Capnography is the monitoring of the concentration or partial pressure of carbon dioxide (CO
2
) in the respiratory gases. Its main development has been as a monitoring tool for use during anesthesia and intensive care. It is usually presented as a graph of CO
2
(measured in kilopascals, "kPa" or millimeters of mercury, "mmHg") plotted against time, or, less commonly, but more usefully, expired volume (known as volumetric capnography). The plot may also show the inspired CO
2
, which is of interest when rebreathing systems are being used. When the measurement is taken at the end of a breath (exhaling), it is called "end tidal" CO
2
(PETCO2).[1]

The capnogram is a direct monitor of the inhaled and exhaled concentration or partial pressure of CO
2
, and an indirect monitor of the CO
2
partial pressure in the arterial blood. In healthy individuals, the difference between arterial blood and expired gas CO
2
partial pressures is very small (normal difference 4-5 mmHg). In the presence of most forms of lung disease, and some forms of congenital heart disease (the cyanotic lesions) the difference between arterial blood and expired gas increases which can be an indication of new pathology or change in the cardiovascular-ventilation system.[2][3]

Medical Use

[edit]

Oxygenation and capnography, although related, remain distinct elements in the physiology of respiration. Ventilation refers to the mechanical process of which the lungs expand and exchange volumes of gasses, however respiration further describes the exchange of gasses (mainly CO
2
and O
2
) at the level of the alveoli. The process of respiration can be divided into two main functions: elimination of CO
2
waste and replenishing tissues with fresh O
2
. Oxygenation (typically measured via pulse oximetry) measures the latter portion of this system. Capnography measures the elimination of CO
2
which may be of greater clinical usefulness than oxygenation status.[4]

During the normal cycle of respiration, a single breath can be divided into two phases: inspiration and expiration. At the beginning of inspiration, the lungs expand and CO
2
free gasses fill the lungs. As the alveoli are filled with this new gas, the concentration of CO
2
that fills the alveoli is dependent on the ventilation of the alveoli and the perfusion (blood flow) that is delivering the CO
2
for exchange. Once expiration begins to occur, the lung volume decreases as air is forced out the respiratory tract. The volume of CO
2
that is exhaled at the end of exhalation is generated as a by product of metabolism from tissue throughout the body. The delivery of CO
2
to the alveoli for exhalation is dependent on an intact cardiovascular system to ensure adequate blood flow from the tissue to the alveoli. If cardiac output (the amount of blood that is pumped out of the heart) is decreased, the ability to transport CO
2
is also decreased which is reflected in a decreased expired amount of CO
2
. The relationship of cardiac output and end tidal CO
2
is linear, such that as cardiac output increases or decreases, the amount of CO
2
is also adjusted in the same manner. Therefore the monitoring of end tidal CO
2
can provide vital information on the integrity of the cardiovascular system, specifically how well the heart is able to pump blood.[5]

The amount of CO
2
that is measured during each breath requires an intact cardiovascular system to delivery the CO
2
to the alveoli which is the functional unit of the lungs. During phase I of expiration, the CO
2
transported to the lungs gas occupies a given space that is not involved in gas exchange, called dead space. Phase II of expiration is when the CO
2
within the lungs is forced up the respiratory tract on its way to leave the body, which causes mixing of the air from the dead space with the air in the functional alveoli responsible for gas exchange. Phase III is the final portion of expiration which reflects CO
2
only from the alveoli and not the dead space. These three phases are important to understand in clinical scenarios since a change in the shape and absolute values can indicate respiratory and/or cardiovascular compromise.[6]

Source of CO2 Content During Exhalation
Capnography

Applications

[edit]
  • Assessing Airway Integrity
  • Confirmation of Endotracheal Tube Placement
  • Predictor of Outcomes in the Intensive Care Unit
  • Intraoperative Complications (ie. air embolism, thromboembolism, etc.)
  • CPR use in ACLS (Advanced Cardiovascular Life Support)
  • Procedural Sedation Monitoring

Anesthesia

[edit]
Capnograph

During anesthesia, there is interplay between two components: the patient and the anesthesia administration device (which is usually a breathing circuit and a ventilator). The critical connection between the two components is either an endotracheal tube or a mask, and CO
2
is typically monitored at this junction. Capnography directly reflects the elimination of CO
2
by the lungs to the anesthesia device. Indirectly, it reflects the production of CO
2
by tissues and the circulatory transport of CO
2
to the lungs.[7]

When expired CO
2
is related to expired volume rather than time, the area beneath the curve represents the volume of CO
2
in the breath, and thus over the course of a minute, this method can yield the CO
2
per minute elimination, an important measure of metabolism. Sudden changes in CO
2
elimination during lung or heart surgery usually imply important changes in cardiorespiratory function.[8]

Capnography has been shown to be more effective than clinical judgement alone in the early detection of adverse respiratory events such as hypoventilation, esophageal intubation and circuit disconnection; thus allowing patient injury to be prevented. During procedures done under sedation, capnography provides more useful information, e.g. on the frequency and regularity of ventilation, than pulse oximetry.[9][10]

Capnography provides a rapid and reliable method to detect life-threatening conditions (malposition of tracheal tubes, unsuspected ventilatory failure, circulatory failure and defective breathing circuits) and to circumvent potentially irreversible patient injury.

Capnography and pulse oximetry together could have helped in the prevention of 93% of avoidable anesthesia mishaps according to an ASA (American Society of Anesthesiologists) closed claim study.[11]

Emergency medical services

[edit]

Capnography is increasingly being used by EMS personnel to aid in their assessment and treatment of patients in the prehospital environment. These uses include verifying and monitoring the position of an endotracheal tube or a blind insertion airway device. A properly positioned tube in the trachea guards the patient's airway and enables the paramedic to breathe for the patient. A misplaced tube in the esophagus can lead to the patient's death if it goes undetected. [12]

A study in the March 2005 Annals of Emergency Medicine, comparing field intubations that used continuous capnography to confirm intubations versus non-use showed zero unrecognized misplaced intubations in the monitoring group versus 23% misplaced tubes in the unmonitored group.[13] The American Heart Association (AHA) affirmed the importance of using capnography to verify tube placement in their 2005 CPR and Emergency Cardiovascular Care Guidelines.[14]

The AHA also notes in their new guidelines that capnography, which indirectly measures cardiac output, can also be used to monitor the effectiveness of CPR and as an early indication of return of spontaneous circulation (ROSC). Studies have shown that when a person doing CPR tires, the patient's end-tidal CO
2
(PETCO2, the level of carbon dioxide released at the end of expiration) falls, and then rises when a fresh rescuer takes over. Other studies have shown when a patient experiences return of spontaneous circulation, the first indication is often a sudden rise in the PETCO2 as the rush of circulation washes untransported CO
2
from the tissues. Likewise, a sudden drop in PETCO2 may indicate the patient has lost pulses and CPR may need to be initiated.[15]

Paramedics are also now beginning to monitor the PETCO2 status of nonintubated patients by using a special nasal cannula that collects the carbon dioxide. A high PETCO2 reading in a patient with altered mental status or severe difficulty breathing may indicate hypoventilation and a possible need for the patient to be intubated. Low PETCO2 readings on patients may indicate hyperventilation.[16]

Capnography, because it provides a breath by breath measurement of a patient's ventilation, can quickly reveal a worsening trend in a patient's condition by providing paramedics with an early warning system into a patient's respiratory status. When compared to oxygenation which is measured by pulse oximetry, there are several disadvantages that capnography can help address to provide a more accurate reflection of cardiovascular integrity. One shortcoming of measuring pulse oximetry alone is that administration of supplemental oxygen (ie. via nasal cannula) can delay desaturation in a patient if they stopped breathing, therefore delaying medical intervention. Capnography provides a rapid way to directly assess ventilation status and indirectly assess cardiac function. Clinical studies are expected to uncover further uses of capnography in asthma, congestive heart failure, diabetes, circulatory shock, pulmonary embolus, acidosis, and other conditions, with potential implications for the prehospital use of capnography.[17]

Registered nurses

[edit]

Registered nurses, but more so RRTs (respiratory therapists), in critical care settings may use capnography to determine if a nasogastric tube, which is used for feeding, has been placed in the trachea as opposed to the esophagus.[18] Usually a patient will cough or gag if the tube is misplaced, but most patients in critical care settings are sedated or comatose. If a nasogastric tube is accidentally placed in the trachea instead of the esophagus, the tube feedings will go into the lungs, which is a life-threatening situation. If the monitor displays typical CO
2
waveforms then placement should be confirmed.[19]

Diagnostic usage

[edit]

Capnography provides information about CO
2
production, pulmonary (lung) perfusion, alveolar ventilation, respiratory patterns, and elimination of CO
2
from the anesthesia breathing circuit and ventilator. The shape of the curve is affected by some forms of lung disease; in general there are obstructive conditions such as bronchitis, emphysema and asthma, in which the mixing of gases within the lung is affected.[20]

Conditions such as pulmonary embolism and congenital heart disease, which affect perfusion of the lung, do not, in themselves, affect the shape of the curve, but greatly affect the relationship between expired CO
2
and arterial blood CO
2
. Capnography can also be used to measure carbon dioxide production, a measure of metabolism. Increased CO
2
production is seen during fever and shivering. Reduced production is seen during anesthesia and hypothermia.[21]

Working mechanism

[edit]
Schematic overview of a capnograph

Capnographs work on the principle that CO
2
is a polyatomic gas and therefore absorbs infrared radiation. A beam of infrared light is passed across the gas sample to fall on a sensor. The presence of CO
2
in the gas leads to a reduction in the amount of light falling on the sensor, which changes the voltage in a circuit. The analysis is rapid and accurate, but the presence of nitrous oxide in the gas mix changes the infrared absorption via the phenomenon of collision broadening.[22] This must be corrected for measuring the CO
2
in human breath by measuring its infrared absorptive power. This was established as a reliable technique by John Tyndall in 1864, though 19th and early 20th century devices were too cumbersome for everyday clinical use.[23] Today, technologies have since improved and are able to measure the values of CO
2
near instantaneously and has become a standard practice in medical settings. There are currently two main types of CO
2
sensors that are used in clinical practice: main-stream sensors and side-stream sensors. Both effectively serve the same function to quantify the amount of CO
2
that is being exhaled in each breath.

Capnogram model

[edit]

The capnogram waveform provides information about various respiratory and cardiac parameters. The capnogram double-exponential model attempts to quantitatively explain the relationship between respiratory parameters and the exhalatory segment of a capnogram waveform.[24] According to the model, each exhalatory segment of capnogram waveform follows the analytical expression:

where

  • represents the partial pressure of carbon dioxide measured by the capnogram as a function of time since the beginning of exhalation.
  • represents the alveolar partial pressure of carbon dioxide.
  • represents the inverse of the dead space fraction (i.e. the ratio of tidal volume to dead space volume).
  • represents the pulmonary time constant (i.e. the product of pulmonary resistance and compliance)

In particular, this model explains the rounded "shark-fin" shape of the capnogram observed in patients with obstructive lung disease.

See also

[edit]

Citations

[edit]
  1. ^ Bhavani-Shankar K, Philip J (October 2000). "Defining segments and phases of a time capnogram". Anesth Analg. 91 (4): 973–977. doi:10.1097/00000539-200010000-00038. PMID 11004059. S2CID 46505268.
  2. ^ Nunn J, Hill D (May 1960). "Respiratory dead space and arterial to end-tidal carbon dioxide tension difference in anesthetized man". J Appl Physiol. 15: 383–389. doi:10.1152/jappl.1960.15.3.383. PMID 14427915.
  3. ^ Williams E, Dassios T, Greenough A (October 2021). "Carbon dioxide monitoring in the newborn". Pediatr Pulmonol. 56 (10): 3148–3156. doi:10.1002/ppul.25605. PMID 34365738. S2CID 236960627.
  4. ^ Lam T, Nagappa M, Wong J, Singh M, Wong D, Chung F (December 2017). "Continuous Pulse Oximetry and Capnography Monitoring for Postoperative Respiratory Depression and Adverse Events: A Systematic Review and Meta-analysis". Anesthesia & Analgesia. 125 (6): 2019–2029. doi:10.1213/ANE.0000000000002557. ISSN 0003-2999. PMID 29064874. S2CID 13950478.
  5. ^ Siobal M (October 2016). "Monitoring Exhaled Carbon Dioxide". Respir Care. 61 (10): 1397–1416. doi:10.4187/respcare.04919. PMID 27601718. S2CID 12532311.
  6. ^ Benumof J (April 1998). "Interpretation of capnography". AANA J. 661 (2): 169–176.
  7. ^ Weil M, Bisera J, Trevino, Rackow E (October 2016). "Cardiac output and end-tidal carbon dioxide". Crit Care Med. 13 (11): 907–909. doi:10.1097/00003246-198511000-00011. PMID 3931979. S2CID 34223367.
  8. ^ J. S. Gravenstein, Michael B. Jaffe, Nikolaus Gravenstein, David A. Paulus, eds. (17 March 2011). Capnography (2 ed.). Cambridge University Press. ISBN 978-0-521-51478-1. OCLC 1031490358.
  9. ^ Lightdale JR, Goldmann DA, Feldman HA, Newburg AR, DiNardo JA, Fox VL (June 2006). "Microstream capnography improves patient monitoring during moderate sedation: a randomized, controlled trial". Pediatrics. 117 (6): e1170–1178. doi:10.1542/peds.2005-1709. ISSN 1098-4275. PMID 16702250. S2CID 2857581.
  10. ^ Burton JH, Harrah JD, Germann CA, Dillon DC (May 2006). "Does end-tidal carbon dioxide monitoring detect respiratory events prior to current sedation monitoring practices?". Academic Emergency Medicine. 13 (5): 500–504. doi:10.1197/j.aem.2005.12.017. ISSN 1553-2712. PMID 16569750.
  11. ^ Tinker JH, Dull DL, Caplan RA, Ward RJ, Cheney FW (1989). "Role of Monitoring Devices in Prevention of Anesthetic Mishaps". Anesthesiology. 71 (4): 541–546. doi:10.1097/00000542-198910000-00010. PMID 2508510.
  12. ^ Katz S, Falk J (January 2001). "Misplaced endotracheal tubes by paramedics in an urban emergency medical services system". Ann Emerg Med. 37 (1): 32–37. doi:10.1067/mem.2001.112098. PMID 11145768.
  13. ^ Silvestri S, Ralls GA, Krauss B, Thundiyil J, Rothrock SG, Senn A, Carter E, Falk J (May 2005). "The effectiveness of out-of-hospital use of continuous end-tidal carbon dioxide monitoring on the rate of unrecognized misplaced intubation within a regional emergency medical services system". Annals of Emergency Medicine. 45 (5): 497–503. doi:10.1016/j.annemergmed.2004.09.014. ISSN 1097-6760. PMID 15855946.
  14. ^ Hazinski MF, Nadkarni VM, Hickey RW, O'Connor R, Becker LB, Zaritsky A (13 December 2005). "Major Changes in the 2005 AHA Guidelines for CPR and ECC". Circulation. 112 (24_supplement): IV–206. doi:10.1161/CIRCULATIONAHA.105.170809. PMID 16314349. S2CID 934519.
  15. ^ Long B, Koyfman A, Vivirito MA (December 2017). "Capnography in the Emergency Department: A Review of Uses, Waveforms, and Limitations". The Journal of Emergency Medicine. 53 (6): 829–842. doi:10.1016/j.jemermed.2017.08.026. ISSN 0736-4679. PMID 28993038.
  16. ^ Davis D, Dunford J, Ochs M, Park K, Hoyt D (April 2004). "The use of quantitative end-tidal capnometry to avoid inadvertent severe hyperventilation in patients with head injury after paramedic rapid sequence intubation". J Trauma. 56 (4): 808–814. doi:10.1097/01.TA.0000100217.05066.87. PMID 15187747.
  17. ^ "Experts: Where capnography is headed". EMS1. 20 November 2013. Retrieved 16 November 2021.
  18. ^ Potter, Patricia Ann, and Anne Griffin Perry. "Nutrition." Essentials for nursing practice. Eighth ed. St. Louis: Elsevier, 2015. 940. Print.
  19. ^ Roubenoff R, Ravich W (April 1998). "Pneumothorax due to nasogastric feeding tubes. Report of four cases, review of the literature, and recommendations for prevention". Arch Intern Med. 149 (1): 184–188. doi:10.1001/archinte.1989.00390010156022. PMID 2492185.
  20. ^ Yaron M, Padyk P, Hutsinpiller M, Cairns C (October 1996). "Utility of the expiratory capnogram in the assessment of bronchospasm". Ann Emerg Med. 28 (4): 403–407. doi:10.1016/S0196-0644(96)70005-7. PMID 8839525.
  21. ^ Danzl D (February 2002). "Hypothermia system". Semin Respir Crit Care Med. 23 (1): 57–68. doi:10.1055/s-2002-20589. PMID 16088598. S2CID 260321037.
  22. ^ Raemer DB, Calalang I (April 1991). "Accuracy of end-tidal carbon dioxide tension analyzers". J Clin Monit. 7 (2): 195–208. doi:10.1007/BF01618124. PMID 1906531. S2CID 33836449.
  23. ^ Jaffe MB (September 2008). "Infrared measurement of carbon dioxide in the human breath: "breathe-through" devices from Tyndall to the present day". Anesth. Analg. 107 (3): 890–904. doi:10.1213/ane.0b013e31817ee3b3. PMID 18713902. S2CID 15610449.
  24. ^ Abid A (May 2017). "Model-Based Estimation of Respiratory Parameters from Capnography, With Application to Diagnosing Obstructive Lung Disease". IEEE Transactions on Biomedical Engineering. 64 (12): 2957–2967. doi:10.1109/TBME.2017.2699972. hdl:1721.1/134854. PMID 28475040. S2CID 206616144.
[edit]
Revisions and contributorsEdit on WikipediaRead on Wikipedia
from Grokipedia
Capnography is a non-invasive monitoring technique that measures the concentration or of (CO₂) in exhaled respiratory gases, providing both numerical values of end-tidal CO₂ (ETCO₂) and a graphical known as a capnogram. This method reflects the adequacy of ventilation, pulmonary blood flow, and CO₂ elimination, with ETCO₂ typically approximating arterial CO₂ levels (PaCO₂) within 4-5 mm Hg under normal conditions. It has become the for confirming endotracheal tube placement and monitoring ventilation during general , procedural , and critical care scenarios. The capnogram waveform consists of distinct phases: Phase I represents dead-space gas with zero CO₂ during early exhalation; Phase II is the rapid upstroke as alveolar gas mixes in; Phase III forms the alveolar plateau, where ETCO₂ is measured at its peak; and Phase IV is the inspiratory downstroke returning to baseline. Abnormal waveforms can indicate issues such as airway obstruction (prolonged upstroke), (elevated ETCO₂), or esophageal (absent or flat waveform). Devices employ for CO₂ detection, available in mainstream (sensor at the airway) or sidestream (aspirated gas sampling) configurations, with sidestream being more versatile for non-intubated patients. Clinically, capnography enhances patient safety by detecting respiratory depression up to four minutes before desaturation, reducing hypoxia risk by approximately 31% during . It is recommended by guidelines from organizations like the for moderate to deep and by the for CPR quality assessment. In non-intubated settings, such as postoperative care or emergency transport, it aids in identifying apnea or , particularly in high-risk patients like those with or use. Historical development traces back to infrared analyzers in , with widespread clinical adoption following 1991 guidelines establishing it as essential in .

Fundamentals

Definition and Physiological Basis

Capnography is defined as the continuous, noninvasive monitoring of the partial pressure or concentration of carbon dioxide (CO₂) in exhaled respiratory gases, with a primary focus on end-tidal CO₂ (EtCO₂), the maximum CO₂ level at the end of expiration. This technique provides a real-time graphical display known as a capnogram, reflecting ventilatory status breath by breath. The physiological basis of capnography stems from the body's CO₂ dynamics. CO₂ is generated as a metabolic byproduct during cellular aerobic respiration, where oxygen combines with glucose and other substrates to produce energy, water, and CO₂. In the bloodstream, CO₂ is transported mainly as bicarbonate ions (about 70%, formed via carbonic anhydrase converting CO₂ and water to H₂CO₃, which dissociates into HCO₃⁻ and H⁺), dissolved CO₂ (roughly 10%), and carbaminohemoglobin (approximately 20%, where CO₂ binds to hemoglobin's amino groups). Elimination occurs in the lungs, where CO₂ diffuses across the alveolar-capillary membrane from blood (PaCO₂ ≈ 35-45 mmHg) to alveolar air (≈ 40 mmHg) during expiration, driven by the partial pressure gradient. Central to this process are alveolar ventilation—the effective exchange of gases in the alveoli (calculated as VA = VE - VD, where VE is and VD is dead space)—and dead space ventilation, which comprises anatomical dead space (non-gas-exchanging conducting airways, ≈150 mL in adults) and physiological dead space (ventilated but underperfused alveoli). EtCO₂ serves as a surrogate for arterial PaCO₂, with a normal PaCO₂-EtCO₂ gradient of 2-5 mmHg in healthy individuals, reflecting efficient ventilation-perfusion matching; wider gradients indicate mismatches like increased dead space. Clinically, capnography confirms endotracheal tube placement by verifying exhaled CO₂ presence, evaluates ventilation adequacy by tracking EtCO₂ trends, and enables early detection of (rising EtCO₂ >45 mmHg) or (falling EtCO₂ <35 mmHg).

Historical Development

The foundations of capnography trace back to 19th-century experiments in gas analysis. In 1801, John Dalton formulated the law of partial pressures, establishing the principle that the total pressure of a gas mixture is the sum of the partial pressures of its components, which became essential for understanding respiratory gas dynamics. Building on this, physicist John Tyndall demonstrated in the 1860s that carbon dioxide absorbs infrared radiation, laying the groundwork for spectroscopic measurement techniques in exhaled gases. In the early 20th century, John Scott Haldane developed one of the first CO₂ analyzers, advancing the quantitative measurement of respiratory gases. These early insights into gas composition and optical properties set the stage for later quantitative monitoring of carbon dioxide in clinical settings. Advancements accelerated in the 20th century with the development of precise CO2 sensors. Concurrently, Karl Friedrich Luft pioneered modern infrared capnography in 1943 with the creation of the URAS (Ultrarod Absorption) device, a non-dispersive infrared analyzer specifically designed for low-concentration CO2 detection in respiratory gases. Initial clinical applications emerged in the 1950s, with increasing adoption in anesthesia during the 1970s, where capnography was used to monitor end-tidal CO2 during mechanical ventilation, marking the transition from laboratory tools to bedside use. Key milestones in the 1980s and beyond drove widespread integration into practice. The American Society of Anesthesiologists (ASA) mandated capnography as a standard for confirming endotracheal tube placement and monitoring ventilation during general anesthesia in 1986, significantly enhancing patient safety in operating rooms. By the late 1980s, capnography was routinely combined with pulse oximetry in multi-parameter monitors, facilitating comprehensive respiratory assessment. Adoption extended to emergency medical services in the 1990s, where portable devices became standard for out-of-hospital intubation confirmation and resuscitation monitoring. The International Organization for Standardization (ISO) established capnography device requirements in 1993 with ISO 9918, promoting uniformity in accuracy and performance. Post-2000 developments refined sampling methods, with sidestream techniques—using remote sensors for versatile, low-flow aspiration—gaining prominence over mainstream approaches for non-intubated patients, exemplified by Oridion's Microstream technology introduced in 1997 and optimized in subsequent years for enhanced reliability in diverse clinical environments.

Technology

Instrumentation and Devices

Capnography instrumentation primarily consists of two main types: sidestream and mainstream devices. Sidestream capnographs aspirate a small continuous sample of respiratory gases (typically 50-250 mL/min) through a sampling line to a remote analyzer, allowing for non-invasive monitoring in non-intubated patients via nasal cannulas or masks. In contrast, mainstream capnographs employ an in-line sensor positioned directly at the airway, such as on the endotracheal tube connector, providing immediate measurement without gas diversion. Devices are also categorized as portable or stationary; portable units, often sidestream-based, are compact and battery-powered for use in emergency medical services (EMS) or transport, while stationary systems integrate into operating room or ICU monitors for fixed clinical environments. Key components include CO2 sensors, displays, and patient interfaces. The primary sensor technology is infrared spectroscopy, which detects CO2 concentration by measuring absorption of infrared light at approximately 4.26 μm wavelength, though mass spectrometry—ionizing and separating gas molecules by mass-to-charge ratio—serves as an alternative in some multi-gas systems. Displays typically show numerical end-tidal CO2 (EtCO2) values in mmHg or kPa alongside real-time waveforms on integrated screens. Adapters facilitate connections, such as nasal cannulas with integrated sampling lines for spontaneous breathing patients or specialized endotracheal tube connectors with T-pieces for intubated cases, ensuring secure and low-dead-space sampling. Setup involves connecting the device to patient breathing circuits: for sidestream, attaching the sampling line to the airway adapter and ensuring the analyzer is positioned away from the patient with proper scavenging of aspirated gases; for mainstream, securing the sensor cuvette directly in the circuit. Calibration includes a zeroing procedure by exposing the sensor to room air (0% CO2) for 15-20 seconds to establish baseline, followed by optional span calibration using a known CO2 concentration gas mixture. Response times differ by type, with mainstream offering near real-time detection (rise time <100 ms) and sidestream introducing a transport delay of 100-500 ms plus rise time, depending on tubing length and flow rate. Modern capnography devices incorporate advanced features for enhanced utility. Many integrate seamlessly with mechanical ventilators, automatically syncing EtCO2 data to ventilation parameters. Wireless portability is common in EMS-oriented models, enabling tetherless monitoring via Bluetooth connectivity to central stations. Additionally, multi-gas analyzers extend beyond CO2 to measure oxygen, nitrous oxide, and volatile anesthetics using combined infrared and paramagnetic sensors, supporting comprehensive perioperative gas analysis.

Measurement Principles

Capnography primarily relies on infrared (IR) spectroscopy to detect and quantify carbon dioxide (CO₂) in respiratory gases, as CO₂ molecules exhibit strong absorption of IR light at a wavelength of 4.26 μm. In this non-dispersive IR method, a light source emits IR radiation through a sample chamber containing the respiratory gas, and a photodetector measures the transmitted light intensity. The degree of absorption is directly proportional to the CO₂ concentration, enabling real-time monitoring. This technique is favored for its specificity, as the 4.26 μm band is relatively isolated from absorption by other common gases like oxygen or nitrogen. The quantitative relationship in IR spectroscopy follows the Beer-Lambert law, which describes how light absorption depends on the properties of the absorbing material: I=I0eϵclI = I_0 e^{-\epsilon c l} Here, II is the intensity of transmitted light, I0I_0 is the incident light intensity, ϵ\epsilon is the molar absorptivity coefficient specific to CO₂ at 4.26 μm, cc is the CO₂ concentration, and ll is the optical path length through the sample. Modern capnographs apply this law computationally to convert measured absorption into CO₂ levels, often displayed as waveforms over time. Calibration against known CO₂ concentrations ensures accuracy, typically within ±2 mmHg or ±0.2% for clinical devices. Alternative detection methods exist but are less prevalent in routine capnography. Mass spectrometry ionizes gas molecules in a vacuum and separates them based on their mass-to-charge ratio, allowing precise identification and quantification of CO₂ alongside other gases; however, it requires complex, expensive equipment often shared across multiple operating rooms. Electrochemical sensors, which generate electrical signals via CO₂-induced chemical reactions at an electrode, are infrequently used due to baseline drift from electrolyte degradation and sensitivity to environmental changes, limiting their reliability for continuous monitoring. CO₂ measurements in capnography are expressed either as partial pressure (EtPCO₂ in mmHg) or volumetric concentration (EtCO₂ in %), with conversion necessary for clinical interpretation. The relationship accounts for total atmospheric pressure and water vapor saturation at body temperature (47 mmHg): EtPCO₂ (mmHg) = EtCO₂ (%) × (P_atm - 47) / 100, where P_atm is the barometric pressure (typically 760 mmHg at sea level). This adjustment ensures equivalence, as partial pressure reflects the effective tension driving gas exchange, while concentration is directly measured by sensors. Measurement accuracy can be compromised by several factors, necessitating built-in compensations. Temperature variations alter gas density and IR absorption coefficients, requiring sensors to incorporate thermistors for real-time correction to standardize readings at 37°C. Humidity introduces interference through water vapor absorption near the CO₂ band or condensation in sampling lines, often mitigated by hydrophobic filters or drying chambers in sidestream systems. Sensor drift, a progressive calibration shift over hours to days due to optical component aging or contamination, demands periodic zeroing and calibration with reference gases to maintain precision within clinical tolerances of ±0.2 kPa.

Waveform Interpretation

Normal Capnogram Characteristics

The normal capnogram in healthy individuals displays a rectangular waveform that graphically represents carbon dioxide (CO₂) concentrations during the respiratory cycle, providing insights into ventilation and gas exchange efficiency. This waveform is characterized by distinct phases corresponding to inspiration and expiration, with a baseline near zero during inspiration and a rise to end-tidal CO₂ (EtCO₂) levels during exhalation. The overall shape is nearly square, oscillating at the respiratory rate, and reflects the sequential elimination of anatomical dead space gas followed by alveolar gas. The capnogram is divided into five phases. Phase 0 denotes the inspiratory baseline, where inspired gas devoid of CO₂ maintains a flat line at approximately 0 mmHg. Phase I follows as the initial portion of exhalation, expelling dead space gas from the airways with no CO₂, resulting in a continued flat segment at baseline. Phase II represents the rapid, exponential upstroke as alveolar gas mixes with residual dead space gas, causing a steep rise in CO₂ concentration. Phase III forms the alveolar plateau, a relatively flat segment where CO₂ levels stabilize as pure alveolar gas is exhaled, reaching the EtCO₂ peak of 35-45 mmHg. Phase 0' marks the abrupt return to baseline at the onset of the next inspiration. Key parameters of the normal capnogram include the EtCO₂ value, which quantifies the maximum CO₂ at the end of expiration; the respiratory rate, typically 12-20 breaths per minute in adults; the slope of Phase II, which is steep and exponential, often with an alpha angle (transition to Phase III) approaching 100-110° for a sharp rise; and the consistency of Phase III plateau height across cycles, indicating stable alveolar ventilation. These features ensure the waveform's reliability as a monitor of normal pulmonary function. Quantitative norms for a healthy capnogram encompass an EtCO₂ of 35-45 mmHg, reflecting arterial CO₂ partial pressure under normal conditions; a Pa-EtCO₂ gradient of 2-5 mmHg, accounting for minor alveolar dead space; and inspired CO₂ below 2 mmHg, confirming no rebreathing. Normal variations include a slight upward slope in Phase III during spontaneous breathing, attributable to continued CO₂ diffusion from blood to alveoli throughout expiration, contrasting with the flatter plateau in controlled mechanical ventilation. EtCO₂ values also decrease at high altitudes due to lower barometric pressure and hyperventilation responses, potentially dropping to around 25 mmHg above 3,000 meters.

Abnormal Waveform Patterns

Abnormal capnography waveforms deviate from the standard phases observed in a normal capnogram, where Phase 0 represents inspiration at baseline, Phase I the dead-space gas exhalation, Phase II the rapid upstroke of mixed gas, Phase III the alveolar plateau with end-tidal CO₂ (EtCO₂) at 35–45 mmHg, and Phase 0' the return to baseline. These deviations reflect underlying pathophysiological processes affecting ventilation, perfusion, or equipment integrity. Hypoventilation manifests as elevated EtCO₂ levels exceeding 45 mmHg, often with a prolonged Phase III due to delayed CO₂ elimination from inadequate alveolar ventilation. In cases of airway obstruction, the waveform assumes a "shark-fin" appearance, characterized by a slanted, upward-curving Phase III as expiratory flow resistance prolongs the rise in CO₂ concentration. This pattern arises from reduced tidal volume or respiratory rate, leading to CO₂ accumulation in the lungs. Conversely, hyperventilation produces low EtCO₂ values below 35 mmHg, with a steep Phase II upstroke and a reduced or absent alveolar plateau (Phase III) due to excessive CO₂ washout. The waveform narrows as rapid, deep breaths accelerate alveolar emptying, minimizing the time for CO₂ equilibration. This reflects increased minute ventilation relative to CO₂ production. Rebreathing is indicated by an inspired CO₂ level above 2 mmHg, causing a gradual rise in the baseline that does not return to zero during inspiration. This occurs when CO₂ is not fully cleared from the breathing circuit, such as from an exhausted absorber, resulting in progressive EtCO₂ elevation and a "crescendo" waveform. The pattern signals inefficient gas exchange, increasing the risk of hypercapnia. Esophageal intubation leads to an abrupt loss of the waveform after possibly a few initial detections, flattening the trace as no pulmonary CO₂ is delivered to the detector. This sudden absence distinguishes it from gradual declines, stemming from misplacement of the airway device outside the trachea. Bronchospasm produces a "scooped" or shark-fin shaped Phase III, with a gradual, sloped rise instead of a flat plateau, due to uneven alveolar emptying from airway constriction. The increased expiratory resistance delays CO₂ exhalation from affected lung regions, altering the waveform's contour. Low cardiac output results in a gradual decrease in EtCO₂ over successive breaths, reflecting diminished pulmonary perfusion and reduced CO₂ transport to the alveoli. The waveform amplitude diminishes without shape distortion, as systemic hypoperfusion limits CO₂ delivery. Artifacts include small oscillations superimposed on Phase III from cardiac pulsations, caused by heartbeat-induced pressure changes in the pulmonary vasculature, and intermittent baseline shifts from loose connections disrupting gas flow. These do not alter EtCO₂ but introduce noise in the trace. Diagnostic thresholds for abnormalities include EtCO₂ below 10 mmHg indicating severe circulatory compromise such as cardiac arrest. In obstructive processes like bronchospasm, the Phase II upstroke slope decreases (shallower angle), contributing to the shark-fin shape. These metrics provide quantitative cues for waveform analysis.

Clinical Applications

Anesthesia and Perioperative Monitoring

Capnography plays a central role in anesthesia by confirming endotracheal intubation through either colorimetric devices, which detect exhaled CO2 via color change, or quantitative waveform capnography, which verifies tracheal placement by displaying persistent CO2 waveforms within six breaths, distinguishing it from brief esophageal intubation signals. During controlled mechanical ventilation under general anesthesia, continuous capnography monitors end-tidal CO2 (EtCO2) levels to ensure adequate alveolar ventilation and detect hypoventilation early, while sudden drops to zero EtCO2 indicate circuit disconnections or leaks, allowing prompt intervention to prevent hypoxia. In procedures involving anesthetics, capnography tracks CO2 dynamics during laparoscopy, where CO2 insufflation increases absorption and risks hypercapnia; vigilant monitoring guides minute ventilation adjustments to maintain normocapnia and avoid acidosis or hemodynamic instability. Opioids administered perioperatively can induce hypoventilation, and capnography provides real-time detection of rising EtCO2 or waveform prolongation, enabling timely reversal to mitigate respiratory depression. Perioperative adjustments include baseline EtCO2 recalibration for patient positioning; for instance, the Trendelenburg position can elevate EtCO2 due to reduced venous return and ventilation-perfusion mismatch, necessitating increased ventilator rates. During post-operative emergence, capnography monitors for residual effects like hypoventilation, supporting safe extubation by confirming return to spontaneous breathing with stable EtCO2 waveforms. The American Society of Anesthesiologists (ASA) has mandated continuous capnography for all patients under general anesthesia since its 1986 Standards for Basic Anesthetic Monitoring, emphasizing its role in verifying ventilation efficacy. Under general anesthesia, target EtCO2 is typically maintained at 30-40 mmHg to balance ventilation and avoid extremes of hypocapnia or hypercapnia.

Emergency Medicine and Resuscitation

In emergency medicine, capnography serves as the gold standard for confirming endotracheal tube placement, particularly in pre-hospital and emergency department settings where rapid airway management is critical. Detection of end-tidal CO₂ (EtCO₂) greater than 10 mmHg reliably rules out esophageal intubation, as esophageal placement typically yields no sustained CO₂ waveform due to the absence of pulmonary gas exchange. This method outperforms clinical auscultation or chest rise visualization alone, reducing misplacement errors that can compromise ventilation during resuscitation. Capnography is also applied to supraglottic airways, such as laryngeal masks, where persistent EtCO₂ waveforms confirm effective gas exchange and guide adjustments for optimal seal and positioning. During cardiopulmonary resuscitation (CPR), quantitative waveform capnography monitors EtCO₂ as a surrogate for pulmonary blood flow, reflecting the efficacy of chest compressions in delivering oxygenated blood to the lungs. Effective compressions maintain EtCO₂ levels above 10-20 mmHg, with values below 10 mmHg signaling inadequate perfusion and prompting technique optimization, such as deeper or faster compressions. A sudden rise in EtCO₂ exceeding 40 mmHg often indicates return of spontaneous circulation (ROSC), providing an early, non-invasive cue for pulse checks even before palpable signs. Persistent EtCO₂ below 10 mmHg after 20 minutes of CPR predicts poor outcomes with high sensitivity, guiding decisions to continue or terminate efforts. In trauma scenarios and patient transport, capnography aids in detecting life-threatening conditions like tension pneumothorax and hypovolemia through characteristic decreases in EtCO₂. For tension pneumothorax, progressive waveform flattening and EtCO₂ dropping below 35 mmHg result from impaired venous return and reduced cardiac output, with rapid EtCO₂ elevation post-decompression confirming intervention success. Hypovolemia similarly presents with low EtCO₂ waveforms due to diminished pulmonary perfusion, often below 25 mmHg, correlating with hemodynamic instability and higher mortality risk in pre-hospital trauma patients. These applications extend to ambulance transfers, where portable capnography devices enable continuous monitoring to detect deteriorations en route. The American Heart Association (AHA) integrated capnography into Advanced Cardiovascular Life Support (ACLS) protocols starting in 2010, recommending its use for ongoing airway verification and CPR quality assessment in intubated patients. This was reinforced in subsequent guidelines, emphasizing quantitative waveform capnography over colorimetric devices for real-time feedback. In emergency medical services (EMS), compact, battery-powered devices like the EMMA capnograph facilitate seamless integration into transport protocols, providing rugged, sidestream sampling for EtCO₂ trends without interrupting care.

Critical Care and Mechanical Ventilation

In intensive care units (ICUs), capnography serves as a vital tool for optimizing mechanical ventilation by providing real-time feedback on end-tidal CO₂ (EtCO₂) levels, enabling clinicians to adjust parameters such as tidal volume and fraction of inspired oxygen (FiO₂) to prevent complications like auto-positive end-expiratory pressure (auto-PEEP) or barotrauma. For instance, targeting an EtCO₂ of approximately 30 mmHg during initial post-intubation ventilation often aligns with a normal arterial partial pressure of CO₂ (PaCO₂) range of 35-45 mmHg, allowing for safe titration of ventilator settings while minimizing lung injury. This approach facilitates early detection of ventilation-perfusion mismatches through continuous waveform analysis, promoting lung-protective strategies in critically ill patients. Volumetric capnography enhances ventilation optimization by enabling breath-by-breath calculation of dead space ventilation, quantified as the dead space-to-tidal volume ratio (Vd/Vt) using the Bohr-Enghoff equation: Vd/Vt = (PaCO₂ - PECO₂)/PaCO₂, where PECO₂ represents mixed expired CO₂ partial pressure. This metric helps assess alveolar dead space and guide adjustments to reduce ineffective ventilation, particularly in patients with heterogeneous lung pathology, thereby improving CO₂ elimination efficiency without relying solely on arterial blood gases (ABGs). During weaning from mechanical ventilation and sedation management, capnography monitors patient readiness by evaluating waveform stability and detecting signs of ventilator dyssynchrony, such as clefts or notching in Phase III of the capnogram, which indicate asynchronous breathing efforts. A stable EtCO₂ waveform with consistent plateau phases signals adequate respiratory drive and CO₂ clearance, serving as a noninvasive adjunct to traditional weaning criteria and reducing the need for frequent ABGs. In cases of suspected hypercapnia during spontaneous breathing trials, capnography can prompt timely interventions, though it typically underestimates PaCO₂ (with a gradient of 2-5 mmHg), necessitating confirmatory blood gas analysis when discrepancies are suspected. In specific ICU scenarios like acute respiratory distress syndrome (ARDS), capnography supports permissive hypercapnia strategies by tracking EtCO₂ trends to tolerate elevated PaCO₂ levels (typically 45-55 mmHg or higher) while limiting tidal volumes to 4-6 mL/kg predicted body weight, with elevated Vd/Vt ratios (>0.60) prognostic of poorer outcomes. For , fluctuating EtCO₂ values often reflect increased intrapulmonary shunting and ventilation-perfusion imbalances, prompting adjustments in positive end-expiratory pressure (PEEP) to optimize oxygenation and CO₂ removal. Capnography integrates effectively with ABG analysis to monitor the PaCO₂-EtCO₂ gradient, where differences exceeding 5-10 mmHg signal increased or shunting, guiding fine-tuning of support. The American Association for Respiratory Care (AARC) recommends routine capnography use in mechanically ventilated patients for ongoing assessment (grade 2B ), emphasizing its role in enhancing and in ICU management.

Diagnostic Applications

Respiratory and Airway Assessment

Capnography serves as a vital diagnostic tool for assessing respiratory conditions, particularly in evaluating airway patency and lung function through analysis of end-tidal CO₂ (EtCO₂) levels and waveform morphology. By measuring the of CO₂ in exhaled breath, it provides non-invasive insights into ventilation-perfusion matching, dead space ventilation, and expiratory dynamics, enabling early detection of abnormalities in patients with suspected pulmonary issues. In airway disorders, capnography identifies characteristic waveform changes that reflect obstructed or irregular airflow. Bronchospasm, often seen in acute exacerbations, produces a "shark-fin" with a progressively sloped Phase III due to heterogeneous alveolar emptying and delayed CO₂ from varying time constants in the airways. Aspiration events can manifest as a prolonged Phase II, where the transition from dead space to alveolar gas mixing is extended, indicating partial airway blockage or retained secretions. Upper airway obstruction, such as from or , typically results in an absent or undetectable alveolar plateau (Phase III), as is curtailed before full alveolar gas elimination. For lung parenchymal diseases, capnography highlights ventilation inefficiencies through quantitative and qualitative metrics. In (COPD) and , an elevated dead space to ratio (Vd/Vt >0.6) is common, reflecting increased physiologic dead space from alveolar destruction and , often accompanied by a sloped Phase III due to uneven ventilation distribution. presents with variable EtCO₂ readings, stemming from regional ventilation-perfusion mismatches caused by consolidated lung segments, leading to fluctuating CO₂ elimination during expiration. Quantitative diagnostics using capnography further aid in pinpointing specific pathologies. An EtCO₂ to PaCO₂ gradient greater than 5 mmHg signals increased alveolar , a hallmark of conditions like where reduced pulmonary perfusion impairs CO₂ exchange. Additionally, capnodynamic indices, derived from volumetric capnography, quantify changes in end-expiratory lung volume during recruitment maneuvers, helping clinicians evaluate alveolar recruitment and optimize in acute respiratory distress. In pediatric and neonatal populations, capnography requires age-specific interpretations due to physiological differences. Normal EtCO₂ values range from 35 to 45 mmHg in neonates, young children, and adults, owing to similar alveolar ventilation mechanics despite higher metabolic rates in younger patients. It is particularly useful for apnea monitoring in these vulnerable groups, as abrupt cessation of the capnogram waveform signals respiratory pauses, allowing prompt intervention to prevent .

Cardiovascular and Metabolic Evaluation

Capnography serves as a non-invasive indicator of circulatory status by reflecting pulmonary blood flow and , with end-tidal CO₂ (EtCO₂) levels typically decreasing in states of hypoperfusion. In patients experiencing shock or , low EtCO₂ values (e.g., below 30 mmHg) often signal inadequate tissue , as reduced limits CO₂ delivery to the lungs for . This threshold correlates with hemorrhagic shock stages, where EtCO₂ below 30 mmHg indicates moderate to severe , serving as an early marker during efforts. Additionally, waveform oscillations or a "curly" appearance in the expiratory phase can correlate with diminished variations in low-output states, highlighting inconsistencies due to or . In cardiac events, capnography distinguishes true (PEA) from pseudo-PEA by revealing a flat waveform despite electrocardiographic activity, indicating absent and confirming the need for aggressive intervention. For , a sudden drop in EtCO₂ occurs with preserved ventilation, resulting from increased dead space ventilation and mismatched ventilation- ratios that impair CO₂ elimination. This abrupt decline prompts immediate diagnostic , as it reflects obstructed pulmonary blood flow without primary respiratory compromise. Capnography also links to metabolic derangements, where compensatory mechanisms alter CO₂ dynamics. In , driven by typically lowers EtCO₂ below 25 mmHg, providing a rapid bedside screen for severity; values above 30 mmHg effectively rule out the with high sensitivity. reduces CO₂ production and slows alveolar emptying, manifesting as a prolonged or delayed rise in Phase II of the waveform, with overall EtCO₂ depression reflecting metabolic slowdown. Advanced applications of volumetric capnography enable estimation through the , where approximates CO₂ elimination rate divided by the arteriovenous CO₂ content difference (VCO₂ / (CvCO₂ - CaCO₂)), with PaCO₂ and PvCO₂ gradients informing mixed venous-arterial differences non-invasively. In , declining EtCO₂ trends inversely correlate with rising lactate levels ( -0.42), predicting mortality and guiding fluid by indicating tissue hypoperfusion before overt hemodynamic instability. This relationship underscores capnography's role in monitoring metabolic distress, where persistent low EtCO₂ signals ongoing anaerobic metabolism and poor prognosis.

Limitations and Considerations

Sources of Error and Artifacts

Capnography measurements can be distorted by various technical, patient-related, and environmental factors, leading to inaccurate end-tidal CO2 (EtCO2) readings and unreliable waveforms. These errors may result in erratic patterns, falsely elevated or reduced values, or complete signal loss, potentially compromising clinical decision-making. Understanding these sources is essential for ensuring reliable monitoring. Technical errors primarily arise from device-related issues. malfunction, including drift, can cause gradual inaccuracies in CO2 detection over time; for instance, in infrared-based systems, uncorrected drift may lead to deviations exceeding acceptable limits during prolonged use. Sampling line occlusion, often due to , moisture, or secretions, produces erratic or absent waveforms by impeding gas flow to the analyzer. failure exacerbates these problems, as improper zeroing to room air or use of known CO2 concentrations results in baseline shifts and overall . Patient factors contribute to dilution or incomplete sampling of exhaled CO2. In non-intubated patients, dilutes EtCO2 by entraining ambient air, leading to lower readings compared to nasal-only . Leaks around masks or airway interfaces cause significant underestimation of EtCO2, as exhaled gas escapes without reaching the sensor, potentially altering ventilation assessments. Environmental influences further complicate accuracy. High promotes in sampling lines, particularly in sidestream systems, clogging the tubing and triggering occlusion alarms or loss. At high altitudes, reduced barometric pressure affects gas analyzer performance, potentially causing EtCO2 underestimation due to altered dynamics. interference in capnographs produces falsely elevated CO2 readings stemming from overlapping absorption spectra. Mitigation strategies focus on preventive maintenance and verification. Regular calibration and line inspections minimize technical errors, while using water traps or dehumidifying materials like tubing prevents condensation buildup. Alternative sampling sites, such as sealed nasal-oral interfaces, reduce dilution from . Cross-verification with arterial blood gas analysis confirms EtCO2 accuracy when artifacts are suspected.

Clinical Guidelines and Future Directions

Clinical guidelines emphasize the integration of capnography as a standard tool for confirming across various healthcare settings. The (AHA) provides a Class I recommendation for using waveform capnography to verify and continuously monitor correct endotracheal tube placement during and , citing its 100% specificity in clinical trials and observational studies (as of 2025). Similarly, the AHA endorses persistent capnographic waveforms with ventilation as the preferred method for confirmation in all environments, including departments and intensive care units. In nursing practice, the American Association of Critical-Care Nurses (AACN) protocols for noninvasive monitoring advocate routine waveform capnography assessments to detect changes in respiratory status early. Capnography's multidisciplinary application spans anesthesiologists, paramedics, and nurses, promoting interprofessional protocols to enhance during . Anesthesiologists routinely employ capnography for procedural and verification, as outlined in society guidelines requiring its use whenever verbal responsiveness is lost. Paramedics follow national EMS clinical guidelines that mandate continuous waveform capnography for end-tidal CO2 monitoring in advanced airway interventions, such as during out-of-hospital intubations. Nurses, particularly in critical care, integrate capnography into routine assessments under AACN standards, fostering collaborative decision-making with physicians to interpret waveforms and adjust ventilation strategies. Emerging technologies are poised to expand capnography's utility beyond traditional settings. Artificial intelligence (AI) algorithms for automated waveform analysis have demonstrated over 90% accuracy in pattern recognition for detecting abnormalities, such as in combined capnography and pulse oximetry data, enabling real-time classification of normal versus abnormal respiratory events in clinical trials. Wearable capnographs facilitate ambulatory monitoring by providing continuous EtCO2 and respiratory rate data comparable to standard capnography, with pilot studies showing feasibility in non-intubated patients for early detection of deterioration outside acute care environments. Post-2020, integration with telemedicine has advanced through remote patient monitoring systems that incorporate capnography for chronic respiratory conditions, allowing real-time data transmission to clinicians via low-cost breath analyzers in home settings. The 2025 AHA guidelines reaffirm capnography's role in resuscitation, including integration with other data for arrest management. Key research gaps persist, particularly regarding long-term outcomes in (NIV), where capnography's role in optimizing therapy adherence and predicting survival in conditions like remains underexplored despite evidence of improved management. Advancements since 2015, such as refined Microstream technology with improved sampling lines for molecular correlation spectroscopy, have enhanced accuracy in low-flow and neonatal applications but require further validation in diverse populations to address these gaps.