds, substituents, or geometries—encounter differing degrees of shielding or deshielding from surrounding electrons, leading to shifts in their absorption frequencies.[3] This phenomenon allows NMR to distinguish between nuclei in unique chemical environments, providing structural insights into molecules. The term "chemical shift" was first coined in 1950 by Warren G. Proctor and Fu-Chun Yu during early NMR experiments, where they unexpectedly observed distinct resonance frequencies for the two nitrogen nuclei in ammonium nitrate (NH₄NO₃), attributing the difference to the chemical environment.[4] Their discovery marked the recognition of how chemical structure influences NMR signals, laying the foundation for high-resolution NMR spectroscopy.[4] The chemical shift is quantitatively expressed by the equation: where is the resonance frequency of the sample nucleus in Hz, is the resonance frequency of the reference standard in Hz, and is the operating frequency of the spectrometer (typically in MHz for protons).[3] This formula derives from the Larmor frequency relation , where small perturbations in the local field (with as the shielding constant) cause frequency differences ; dividing by and scaling by yields the ppm scale for practical measurement. By convention, deshielded nuclei (experiencing higher effective magnetic fields) appear at higher δ values (downfield), while shielded nuclei (experiencing lower effective fields) appear upfield.[3] A representative example illustrates this principle for protons (^1H): in alkanes, methyl (CH₃) protons typically resonate at 0.9–1.0 ppm and methylene (CH₂) protons at 1.2–1.4 ppm due to their non-polar hydrocarbon environment, whereas aldehyde protons (RCHO) appear far downfield at 9–10 ppm owing to the deshielding effect of the electron-withdrawing carbonyl group.[5] Tetramethylsilane (TMS) serves as the universal reference standard, assigned δ = 0 ppm for protons.[6]Measurement and Units
In nuclear magnetic resonance (NMR) spectroscopy, chemical shifts are measured by recording the resonance frequency of a nucleus relative to a reference standard, initially expressed as a frequency difference in hertz (Hz).[7] This difference arises from the local magnetic environment of the nucleus and is determined during the acquisition of the free induction decay (FID) signal in the spectrometer. Modern NMR instruments, such as Fourier transform (FT) spectrometers, detect these frequencies in the time domain and convert them to the frequency domain via Fourier transformation, yielding spectra where peaks correspond to specific chemical environments.[7] To achieve field-independent reporting, the frequency difference in Hz is converted to the chemical shift scale in parts per million (ppm), a dimensionless unit defined by the International Union of Pure and Applied Chemistry (IUPAC).[8] The conversion is performed automatically by spectrometer software using the formula: where is the resonance frequency of the sample, is the reference frequency (both in Hz), and is the spectrometer's operating frequency (in MHz).[7] Positive values indicate deshielding (downfield shifts, higher frequency relative to the reference), while negative values denote shielding (upfield shifts).[8] For H NMR in organic compounds, chemical shifts typically span 0 to 12 ppm, with rare negative shifts (e.g., -1 to -2 ppm) in cases like protons influenced by aromatic ring currents, and more negative values (down to -10 ppm) in organometallic hydride protons.[7][9] The accuracy of these measurements relies on proper instrument preparation, including locking and shimming. Locking involves using a deuterium lock signal (from a deuterated solvent) to stabilize the magnetic field against drifts, while shimming adjusts gradient coils to homogenize the field, minimizing linewidths and ensuring precise peak positions. Without adequate locking and shimming, spectral distortions can shift apparent peak frequencies by several Hz, propagating errors into the ppm scale. A key advantage of the ppm scale is its independence from the magnetic field strength, allowing consistent chemical shift values across instruments operating at different frequencies, such as 60 MHz or 900 MHz spectrometers.[7] For instance, a 600 Hz difference at 60 MHz corresponds to 10 ppm, just as a 9000 Hz difference does at 900 MHz, facilitating direct comparison of data from diverse setups. Common sources of error in chemical shift measurement include sample impurities, which introduce extraneous peaks or broaden signals, complicating accurate peak identification and positioning.[10] Poor spectral resolution, often due to magnetic field inhomogeneities or low signal-to-noise ratios, can also lead to imprecise determination of peak maxima, with errors up to 0.1-0.5 ppm in low-quality spectra.[11] These issues are mitigated by employing high-field NMR spectrometers (e.g., 500 MHz or higher), which enhance resolution through increased chemical shift dispersion and better separation of overlapping signals.Theoretical Basis
Induced Magnetic Field
In nuclear magnetic resonance (NMR) spectroscopy, the applied external magnetic field induces currents in the electrons surrounding a nucleus, generating a secondary local magnetic field that modifies the field experienced by the nucleus.[12] This induced field arises from the response of the molecular electron cloud to , creating circulating electron currents analogous to those in a loop of wire, which produce according to the Biot-Savart law.[13] The direction and magnitude of depend on the electronic structure, typically opposing at the nucleus in diamagnetic systems, thereby shielding it from the full external field.[12] The effective magnetic field at the nucleus, , is the vector sum , where accounts for any additional external contributions, though it is often negligible in standard NMR setups. The resonance frequency of the nucleus is then given by the Larmor equation: where is the gyromagnetic ratio of the nucleus. Since modulates , variations in lead to shifts in , with the induced field typically reducing and thus lowering the observed frequency compared to a bare nucleus.[12] Approximating , where is the shielding constant (a dimensionless quantity between 0 and 1 for most cases), yields , directly linking the induced field to the chemical shift.[13] The induced magnetic field encompasses both diamagnetic and paramagnetic contributions. In diamagnetic molecules, which lack unpaired electrons and dominate organic compounds, the diamagnetic term arises from the orbital motion of paired electrons and typically produces an opposing that shields the nucleus. Paramagnetic contributions, stemming from unpaired electrons, can enhance or reverse the field direction, leading to deshielding, but these are rare in standard diamagnetic samples used in routine NMR. Visual representations of the induced field often depict field lines circulating around the nucleus due to electron loops, with pointing opposite to within the electron cloud, akin to the demagnetizing field in a superconducting ring. This circulation creates a local environment where the field strength diminishes toward the nucleus, emphasizing the shielding effect.[13] The quantum mechanical foundation for was established using second-order perturbation theory to calculate the induced electron currents and their magnetic effects on the nucleus, as introduced by Norman Ramsey in 1950. This approach treats the external field as a perturbation that mixes excited electronic states into the ground state wavefunction, yielding the shielding constant through integrals over electronic orbitals.[12]Diamagnetic Shielding
Diamagnetic shielding arises from the weak opposition to the external magnetic field generated by induced loops of electron circulation around the nucleus, which reduces the effective magnetic field at the nucleus. This contribution, denoted as , is a key component of the total magnetic shielding in NMR spectroscopy. The quantitative expression for diamagnetic shielding in atoms is given by the Lamb formula: where is the elementary charge, the electron mass, the speed of light, and the expectation value of the inverse distance between an electron and the nucleus, averaged over the ground-state electron density for all electrons.[14] This formula originates from a classical treatment of electron Larmor precession in the magnetic field, assuming spherical symmetry of the electron distribution; it provides the leading-order term for the induced magnetic moment opposing . In atomic units, the expression involves the fine-structure constant : (converted to ppm). The magnitude of is typically 0–100 ppm for light nuclei like and , reflecting compact orbitals with smaller ; for heavier atoms, it increases significantly due to more diffuse orbitals that enhance this average. The total shielding is , with the paramagnetic term arising from excited-state contributions; in closed-shell molecules, dominates because filled orbitals minimize paramagnetic effects. In noble gases, diamagnetic shielding exemplifies this dominance, as seen in , where the atomic shielding reaches approximately 6600 ppm, driven by the high electron density in the closed-shell configuration.Factors Influencing Chemical Shifts
The chemical shift in NMR spectroscopy is modulated by several molecular and environmental factors that alter the electron density and local magnetic field around the nucleus, primarily affecting the diamagnetic shielding component. These factors lead to deshielding (downfield shifts) or shielding (upfield shifts) relative to a reference, enabling structural elucidation. Electronegativity of adjacent atoms plays a crucial role through inductive effects, where electron-withdrawing groups reduce electron density around the observed nucleus, causing deshielding. For instance, the high electronegativity of fluorine deshields the protons in CH₃F, resulting in a chemical shift of 4.26 ppm compared to 0.23 ppm for the protons in CH₄.[15] This effect diminishes with distance from the electronegative atom, typically influencing protons within two or three bonds. Hybridization influences chemical shifts by changing the s-character of bonding orbitals, which pulls electrons closer to the nucleus and enhances deshielding. sp³-hybridized carbons exhibit shifts in the 0–50 ppm range for ¹³C NMR, while sp²-hybridized carbons appear at 100–200 ppm due to the higher 33% s-character compared to 25% in sp³. In aromatic systems, the pi-electron ring current generates magnetic anisotropy that deshields protons in the ring plane; for example, the methyl protons in toluene resonate at 2.34 ppm, deshielded by approximately 1.4 ppm relative to typical alkane methyl protons near 0.9 ppm.[16] Hydrogen bonding significantly deshields protons involved in the interaction by polarizing the electron cloud and reducing shielding. In alcohols, OH protons typically appear between 1 and 5 ppm, with the exact position varying with concentration: dilute solutions show upfield shifts (near 1 ppm) for monomeric forms, while concentrated samples exhibit downfield shifts (2–5 ppm) due to intermolecular hydrogen bonds.[17] Steric and conformational effects arise from spatial arrangements that alter local electron distribution or anisotropic fields. In cyclohexane derivatives at low temperatures, axial and equatorial protons differ by about 0.5 ppm due to distinct orientations relative to surrounding bonds, with axial protons generally more deshielded.[18] Solvent effects stem from interactions like hydrogen bonding or changes in the dielectric constant, which can deshield or shield nuclei. Polar solvents such as DMSO often cause greater deshielding than nonpolar ones like chloroform; for many ¹H signals, this results in shifts of 0.5–1 ppm, with variations up to 4.6 ppm for polar protons.| Factor | Description | Example |
|---|---|---|
| Electronegativity | Inductive withdrawal by electron-withdrawing groups reduces electron density, deshielding nuclei. | CH₃F protons at 4.26 ppm vs. CH₄ at 0.23 ppm (¹H NMR).[15] |
| Hybridization/Anisotropy | Higher s-character in sp² vs. sp³ increases deshielding; ring currents in aromatics create anisotropic fields. | sp³ carbons 0–50 ppm, sp² 100–200 ppm (¹³C NMR); toluene CH₃ at 2.34 ppm (¹H NMR).[16] |
| Hydrogen Bonding | Polarizes bonds, deshielding involved protons; concentration-dependent. | Alcohol OH 1–5 ppm, shifts downfield with increasing H-bonding (¹H NMR).[17] |
| Steric/Conformational | Spatial orientation alters local fields, causing differences in conformers. | Axial vs. equatorial protons in cyclohexane differ by ~0.5 ppm (¹H NMR).[18] |
| Solvent Effects | Polarity and H-bonding capability modulate shielding via solute-solvent interactions. | ¹H shifts of 0.5–1 ppm from CDCl₃ to DMSO-d₆. |
Referencing and Standardization
Chemical Shift Referencing
Chemical shift referencing is essential in nuclear magnetic resonance (NMR) spectroscopy to establish a universal zero point on the chemical shift scale (δ), measured in parts per million (ppm), thereby allowing consistent comparison of spectral data across different samples, solvents, instruments, and laboratories. Without a reference standard, the observed resonance frequencies would be arbitrary and dependent on the spectrometer's magnetic field strength, rendering inter-sample comparisons impossible. The IUPAC recommends a unified chemical shift scale for all nuclei based on the ¹H resonance of tetramethylsilane (TMS) as the primary reference, set at δ = 0 ppm, to ensure reproducibility and standardization.[19] Internal referencing, where the standard is added directly to the sample solution, is the preferred method for most routine NMR experiments due to its simplicity and the homogeneity it provides in the magnetic environment, minimizing susceptibility differences. For example, a small amount of TMS (typically <1% v/v) is added to organic solvents like CDCl₃, where its methyl protons and carbon produce a single sharp peak at 0 ppm for ¹H and ¹³C NMR, respectively. In contrast, external referencing involves measuring the sample and standard separately—often using coaxial tubes or the substitution method—and requires corrections for bulk magnetic susceptibility differences between the sample and reference compartments. This approach is particularly useful for air-sensitive or reactive samples where adding an internal standard could compromise the sample integrity or introduce contaminants.[19][1] Ideal reference standards must exhibit several key properties to ensure accurate and reliable calibration: chemical inertness to avoid reactions with the sample, low volatility to prevent evaporation during measurement (though TMS is used in dilute solutions despite its moderate volatility), and a single, sharp, intense resonance peak that does not overlap with typical sample signals. TMS satisfies these criteria exceptionally well due to its tetrahedral symmetry, which yields equivalent protons and carbons, and the low electronegativity of silicon, which shields its nuclei to position the signal at high field (0 ppm). For aqueous solutions, where TMS has limited solubility, IUPAC guidelines recommend secondary references such as sodium 3-(trimethylsilyl)propane-1-sulfonate (DSS), whose methyl ¹H resonance is set to 0 ppm, providing a water-soluble alternative with similar desirable properties. These 2001 IUPAC recommendations (building on earlier 1999 proposals for reporting standards) emphasize using secondary references only when primary ones like TMS are impractical, with their positions calibrated relative to TMS via the frequency ratio Ξ.[19][1][19] To maintain field homogeneity and stability during spectral acquisition, NMR spectrometers employ a lock signal, typically from the deuterium (²H) resonance of the deuterated solvent (e.g., CDCl₃), which continuously monitors and adjusts for magnetic field drifts. This lock mechanism ensures that chemical shifts remain consistent throughout the experiment, with any necessary corrections applied based on the difference between the lock frequency and the ¹H reference frequency. By integrating referencing with the lock system, variations due to temperature, solvent, or instrumental fluctuations are minimized, upholding the precision of the δ scale.[19]Referencing Methods
Direct referencing involves adding a standard compound directly to the NMR sample to serve as an internal reference for chemical shift calibration. For proton (¹H) and carbon-13 (¹³C) NMR in organic solvents like CDCl₃, tetramethylsilane (TMS) is commonly used at concentrations below 1% volume fraction to avoid signal overlap or distortion, with the methyl resonance set to 0 ppm.[20] In aqueous solutions, particularly for biomolecules, sodium 2,2-dimethyl-2-silapentane-5-sulfonate (DSS) is preferred at approximately 10 mmol/dm³, also set to 0 ppm for the methyl group, due to its solubility and minimal interaction with biological samples.[20] The procedure typically entails acquiring a ¹H spectrum first to confirm the reference peak position before scaling other nuclei spectra accordingly. Indirect referencing relies on the deuterium (²H) lock signal from the deuterated solvent to calibrate the chemical shift scale without adding an internal standard, using predefined frequency ratios (Ξ values) from the IUPAC unified scale. For example, in DMSO-d₆, the residual ¹H signal of the CHD₂ group is set to 2.50 ppm relative to TMS in the ¹H spectrum, and the spectrometer adjusts the ²H lock frequency to maintain consistency across nuclei via the ²H Ξ value of 15.350609%.[20] This method ensures reproducibility without altering the sample composition and is standard for multinuclear experiments. Secondary standards provide alternatives tailored to specific sample types, such as biomolecules in aqueous media. For ¹H NMR in water-based solutions, 3-(trimethylsilyl)propionate-2,2,3,3-d₄ (TSP) is often employed at low concentrations with its methyl resonance defined as 0 ppm, offering advantages in solubility for biological matrices despite some pH sensitivity that requires careful control. In contrast, DSS serves as a more robust secondary option in such environments due to its lower sensitivity to pH variations, maintaining shift stability across typical biological pH ranges (e.g., 4–8). Specialized methods address challenging sample states like solids or gases. In solid-state NMR under magic angle spinning (MAS), adamantane is used as a reference for ¹³C shifts, with its methylene (CH₂) carbon signal at 37.77 ppm relative to TMS, enabling high-resolution calibration without bulk magnetic susceptibility corrections.[20] For volatile or gas-phase samples, low-pressure ³He gas serves as a temperature-independent standard, with a ³He Ξ value of 76.178976%, facilitating precise referencing in low-density environments.[20] Common pitfalls in referencing include using impure standards, which can lead to peak broadening from contaminants interfering with the lock or reference signals, necessitating high-purity reagents verified by prior spectroscopy.[20] Additionally, temperature dependence requires corrections; for instance, the TMS ¹H shift varies by approximately -0.0005 ppm/°C, potentially causing up to 0.01 ppm deviation over a 20°C range, so calibrations should specify and account for the measurement temperature.[20]Practical Considerations
Operating Frequency
The operating frequency of an NMR spectrometer, denoted as , is the Larmor frequency at which nuclei resonate in the applied magnetic field , given by the equation , where is the gyromagnetic ratio of the nucleus.[21] Higher operating frequencies correspond to stronger magnetic fields; for example, a 400 MHz spectrometer for H nuclei operates at approximately twice the field strength of a 100 MHz instrument, leading to greater separation of chemical shift differences when measured in hertz (Hz).[22] However, the chemical shift in parts per million (ppm) remains constant across field strengths because it is a relative measure, independent of , ensuring comparability of spectra regardless of the instrument used.[23] Increased operating frequency enhances spectral resolution by dispersing peaks over a wider frequency range in Hz, facilitating the separation of closely spaced signals. At 900 MHz for H NMR, aromatic protons in complex organic molecules exhibit significantly better peak separation compared to lower fields, allowing clearer identification of subtle structural differences.[24] Digital resolution, which determines the precision of peak definition in the processed spectrum, is calculated as the sweep width (the frequency range covered during acquisition) divided by the number of data points acquired.[25] This parameter improves with higher fields due to the expanded dispersion, though it requires careful optimization of acquisition parameters to avoid truncation artifacts. While higher frequencies boost sensitivity, with signal-to-noise ratio (S/N) scaling approximately as , they introduce practical challenges such as more difficult shimming to achieve field homogeneity.[26] At ultra-high fields beyond 1 GHz, magnetic field instabilities and sample-specific susceptibilities complicate shimming, potentially degrading resolution despite the theoretical gains.[27] In practice, low-field instruments operating at 60 MHz are commonly used in educational settings for straightforward spectra of small molecules, whereas high-field systems at 1 GHz or above are essential for resolving the crowded chemical shifts in complex mixtures like proteins.[28] The use of ppm for chemical shifts is particularly advantageous here, as the raw frequency differences in Hz scale linearly with , making absolute Hz values instrument-dependent and less useful for standardization.[29]Magnetic Properties of Common Nuclei
The magnetic properties of atomic nuclei, particularly their nuclear spin quantum number , gyromagnetic ratio , and natural isotopic abundance, fundamentally determine their suitability for nuclear magnetic resonance (NMR) spectroscopy and the observable range of chemical shifts. Nuclei with produce sharp, well-resolved signals because they lack a nuclear quadrupole moment, avoiding relaxation-induced broadening from interactions with electric field gradients.[19] Prominent examples include , , , and , which are routinely used in multi-nuclear NMR for structural analysis due to their favorable properties. In contrast, nuclei with , such as () and (), possess a quadrupole moment that often leads to significant line broadening in solution-state NMR, complicating spectral interpretation unless in highly symmetric environments.[19] The gyromagnetic ratio dictates the resonance frequency for a given magnetic field strength and influences signal sensitivity, as the equilibrium magnetization scales with . High values, as seen in and , yield stronger signals and wider chemical shift dispersions, enhancing resolution for distinguishing subtle electronic environments. Natural abundance further modulates detectability; low-abundance isotopes like (1.1%) require longer acquisition times or enhancements such as the nuclear Overhauser effect (NOE) via proton decoupling to achieve practical signal-to-noise ratios.[19] The following table summarizes key properties for common NMR-active nuclei, including typical chemical shift ranges observed in organic and biochemical contexts. These ranges reflect environmental influences on shielding, with broader dispersions for nuclei like providing valuable structural insights in multi-nuclear studies. Shift values are relative to standard references (e.g., TMS for and , 85% HPO for , CFCl for ).| Nucleus | Spin | Natural Abundance (%) | (MHz/T) | Typical Shift Range (ppm) |
|---|---|---|---|---|
| 1/2 | 99.99 | 42.58 | 0–12 | |
| 1/2 | 1.1 | 10.71 | 0–220 | |
| 1/2 | 100 | 17.24 | –50 to +100 | |
| 1/2 | 100 | 40.08 | –300 to +300 | |
| 1 | 0.015 | 6.54 | 0–10 (broadened) | |
| 1 | 99.6 | 3.08 | Variable, often broadened |


