Hubbry Logo
Cardiac outputCardiac outputMain
Open search
Cardiac output
Community hub
Cardiac output
logo
8 pages, 0 posts
0 subscribers
Be the first to start a discussion here.
Be the first to start a discussion here.
Contribute something
Cardiac output
Cardiac output
from Wikipedia

Major factors influencing cardiac output – heart rate and stroke volume, both of which are variable.[1]

In cardiac physiology, cardiac output (CO), also known as heart output and often denoted by the symbols , , or ,[2] is the volumetric flow rate of the heart's pumping output: that is, the volume of blood being pumped by a single ventricle of the heart, per unit time (usually measured per minute). Cardiac output (CO) is the product of the heart rate (HR), i.e. the number of heartbeats per minute (bpm), and the stroke volume (SV), which is the volume of blood pumped from the left ventricle per beat; thus giving the formula:

[3]

Values for cardiac output are usually denoted as L/min. For a healthy individual weighing 70 kg, the cardiac output at rest averages about 5 L/min; assuming a heart rate of 70 beats/min, the stroke volume would be approximately 70 mL.

Because cardiac output is related to the quantity of blood delivered to various parts of the body, it is an important component of how efficiently the heart can meet the body's demands for the maintenance of adequate tissue perfusion. Body tissues require continuous oxygen delivery which requires the sustained transport of oxygen to the tissues by systemic circulation of oxygenated blood at an adequate pressure from the left ventricle of the heart via the aorta and arteries. Oxygen delivery (DO2 mL/min) is the resultant of blood flow (cardiac output CO) times the blood oxygen content (CaO2). Mathematically this is calculated as follows: oxygen delivery = cardiac output × arterial oxygen content, giving the formula:

[4]

With a resting cardiac output of 5 L/min, a 'normal' oxygen delivery is around 1 L/min. The amount/percentage of the circulated oxygen consumed (VO2) per minute through metabolism varies depending on the activity level but at rest is circa 25% of the DO2. Physical exercise requires a higher than resting-level of oxygen consumption to support increased muscle activity. Regular aerobic exercise can induce physiological adaptations such as improved stroke volume and myocardial efficiency that increase cardiac output.[5] In the case of heart failure, actual CO may be insufficient to support even simple activities of daily living; nor can it increase sufficiently to meet the higher metabolic demands stemming from even moderate exercise.

Cardiac output is a global blood flow parameter of interest in hemodynamics, the study of the flow of blood. The factors affecting stroke volume and heart rate also affect cardiac output. The figure at the right margin illustrates this dependency and lists some of these factors. A detailed hierarchical illustration is provided in a subsequent figure.

There are many methods of measuring CO, both invasively and non-invasively; each has advantages and drawbacks as described below.

Trend of central venous pressure as a consequence of variations in cardiac output. The three functions indicate the trend in physiological conditions (in the centre), in those of decreased preload (e.g. in hemorrhage, bottom curve) and in those of increased preload (e.g. following transfusion, top curve).

Definition

[edit]

The function of the heart is to drive blood through the circulatory system in a cycle that delivers oxygen, nutrients and chemicals to the body's cells and removes cellular waste. Because it pumps out whatever blood comes back into it from the venous system, the quantity of blood returning to the heart effectively determines the quantity of blood the heart pumps out – its cardiac output, Q. Cardiac output is classically defined alongside stroke volume (SV) and the heart rate (HR) as:[citation needed]

In standardizing what CO values are considered to be within normal range independent of the size of the subject's body, the accepted convention is to further index equation (1) using body surface area (BSA), giving rise to the Cardiac index (CI). This is detailed in equation (2) below.

Measurement

[edit]

There are a number of clinical methods to measure cardiac output, ranging from direct intracardiac catheterization to non-invasive measurement of the arterial pulse. Each method has advantages and drawbacks. Relative comparison is limited by the absence of a widely accepted "gold standard" measurement. Cardiac output can also be affected significantly by the phase of respiration – intra-thoracic pressure changes influence diastolic filling and therefore cardiac output. This is especially important during mechanical ventilation, in which cardiac output can vary by up to 50% across a single respiratory cycle.[citation needed] Cardiac output should therefore be measured at evenly spaced points over a single cycle or averaged over several cycles.[citation needed]

Invasive methods are well accepted, but there is increasing evidence that these methods are neither accurate nor effective in guiding therapy. Consequently, the focus on development of non-invasive methods is growing.[6][7][8]

Doppler ultrasound

[edit]
Doppler signal in the left ventricular outflow tract: Velocity Time Integral (VTI)

This method uses ultrasound and the Doppler effect to measure cardiac output. The blood velocity through the heart causes a Doppler shift in the frequency of the returning ultrasound waves. This shift can then be used to calculate flow velocity and volume, and effectively cardiac output, using the following equations:[citation needed]

where:

  • CSA is the valve orifice cross sectional area,
  • r is the valve radius, and,
  • VTI is the velocity time integral of the trace of the Doppler flow profile.

Being non-invasive, accurate and inexpensive, Doppler ultrasound is a routine part of clinical ultrasound; it has high levels of reliability and reproducibility, and has been in clinical use since the 1960s.[citation needed]

Echocardiography

[edit]

Echocardiography is a non-invasive method of quantifying cardiac output using ultrasound. Two-dimensional (2D) ultrasound and Doppler measurements are used together to calculate cardiac output. 2D measurement of the diameter (d) of the aortic annulus allows calculation of the flow cross-sectional area (CSA), which is then multiplied by the VTI of the Doppler flow profile across the aortic valve to determine the flow volume per beat (stroke volume, SV). The result is then multiplied by the heart rate (HR) to obtain cardiac output. Although used in clinical medicine, it has a wide test-retest variability.[9] It is said to require extensive training and skill, but the exact steps needed to achieve clinically adequate precision have never been disclosed. 2D measurement of the aortic valve diameter is one source of noise; others are beat-to-beat variation in stroke volume and subtle differences in probe position. An alternative that is not necessarily more reproducible is the measurement of the pulmonary valve to calculate right-sided CO. Although it is in wide general use, the technique is time-consuming and is limited by the reproducibility of its component elements. In the manner used in clinical practice, precision of SV and CO is of the order of ±20%.[citation needed]

Transcutaneous

[edit]

Ultrasonic Cardiac Output Monitor (USCOM) uses continuous wave Doppler to measure the Doppler flow profile VTI. It uses anthropometry to calculate aortic and pulmonary valve diameters and CSAs, allowing right-sided and left-sided Q measurements. In comparison to the echocardiographic method, USCOM significantly improves reproducibility and increases sensitivity of the detection of changes in flow. Real-time, automatic tracing of the Doppler flow profile allows beat-to-beat right-sided and left-sided Q measurements, simplifying operation and reducing the time of acquisition compared to conventional echocardiography. USCOM has been validated from 0.12 L/min to 18.7 L/min[10] in new-born babies,[11] children[12] and adults.[13] The method can be applied with equal accuracy to patients of all ages for the development of physiologically rational haemodynamic protocols. USCOM is the only method of cardiac output measurement to have achieved equivalent accuracy to the implantable flow probe.[14] This accuracy has ensured high levels of clinical use in conditions including sepsis, heart failure and hypertension.[15][16][17]

Transoesophageal

[edit]
A Transesophageal echocardiogram (BrE: TOE, AmE: TEE) probe.
A transoesophageal echocardiogram probe.

The Transoesophageal Doppler includes two main technologies; transoesophageal echocardiogram—which is primarily used for diagnostic purposes, and oesophageal Doppler monitoring—which is primarily used for the clinical monitoring of cardiac output. The latter uses continuous wave Doppler to measure blood velocity in the descending thoracic aorta. An ultrasound probe is inserted either orally or nasally into the oesophagus to mid-thoracic level, at which point the oesophagus lies alongside the descending thoracic aorta. Because the transducer is close to the blood flow, the signal is clear. The probe may require re-focussing to ensure an optimal signal. This method has good validation, is widely used for fluid management during surgery with evidence for improved patient outcome,[18][19][20][21][22][23][24][25] and has been recommended by the UK's National Institute for Health and Clinical Excellence (NICE).[26] Oesophageal Doppler monitoring measures the velocity of blood and not true Q, therefore relies on a nomogram[27] based on patient age, height and weight to convert the measured velocity into stroke volume and cardiac output. This method generally requires patient sedation and is accepted for use in both adults and children.[citation needed]

Pulse pressure methods

[edit]

Pulse pressure (PP) methods measure the pressure in an artery over time to derive a waveform and use this information to calculate cardiac performance. However, any measure from the artery includes changes in pressure associated with changes in arterial function, for example compliance and impedance. Physiological or therapeutic changes in vessel diameter are assumed to reflect changes in Q. PP methods measure the combined performance of the heart and the blood vessels, thus limiting their application for measurement of Q. This can be partially compensated for by intermittent calibration of the waveform to another Q measurement method then monitoring the PP waveform. Ideally, the PP waveform should be calibrated on a beat-to-beat basis. There are invasive and non-invasive methods of measuring PP.[citation needed]

Finapres methodology

[edit]

In 1967, the Czech physiologist Jan Peňáz invented and patented the volume clamp method of measuring continuous blood pressure. The principle of the volume clamp method is to dynamically provide equal pressures, on either side of an artery wall. By clamping the artery to a certain volume, inside pressure—intra-arterial pressure—balances outside pressure—finger cuff pressure. Peñáz decided the finger was the optimal site to apply this volume clamp method. The use of finger cuffs excludes the device from application in patients without vasoconstriction, such as in sepsis or in patients on vasopressors.[citation needed]

In 1978, scientists at BMI-TNO, the research unit of Netherlands Organisation for Applied Scientific Research at the University of Amsterdam, invented and patented a series of additional key elements that make the volume clamp work in clinical practice. These methods include the use of modulated infrared light in the optical system inside the sensor, the lightweight, easy-to-wrap finger cuff with velcro fixation, a new pneumatic proportional control valve principle, and a set point strategy for the determining and tracking the correct volume at which to clamp the finger arteries—the Physiocal system. An acronym for physiological calibration of the finger arteries, this Physiocal tracker was found to be accurate, robust and reliable.[citation needed]

The Finapres methodology was developed to use this information to calculate arterial pressure from finger cuff pressure data. A generalised algorithm to correct for the pressure level difference between the finger and brachial sites in patients was developed. This correction worked under all of the circumstances it was tested in—even when it was not designed for it—because it applied general physiological principles. This innovative brachial pressure waveform reconstruction method was first implemented in the Finometer, the successor of Finapres that BMI-TNO introduced to the market in 2000.[citation needed]

The availability of a continuous, high-fidelity, calibrated blood pressure waveform opened up the perspective of beat-to-beat computation of integrated haemodynamics, based on two notions: pressure and flow are inter-related at each site in the arterial system by their so-called characteristic impedance. At the proximal aortic site, the 3-element Windkessel model of this impedance can be modelled with sufficient accuracy in an individual patient with known age, gender, height and weight. According to comparisons of non-invasive peripheral vascular monitors, modest clinical utility is restricted to patients with normal and invariant circulation.[28]

Invasive

[edit]

Invasive PP monitoring involves inserting a manometer pressure sensor into an artery—usually the radial or femoral artery—and continuously measuring the PP waveform. This is generally done by connecting the catheter to a signal processing device with a display. The PP waveform can then be analysed to provide measurements of cardiovascular performance. Changes in vascular function, the position of the catheter tip or damping of the pressure waveform signal will affect the accuracy of the readings. Invasive PP measurements can be calibrated or uncalibrated.[citation needed]

Calibrated PP – PiCCO, LiDCO
[edit]

PiCCO (PULSION Medical Systems AG, Munich, Germany) and PulseCO (LiDCO Ltd, London, England) generate continuous Q by analysing the arterial PP waveform. In both cases, an independent technique is required to provide calibration of continuous Q analysis because arterial PP analysis cannot account for unmeasured variables such as the changing compliance of the vascular bed. Recalibration is recommended after changes in patient position, therapy or condition.[citation needed]

In PiCCO, transpulmonary thermodilution, which uses the Stewart-Hamilton principle but measures temperatures changes from central venous line to a central arterial line, i.e., the femoral or axillary arterial line, is used as the calibrating technique. The Q value derived from cold-saline thermodilution is used to calibrate the arterial PP contour, which can then provide continuous Q monitoring. The PiCCO algorithm is dependent on blood pressure waveform morphology (mathematical analysis of the PP waveform), and it calculates continuous Q as described by Wesseling and colleagues.[29] Transpulmonary thermodilution spans right heart, pulmonary circulation and left heart, allowing further mathematical analysis of the thermodilution curve and giving measurements of cardiac filling volumes (GEDV), intrathoracic blood volume and extravascular lung water. Transpulmonary thermodilution allows for less invasive Q calibration but is less accurate than PA thermodilution and requires a central venous and arterial line with the accompanied infection risks.[citation needed]

In LiDCO, the independent calibration technique is lithium chloride dilution using the Stewart-Hamilton principle. Lithium chloride dilution uses a peripheral vein and a peripheral arterial line. Like PiCCO, frequent calibration is recommended when there is a change in Q.[30] Calibration events are limited in frequency because they involve the injection of lithium chloride and can be subject to errors in the presence of certain muscle relaxants. The PulseCO algorithm used by LiDCO is based on pulse power derivation and is not dependent on waveform morphology.[citation needed]

Statistical analysis of arterial pressure – FloTrac/Vigileo
[edit]

FloTrac/Vigileo (Edwards Lifesciences) is an uncalibrated, haemodynamic monitor based on pulse contour analysis. It estimates cardiac output (Q) using a standard arterial catheter with a manometer located in the femoral or radial artery. The device consists of a high-fidelity pressure transducer, which, when used with a supporting monitor (Vigileo or EV1000 monitor), derives left-sided cardiac output (Q) from a sample of arterial pulsations. The device uses an algorithm based on the Frank–Starling law of the heart, which states pulse pressure (PP) is proportional to stroke volume (SV). The algorithm calculates the product of the standard deviation of the arterial pressure (AP) wave over a sampled period of 20 seconds and a vascular tone factor (Khi, or χ) to generate stroke volume. The equation in simplified form is: , or, . Khi is designed to reflect arterial resistance; compliance is a multivariate polynomial equation that continuously quantifies arterial compliance and vascular resistance. Khi does this by analyzing the morphological changes of arterial pressure waveforms on a bit-by-bit basis, based on the principle that changes in compliance or resistance affect the shape of the arterial pressure waveform. By analyzing the shape of said waveforms, the effect of vascular tone is assessed, allowing the calculation of SV. Q is then derived using equation (1). Only perfused beats that generate an arterial waveform are counted for in HR.[citation needed]

This system estimates Q using an existing arterial catheter with variable accuracy. These arterial monitors do not require intracardiac catheterisation from a pulmonary artery catheter. They require an arterial line and are therefore invasive. As with other arterial waveform systems, the short set-up and data acquisition times are benefits of this technology. Disadvantages include its inability to provide data regarding right-sided heart pressures or mixed venous oxygen saturation.[31][32] The measurement of Stroke Volume Variation (SVV), which predicts volume responsiveness is intrinsic to all arterial waveform technologies. It is used for managing fluid optimisation in high-risk surgical or critically ill patients. A physiologic optimization program based on haemodynamic principles that incorporates the data pairs SV and SVV has been published.[33]

Arterial monitoring systems are unable to predict changes in vascular tone; they estimate changes in vascular compliance. The measurement of pressure in the artery to calculate the flow in the heart is physiologically irrational and of questionable accuracy,[34] and of unproven benefit.[35] Arterial pressure monitoring is limited in patients off-ventilation, in atrial fibrillation, in patients on vasopressors, and in those with a dynamic autonomic system such as those with sepsis.[30]

Uncalibrated, pre-estimated demographic data-free – PRAM
[edit]

Pressure Recording Analytical Method (PRAM), estimates Q from the analysis of the pressure wave profile obtained from an arterial catheter—radial or femoral access. This PP waveform can then be used to determine Q. As the waveform is sampled at 1000 Hz, the detected pressure curve can be measured to calculate the actual beat-to-beat stroke volume. Unlike FloTrac, neither constant values of impedance from external calibration, nor form pre-estimated in vivo or in vitro data, are needed.[citation needed]

PRAM has been validated against the considered gold standard methods in stable condition[36] and in various haemodynamic states.[37] It can be used to monitor pediatric and mechanically supported patients.[38][39]

Generally monitored haemodynamic values, fluid responsiveness parameters and an exclusive reference are provided by PRAM: Cardiac Cycle Efficiency (CCE). It is expressed by a pure number ranging from 1 (best) to -1 (worst) and it indicates the overall heart-vascular response coupling. The ratio between heart performance and consumed energy, represented as CCE "stress index", can be of paramount importance in understanding the patient's present and future courses.[40]

Impedance cardiography

[edit]

Impedance cardiography (often abbreviated as ICG, or Thoracic Electrical Bioimpedance (TEB)) measures changes in electrical impedance across the thoracic region over the cardiac cycle. Lower impedance indicates greater intrathoracic fluid volume and blood flow. By synchronizing fluid volume changes with the heartbeat, the change in impedance can be used to calculate stroke volume, cardiac output and systemic vascular resistance.[41]

Both invasive and non-invasive approaches are used.[42] The reliability and validity of the non-invasive approach has gained some acceptance,[43][44][45][46] although there is not complete agreement on this point.[47] The clinical use of this approach in the diagnosis, prognosis and therapy of a variety of diseases continues.[48]

Non-invasive ICG equipment includes the Bio-Z Dx,[49] the Niccomo,[50] and TEBCO products by BoMed.[51][52]

Ultrasound dilution

[edit]

Ultrasound dilution (UD) uses body-temperature normal saline (NS) as an indicator introduced into an extracorporeal loop to create an atrioventricular (AV) circulation with an ultrasound sensor, which is used to measure the dilution then to calculate cardiac output using a proprietary algorithm. A number of other haemodynamic variables, such as total end-diastole volume (TEDV), central blood volume (CBV) and active circulation volume (ACVI) can be calculated using this method.[citation needed]

The UD method was firstly introduced in 1995.[53] It was extensively used to measure flow and volumes with extracorporeal circuit conditions, such as ECMO[54][55] and Haemodialysis,[56][57] leading more than 150 peer reviewed publications. UD has now been adapted to intensive care units (ICU) as the COstatus device.[58]

The UD method is based on ultrasound indicator dilution.[59] Blood ultrasound velocity (1560–1585 m/s) is a function of total blood protein concentration—sums of proteins in plasma and in red blood red cells—and temperature. Injection of body-temperature normal saline (ultrasound velocity of saline is 1533 m/s) into a unique AV loop decreases blood ultrasound velocity, and produces dilution curves.[citation needed]

UD requires the establishment of an extracorporeal circulation through its unique AV loop with two pre-existing arterial and central venous lines in ICU patients. When the saline indicator is injected into the AV loop, it is detected by the venous clamp-on sensor on the loop before it enters the patient's heart's right atrium. After the indicator traverses the heart and lung, the concentration curve in the arterial line is recorded and displayed on the COstatus HCM101 Monitor. Cardiac output is calculated from the area of the concentration curve using the Stewart-Hamilton equation. UD is a non-invasive procedure, requiring only a connection to the AV loop and two lines from a patient. UD has been specialised for application in pediatric ICU patients and has been demonstrated to be relatively safe although invasive and reproducible.[citation needed]

Electrical cardiometry

[edit]

Electrical cardiometry is a non-invasive method similar to Impedance cardiography; both methods measure thoracic electrical bioimpedance (TEB). The underlying model differs between the two methods; Electrical cardiometry attributes the steep increase of TEB beat-to-beat to the change in orientation of red blood cells. Four standard ECG electrodes are required for measurement of cardiac output. Electrical Cardiometry is a method trademarked by Cardiotronic, Inc., and shows promising results in a wide range of patients. It is currently approved in the US for use in adults, children and babies. Electrical cardiometry monitors have shown promise in postoperative cardiac surgical patients, in both haemodynamically stable and unstable cases.[60]

Magnetic resonance imaging

[edit]

Velocity-encoded phase contrast Magnetic resonance imaging (MRI)[61] is the most accurate technique for measuring flow in large vessels in mammals. MRI flow measurements have been shown to be highly accurate compared to measurements made with a beaker and timer,[62] and less variable than the Fick principle[63] and thermodilution.[64]

Velocity-encoded MRI is based on the detection of changes in the phase of proton precession. These changes are proportional to the velocity of the protons' movement through a magnetic field with a known gradient. When using velocity-encoded MRI, the result is two sets of images, one for each time point in the cardiac cycle. One is an anatomical image and the other is an image in which the signal intensity in each pixel is directly proportional to the through-plane velocity. The average velocity in a vessel, i.e., the aorta or the pulmonary artery, is quantified by measuring the average signal intensity of the pixels in the cross-section of the vessel then multiplying by a known constant. The flow is calculated by multiplying the mean velocity by the cross-sectional area of the vessel. This flow data can be used in a flow-versus-time graph. The area under the flow-versus-time curve for one cardiac cycle is the stroke volume. The length of the cardiac cycle is known and determines heart rate; Q can be calculated using equation (1). MRI is typically used to quantify the flow over one cardiac cycle as the average of several heart beats. It is also possible to quantify the stroke volume in real-time on a beat-for-beat basis.[65]

While MRI is an important research tool for accurately measuring Q, it is currently not clinically used for haemodynamic monitoring in emergency or intensive care settings. As of 2015, cardiac output measurement by MRI is routinely used in clinical cardiac MRI examinations.[66]

Dye dilution method

[edit]

The dye dilution method is done by rapidly injecting a dye, indocyanine green, into the right atrium of the heart. The dye flows with the blood into the aorta. A probe is inserted into the aorta to measure the concentration of the dye leaving the heart at equal time intervals [0, T] until the dye has cleared. Let c(t) be the concentration of the dye at time t. By dividing the time intervals from [0, T] into subintervals Δt, the amount of dye that flows past the measuring point during the subinterval from to is:

where is the rate of flow that is being calculated. The total amount of dye is:

and, letting , the amount of dye is:

Thus, the cardiac output is given by:

where the amount of dye injected is known, and the integral can be determined using the concentration readings.[67]

The dye dilution method is one of the most accurate methods of determining cardiac output during exercise. The error of a single calculation of cardiac output values at rest and during exercise is less than 5%. This method does not allow measurement of 'beat to beat' changes, and requires a cardiac output that is stable for approximately 10 s during exercise and 30 s at rest.[citation needed]

Factors influencing cardiac output

[edit]

Hierarchical summary of major factors influencing cardiac output.
Hierarchical summary of major factors influencing cardiac output.

Cardiac output is primarily controlled by the oxygen requirement of tissues in the body. In contrast to other pump systems, the heart is a demand pump that does not regulate its own output.[68] When the body has a high metabolic oxygen demand, the metabolically controlled flow through the tissues is increased, leading to a greater flow of blood back to the heart, leading to higher cardiac output.

The capacitance, also known as compliance, of the arterio-vascular channels that carry the blood also controls cardiac output. As the body's blood vessels actively expand and contract, the resistance to blood flow decreases and increases respectively. Thin-walled veins have about eighteen times the capacitance of thick-walled arteries because they are able to carry more blood by virtue of being more distensible.[69]

From this formula, it is clear the factors affecting stroke volume and heart rate also affect cardiac output. The figure to the right illustrates this dependency and lists a few of these factors. A more detailed hierarchical illustration is provided in a subsequent figure.

Equation (1) reveals HR and SV to be the primary determinants of cardiac output Q. A detailed representation of these factors is illustrated in the figure to the right. The primary factors that influence HR are autonomic innervation plus endocrine control. Environmental factors, such as electrolytes, metabolic products, and temperature are not shown. The determinants of SV during the cardiac cycle are the contractility of the heart muscle, the degree of preload of myocardial distention prior to shortening and the afterload during ejection.[70] Other factors such as electrolytes may be classified as either positive or negative inotropic agents.[71]

Cardiac response

[edit]
Table 3: Cardiac response to decreasing blood flow and pressure due to decreasing cardiac output[1]
Baroreceptors (aorta, carotid arteries, venae cavae, and atria) Chemoreceptors (both central nervous system and in proximity to baroreceptors)
Sensitive to Decreasing stretch[1] Decreasing O2 and increasing CO2, H+, and lactic acid[1]
Target Parasympathetic stimulation suppressed[1] Sympathetic stimulation increased[1]
Response of heart Increasing heart rate and increasing stroke volume[1] Increasing heart rate and increasing stroke volume[1]
Overall effect Increasing blood flow and pressure due to increasing cardiac output; haemostasis restored[1] Increasing blood flow and pressure due to increasing cardiac output; haemostasis restored[1]
Table 4: Cardiac response to increasing blood flow and pressure due to increasing cardiac output[1]
Baroreceptors (aorta, carotid arteries, venae cavae, and atria) Chemoreceptors (both central nervous system and in proximity to baroreceptors)
Sensitive to Increasing stretch[1] Increasing O2 and decreasing CO2, H+, and lactic acid[1]
Target Parasympathetic stimulation increased[1] Sympathetic stimulation suppressed[1]
Response of heart Decreasing heart rate and decreasing stroke volume[1] Decreasing heart rate and decreasing stroke volume[1]
Overall effect Decreasing blood flow and pressure due to decreasing cardiac output; haemostasis restored[1] Decreasing blood flow and pressure due to decreasing cardiac output; haemostasis restored[1]

Clinical significance

[edit]

When Q increases in a healthy but untrained individual, most of the increase can be attributed to an increase in heart rate (HR). Change of posture, increased sympathetic nervous system activity, and decreased parasympathetic nervous system activity can also increase cardiac output. HR can vary by a factor of approximately 3—between 60 and 180 beats per minute—while stroke volume (SV) can vary between 70 and 120 mL (2.5 and 4.2 imp fl oz; 2.4 and 4.1 US fl oz), a factor of only 1.7.[72][73][74]

Diseases of the cardiovascular system are often associated with changes in Q, particularly the pandemic diseases hypertension and heart failure. Increased Q can be associated with cardiovascular disease that can occur during infection and sepsis. Decreased Q can be associated with cardiomyopathy and heart failure.[70] Sometimes, in the presence of ventricular disease associated with dilatation, EDV may vary. An increase in EDV could counterbalance LV dilatation and impaired contraction. From equation (3), the resulting cardiac output Q may remain constant. The ability to accurately measure Q is important in clinical medicine because it provides for improved diagnosis of abnormalities and can be used to guide appropriate management.[75]

Example values

[edit]
Ventricular volumes
Measure Right ventricle Left ventricle
End-diastolic volume 144 mL (± 23 mL)[76] 142 mL (± 21 mL)[77]
End-diastolic volume / body surface area (mL/m2) 78 mL/m2 (± 11 mL/m2)[76] 78 mL/m2 (± 8.8 mL/m2)[77]
End-systolic volume 50 mL (± 14 mL)[76] 47 mL (± 10 mL)[77]
End-systolic volume / body surface area (mL/m2) 27 mL/m2 (± 7 mL/m2)[76] 26 mL/m2 (± 5.1 mL/m2)[77]
Stroke volume 94 mL (± 15 mL)[76] 95 mL (± 14 mL)[77]
Stroke volume / body surface area (mL/m2) 51 mL/m2 (± 7 mL/m2)[76] 52 mL/m2 (± 6.2 mL/m2)[77]
Ejection fraction 66% (± 6%)[76] 67% (± 4.6%)[77]
Heart rate 60–100 bpm[78] 60–100 bpm[78]
Cardiac output 4.0–8.0 L/minute[79] 4.0–8.0 L/minute[79]
[edit]

Ejection fraction

[edit]

Ejection fraction (EF) is a parameter related to SV. EF is the fraction of blood ejected by the left ventricle (LV) during the contraction or ejection phase of the cardiac cycle or systole. Prior to the start of systole, during the filling phase (diastole), the LV is filled with blood to the capacity known as end diastolic volume (EDV). During systole, the LV contracts and ejects blood until it reaches its minimum capacity known as end systolic volume (ESV). It does not completely empty. The following equations help translate the effect of EF and EDV on cardiac output Q, via SV.

Cardiac input

[edit]

Cardiac input (CI) is the inverse operation of cardiac output. As cardiac output implies the volumetric expression of ejection fraction, cardiac input implies the volumetric injection fraction (IF).

IF = end diastolic volume (EDV) / end systolic volume (ESV)

Cardiac input is a readily imaged mathematical model of diastole.[clarification needed]

Cardiac index

[edit]

In all resting mammals of normal mass, CO value is a linear function of body mass with a slope of 0.1 L/(min kg).[80][81] Fat has about 65% of oxygen demand per mass in comparison to other lean body tissues. As a result, the calculation of normal CO value in an obese subject is more complex; a single, common "normal" value of SV and CO for adults cannot exist. All blood flow parameters have to be indexed. It is accepted convention to index them by the body surface area, BSA [m2], by DuBois & DuBois Formula, a function of height and weight:

The resulting indexed parameters are stroke index (SI) and cardiac index (CI). Stroke index, measured in mL/beat/m2, is defined as

Cardiac index, measured in L/(min m2), is defined as

The CO equation (1) for indexed parameters then changes to the following.

The normal range for these indexed blood flow parameters are between 35 and 65 mL/beat/m2 for SI and between 2.5 and 4 L/(min m2) for CI.[82]

Combined cardiac output

[edit]

Combined cardiac output is the sum of the outputs of the right and left sides of the heart. It is a useful measurement in fetal circulation, where cardiac outputs from both sides of the heart work partly in parallel by the foramen ovale and ductus arteriosus, which directly supply the systemic circulation.[83]

Historical methods

[edit]

Fick principle

[edit]
An illustration of how spirometry is done
An illustration of how spirometry is done

The Fick principle, first described by Adolf Eugen Fick in 1870, assumes the rate of oxygen consumption is a function of the rate of blood flow and the rate of oxygen picked up by the red blood cells. Application of the Fick principle involves calculating the oxygen consumed over time by measuring the oxygen concentration of venous blood and arterial blood. Q is calculated from these measurements as follows:

  • VO2 consumption per minute using a spirometer (with the subject re-breathing air) and a CO2 absorber
  • the oxygen content of blood taken from the pulmonary artery (representing mixed venous blood)
  • the oxygen content of blood from a cannula in a peripheral artery (representing arterial blood)

From these values, we know that:

where

  • CA is the oxygen content of arterial blood, and,
  • CV is the oxygen content of venous blood.

This allows us to say

and therefore calculate Q. (CACV) is also known as the arteriovenous oxygen difference.[citation needed]

While considered to be the most accurate method of measuring Q, the Fick method is invasive and requires time for sample analysis, and accurate oxygen consumption samples are difficult to acquire. There have been modifications to the Fick method where respiratory oxygen content is measured as part of a closed system and the consumed oxygen is calculated using an assumed oxygen consumption index, which is then used to calculate Q. Other variations use inert gases as tracers and measure the change in inspired and expired gas concentrations to calculate Q (Innocor, Innovision A/S, Denmark).

The calculation of the arterial and venous oxygen content of the blood is a straightforward process. Almost all oxygen in the blood is bound to hæmoglobin molecules in the red blood cells. Measuring the content of hæmoglobin in the blood and the percentage of saturation of hæmoglobin—the oxygen saturation of the blood—is a simple process and is readily available to physicians. Each gram of haemoglobin can carry 1.34 mL of O2; the oxygen content of the blood—either arterial or venous—can be estimated using the following formula:

Pulmonary artery thermodilution (trans-right-heart thermodilution)

[edit]
Diagram of Pulmonary artery catheter (PAC)
Diagram of Pulmonary artery catheter (PAC)

The indicator method was further developed by replacing the indicator dye with heated or cooled fluid. Temperature changes rather than dye concentration are measured at sites in the circulation; this method is known as thermodilution. The pulmonary artery catheter (PAC) introduced to clinical practice in 1970, also known as the Swan-Ganz catheter, provides direct access to the right heart for thermodilution measurements. Continuous, invasive, cardiac monitoring in intensive care units has been mostly phased out. The PAC remains useful in right-heart study done in cardiac catheterisation laboratories.[citation needed]

The PAC is balloon tipped and is inflated, which helps "sail" the catheter balloon through the right ventricle to occlude a small branch of the pulmonary artery system. The balloon is then deflated. The PAC thermodilution method involves the injection of a small amount (10 mL) of cold glucose at a known temperature into the pulmonary artery and measuring the temperature a known distance away 6–10 cm (2.4–3.9 in) using the same catheter with temperature sensors set apart at a known distance.[citation needed]

The historically significant Swan-Ganz multi-lumen catheter allows reproducible calculation of cardiac output from a measured time-temperature curve, also known as the thermodilution curve. Thermistor technology enabled the observations that low CO registers temperature change slowly and high CO registers temperature change rapidly. The degree of temperature change is directly proportional to the cardiac output. In this unique method, three or four repeated measurements or passes are usually averaged to improve accuracy.[84][85] Modern catheters are fitted with heating filaments that intermittently heat up and measure the thermodilution curve, providing serial Q measurements. These instruments average measurements over 2–9 minutes depending on the stability of the circulation, and thus do not provide continuous monitoring.

PAC use can be complicated by arrhythmias, infection, pulmonary artery rupture and damage to the right heart valve. Recent studies in patients with critical illnesses, sepsis, acute respiratory failure and heart failure suggest that use of the PAC does not improve patient outcomes.[6][7][8] This clinical ineffectiveness may relate to its poor accuracy and sensitivity, which have been demonstrated by comparison with flow probes across a sixfold range of Q values.[14] Use of PAC is in decline as clinicians move to less invasive and more accurate technologies for monitoring hæmodynamics.[86]

See also

[edit]

References

[edit]
[edit]
Revisions and contributorsEdit on WikipediaRead on Wikipedia
from Grokipedia
Cardiac output (CO) is the volume of blood pumped by the heart into the systemic circulation per minute, serving as a fundamental measure of cardiovascular performance. It is calculated as the product of (HR), the number of heartbeats per minute, and (SV), the amount of blood ejected by the left ventricle per beat, with the formula CO = HR × SV. In healthy adults at rest, cardiac output typically ranges from 5 to 6 liters per minute, though it can increase dramatically to over 35 liters per minute during intense exercise in elite athletes. Cardiac output is regulated by several physiological factors that influence either or . is primarily controlled by the and modulated by inputs, with normal resting rates between 60 and 100 beats per minute. , in turn, depends on preload (the stretching the ventricle, governed by the Frank-Starling law), (the resistance against which the heart pumps, often related to systemic ), and (the intrinsic force of ventricular contraction). These determinants interact dynamically; for instance, increased preload enhances up to a point, while elevated can reduce it, particularly in compromised hearts. Clinically, cardiac output is crucial for maintaining adequate to vital organs and is often assessed in conditions like , shock, or during surgical monitoring. Low cardiac output can lead to symptoms such as , , and , and is associated with increased morbidity and mortality in cardiovascular diseases. It is measured noninvasively via or cardiac MRI, or invasively using techniques like thermodilution or the , which calculates CO based on oxygen consumption and .

Definition and Physiology

Definition

Cardiac output (CO) is defined as the volume of blood pumped by the heart per minute, serving as the primary mechanism for circulating throughout the body to meet tissue demands. This measure typically refers to the output from the left ventricle, which propels oxygenated into the systemic circulation. The fundamental equation for cardiac output is CO = (HR) × (SV), where HR represents the number of heartbeats per minute and SV is the volume of blood ejected by the ventricle per beat. The resulting CO is expressed in liters per minute (L/min), with normal resting values around 5-6 L/min in adults. Cardiac output plays a crucial role in maintaining systemic , ensuring adequate delivery of oxygen and nutrients to tissues while removing products. In normal physiology, the cardiac outputs of the right and left ventricles are equal, as the closed requires balanced pulmonary and systemic blood flows in the absence of shunts.

Determinants of Cardiac Output

Cardiac output is fundamentally determined by the product of and , where adjustments in either component allow the heart to meet varying physiological demands. represents the volume of blood ejected by the ventricle per beat, calculated as the difference between —the amount of blood in the ventricle at the end of —and end-systolic volume—the residual blood remaining after . This difference typically ranges from 70 to 80 mL in a resting adult, providing the baseline for effective circulation. A primary physiological mechanism governing is the Frank-Starling law, which posits that increased myocardial fiber length due to greater enhances the force of contraction, thereby augmenting up to an optimal stretch point. This intrinsic property ensures that the heart adapts output to incoming venous return, preventing blood pooling in the venous system. Within physiological limits, this stretch-induced potentiation optimizes overlap and actin-myosin interactions, directly linking preload to ejection efficiency. Heart rate, the number of cardiac cycles per minute, is regulated primarily by the , with sympathetic activation accelerating rate through β-adrenergic stimulation of the , while parasympathetic input via the slows it. in the and play a crucial role in this regulation by sensing arterial changes and modulating autonomic outflow: elevated pressure triggers increased parasympathetic activity to decrease heart rate, whereas enhances sympathetic drive to elevate it, thereby stabilizing cardiac output. This operates via the nucleus tractus solitarius in the , providing rapid beat-to-beat adjustments. The interplay between and maintains cardiac output , as an increase in one can compensate for a decrease in the other; for instance, often offsets reduced in conditions like to preserve overall . However, extreme elevations in may shorten diastolic filling time, potentially limiting stroke volume gains and underscoring their interdependent nature. Such compensatory dynamics ensure that cardiac output rises appropriately during exercise or stress, typically from 5 L/min at rest to over 20 L/min.

Factors Influencing Cardiac Output

Preload and Afterload

Preload refers to the initial stretching of the cardiac myocytes prior to contraction, quantified as the ventricular end-diastolic pressure or volume at the end of diastole. It is primarily influenced by venous return, which delivers blood to the heart, and total blood volume, as increases in either expand the end-diastolic volume. Higher preload enhances the overlap of actin and myosin filaments in the sarcomeres, optimizing force generation during systole. Afterload represents the resistance the ventricle must overcome to eject blood, primarily determined by systemic vascular resistance (SVR), which arises from the tone and caliber of peripheral arterioles. It is often approximated by (MAP), the average pressure in the arteries during a , as this reflects the load against which the heart pumps. SVR quantifies this opposition to flow and can be calculated using the formula: SVR=MAPCVPCO×80\text{SVR} = \frac{\text{MAP} - \text{CVP}}{\text{CO}} \times 80 where CVP is central venous pressure, CO is cardiac output, and the result is expressed in dynes·s·cm⁻⁵; this derivation stems from Ohm's law applied to hemodynamics, converting pressure differences to resistance. These factors modulate stroke volume (SV), a key component of cardiac output (CO = SV × heart rate). Increased preload augments SV through the Frank-Starling mechanism, whereby greater end-diastolic volume stretches myocardial fibers, leading to stronger contractions and higher ejected blood volume. Conversely, elevated afterload impedes ventricular ejection, increasing end-systolic volume and thereby reducing SV, as the heart expends more energy against higher resistance without proportionally increasing output.

Heart Rate and Contractility

Myocardial contractility refers to the intrinsic ability of the to generate force during contraction, independent of preload and , through chemo-mechanical processes that are kinetically controlled. This property is enhanced by stimulation via β1-adrenergic receptors, which increase intracellular calcium availability and thereby boost the force of myocardial contraction. Similarly, positive inotropic agents, such as catecholamines or inhibitors, augment contractility by similar mechanisms, leading to improved ejection of blood and higher cardiac output without altering loading conditions. Heart rate (HR), typically ranging from 60 to 100 beats per minute at rest in healthy adults, is a key determinant of cardiac output (CO), calculated as the product of HR and (SV). Within physiological limits, HR and SV exhibit an inverse relationship to maintain stable CO; for instance, a moderate increase in HR is often compensated by a slight decrease in SV due to reduced filling time per beat, preserving overall output. Additionally, elevated HR exerts a positive inotropic effect on the myocardium through the , also known as the treppe or staircase phenomenon, where successive contractions at higher frequencies build increasing force due to enhanced calcium handling in cardiac cells. However, excessive HR can limit CO by disproportionately shortening the diastolic phase, which reduces ventricular filling time and thus impairs preload and SV. This effect becomes particularly pronounced during intense exercise or in pathological states like , where the net result may be diminished CO despite the initial compensatory rise in HR.

Measurement Techniques

Non-Invasive Methods

Non-invasive methods for measuring cardiac output (CO) provide accessible alternatives to invasive techniques, relying on external sensors or imaging to estimate (SV) and without penetrating the body, thereby minimizing risks such as infection or vascular complications. These approaches are particularly valuable in clinical settings like intensive care units or outpatient evaluations, where continuous monitoring is needed but is paramount. Common techniques include Doppler ultrasound, impedance-based methods, and pulse waveform analysis, each leveraging physiological signals to derive CO as the product of SV and heart rate. Doppler ultrasound employs the Doppler effect to assess blood flow velocity, enabling SV estimation through the integration of velocity-time integrals (VTI) across a cross-sectional area (CSA) of the outflow tract, as expressed by the formula SV = CSA × VTI. In transthoracic echocardiography (TTE), a transcutaneous probe is placed on the chest to image the left ventricular outflow tract (LVOT), measuring aortic velocity and LVOT diameter to calculate CO; this method is widely used due to its portability and real-time capabilities. Transesophageal echocardiography (TEE) offers higher resolution by inserting a probe into the esophagus for closer proximity to the heart, improving accuracy in patients with poor acoustic windows, though it requires sedation. These variants provide beat-to-beat CO assessments but depend on operator skill for precise alignment and measurement. Impedance cardiography (ICG) measures changes in thoracic during the , attributing variations to shifts in the and to derive SV. Electrodes are placed on the and to apply a high-frequency current and detect impedance fluctuations, with SV approximated from the first of impedance (dZ/dt) and ejection time, incorporating factors like impedance (Z0) and velocity of blood flow. This technique allows continuous, bedside monitoring without , making it suitable for in hemodynamically unstable patients. Electrical cardiometry represents an advancement over traditional ICG by incorporating the electrical conductivity of , which varies with its orientation in the during , to more accurately estimate SV. Using four dual electrodes on the , it analyzes phase shifts in the electrical field caused by pulsatile flow, applying algorithms to compute CO without assuming constant resistivity. This method has shown good correlation with invasive references in various populations, including and critically ill adults, enhancing reliability in dynamic conditions. Pulse pressure methods, such as those using the Finapres device, analyze continuous arterial waveforms obtained via finger cuff photoplethysmography to estimate CO through pulse contour analysis. The system employs the volume clamp technique to maintain constant arterial volume, deriving SV from the waveform's , aortic compliance, and impedance, often calibrated initially for accuracy. Devices like Finapres enable non-invasive, real-time tracking of hemodynamic changes during procedures or stress tests, though they require validation against reference standards for absolute values. These non-invasive techniques offer key advantages, including ease of bedside application, absence of , and suitability for serial measurements in low-risk , facilitating early detection of hemodynamic instability without procedural hazards. However, limitations include operator dependency in ultrasound-based methods, potential inaccuracies from movement or arrhythmias in impedance techniques, and the need for in waveform analyses, which can affect precision compared to invasive gold standards like . (MRI) serves as a non-invasive reference for CO validation, providing precise volumetric assessments without .

Invasive Methods

Invasive methods for measuring cardiac output involve direct vascular access, providing high-fidelity data essential for managing hemodynamically unstable patients in critical care environments, such as intensive care units (ICUs). These techniques, including thermodilution and indicator dilution, are considered reference standards due to their accuracy in clinical settings with cardiac pathology. Pulmonary artery thermodilution, performed via a Swan-Ganz catheter inserted through a central into the , remains the gold standard for invasive cardiac output monitoring. The procedure entails injecting a known volume of cold saline (typically 5-10 mL at 0-10°C) into the right atrium or proximal , where a at the tip detects the resulting temperature change in the as it flows past. The cardiac output is calculated from the area under the temperature-time curve using the Stewart-Hamilton equation: CO=V×(TBTI)×KΔTdtCO = \frac{V \times (T_B - T_I) \times K}{\int \Delta T \, dt} where VV is the injectate volume, TBT_B and TIT_I are the blood and injectate temperatures, respectively, KK is a correction factor accounting for specific heats and densities, and ΔTdt\int \Delta T \, dt represents the integral of the temperature change over time. Measurements are typically repeated three to five times for averaging to minimize variability, with errors reduced to under 5% in stable conditions. Introduced in 1971, this method excels in ICU settings for real-time assessment of cardiac function during shock or surgery. The dye dilution method, another established invasive approach, injects a known quantity of indicator dye, such as , into the central circulation, followed by serial arterial blood sampling to plot a concentration-time curve. The dye is rapidly mixed in the bloodstream, and cardiac output is derived from the Stewart-Hamilton principle by dividing the injected amount by the curve's area, corrected for recirculation. is preferred for its non-toxicity, rapid hepatic clearance, and detectability via , making it suitable for patients without severe liver dysfunction. This technique, historically significant since the , offers precision comparable to thermodilution but requires blood withdrawal, limiting its use to intermittent measurements. Ultrasound dilution represents a variant of dilution techniques, utilizing saline boluses injected via a central venous line and detected by sensors on arterial and venous lines, often in extracorporeal circuits like . The method measures changes in ultrasound velocity caused by the saline's acoustic properties, enabling transpulmonary cardiac output estimation without dyes or thermistors. Validated in animal models and pediatric patients, it provides reliable readings in the presence of shunts and is particularly useful in ICU or perioperative settings with vascular access. Despite their accuracy, invasive methods carry risks including catheter-related infections, arrhythmias during insertion, , and rare pulmonary artery rupture, necessitating strict aseptic technique and monitoring in high-acuity care. They are primarily employed in ICUs for guiding therapy in conditions like or , where non-invasive Doppler may serve only for initial screening.

Cardiac Index and Ejection Fraction

The cardiac index (CI) is a hemodynamic parameter that normalizes cardiac output (CO) to an individual's body surface area (BSA), providing a size-adjusted measure of cardiac performance essential for comparing patients across varying body sizes. It is calculated using the formula: CI=COBSA\text{CI} = \frac{\text{CO}}{\text{BSA}} where CO is expressed in liters per minute and BSA in square meters, yielding units of L/min/m². The normal range for CI in healthy adults at rest is 2.5 to 4.0 L/min/m², with values below 2.2 L/min/m² often indicating inadequate cardiac function in clinical contexts. BSA is commonly estimated using the Du Bois formula, derived from empirical measurements of : BSA=0.007184×Weight0.425×Height0.725\text{BSA} = 0.007184 \times \text{Weight}^{0.425} \times \text{Height}^{0.725} where weight is in kilograms and height in centimeters, resulting in BSA in square meters; this formula remains a standard in clinical practice for dosing medications and normalizing physiological parameters. The (EF) quantifies the efficiency of the left ventricle's systolic function by measuring the fraction of ejected with each contraction. It is computed as: EF=(SVEDV)×100%\text{EF} = \left( \frac{\text{SV}}{\text{EDV}} \right) \times 100\% where SV is stroke volume and EDV is end-diastolic volume, typically reported as a percentage. Normal EF values, assessed via 2D echocardiography, range from 52% to 72% in men and 54% to 74% in women, reflecting robust ventricular contractility. Clinically, EF is most commonly derived from echocardiography using the modified Simpson's biplane method, which involves tracing ventricular volumes in multiple views to estimate SV and EDV; this metric links directly to cardiac output through SV, as CO incorporates SV as a core component. Stroke volume (SV) is defined as the volume of blood ejected from the left ventricle of the heart during each systolic contraction. In a typical adult male weighing 70 kg, the average SV is approximately 70 mL. This metric serves as a fundamental component in assessing ventricular performance and contributes directly to cardiac output through its multiplication by . Stroke volume of the right ventricle, which propels blood into the . Under normal physiological conditions, without intracardiac shunting, right ventricular cardiac output equals left ventricular cardiac output, ensuring balanced circulation between the pulmonary and systemic systems. However, in the presence of intracardiac shunts or , this balance is disrupted, leading to unequal stroke volumes between the ventricles. In scenarios involving parallel circulations, such as the fetal cardiovascular system, the concept of combined cardiac output emerges, representing the summed outputs from both ventricles to support dual systemic and placental flows. Similarly, during (ECMO) support, particularly in venoarterial configurations, the total effective circulation may involve a combined output from the native heart and the extracorporeal pump, functioning in parallel to maintain systemic perfusion. Stroke volume can be derived indirectly by dividing cardiac output by , providing a calculated estimate in clinical assessments. Alternatively, SV can be measured directly using imaging techniques that quantify ventricular volumes during the .

Clinical Significance

Normal Values and Variations

In healthy adults at rest, cardiac output typically ranges from 4 to 8 liters per minute (L/min), with an average value of approximately 5 L/min. This value varies with factors such as age, , and body size; for instance, larger individuals tend to have higher absolute cardiac output due to greater metabolic demands, while provides a normalized measure accounting for . differences are minimal after , as women generally exhibit smaller stroke volumes compensated by higher s, resulting in comparable overall cardiac output to men. Athletes at rest often display cardiac output values within or slightly above the typical range, reflecting adaptations like increased despite lower heart rates, though absolute differences are modest compared to non-athletes of similar body size. In children, cardiac output norms are best expressed as , ranging from 3.5 to 5.5 L/min per square meter (L/min/m²) depending on age, with higher values in infants and neonates that gradually decline toward levels. Physiological variations in cardiac output occur in response to normal demands, such as increasing substantially during exercise—from about 5 L/min at rest to 20–25 L/min in untrained or moderately trained individuals and over 35 L/min in elite athletes—to meet elevated oxygen needs—without exceeding limits in healthy states. During pregnancy, cardiac output rises by 30-50% above non-pregnant baseline levels by the second trimester to support maternal and fetal circulation. In contrast, cardiac output decreases with advancing age due to reduced myocardial contractility and vascular stiffness, often falling below 4 L/min in older adults at rest. Similarly, in hypothermia, cardiac output declines primarily from bradycardia and impaired contractility, though initial compensatory increases may occur in mild cases. These changes highlight cardiac output's adaptability in resting versus stress states, with normalization via cardiac index aiding comparisons across populations.

Pathophysiological Implications

Low cardiac output (CO) is a hallmark of several critical conditions, including and various forms of shock, where it results in inadequate tissue and end-organ dysfunction. In , reduced CO stems from impaired or structural abnormalities, leading to systemic hypoperfusion, activation of compensatory mechanisms like the renin-angiotensin-aldosterone system, and progressive organ damage such as renal insufficiency and hepatic congestion. In , a primary cardiac disorder, low CO directly causes circulatory failure and multi-organ hypoperfusion, often exacerbated by or valvular dysfunction. Similarly, , triggered by significant fluid or blood loss, diminishes preload and thereby reduces CO, culminating in tissue hypoxia and if untreated. Conversely, elevated CO characterizes hyperdynamic states, where increased metabolic demands or drive compensatory cardiac hyperactivity, potentially overwhelming the heart and leading to failure. In , systemic inflammation induces and myocardial depression alongside an initial high-output phase, resulting in maldistribution of blood flow and tissue hypoperfusion despite elevated CO. Chronic elevates CO through reduced oxygen-carrying capacity, prompting and increased , which can precipitate over time. Thyrotoxicosis similarly boosts CO via thyroid hormone-mediated enhancements in and contractility, risking arrhythmias or in severe cases. Monitoring CO is crucial in these pathologies, with a cardiac index below 2.2 L/min/m² signaling severe compromise and high mortality risk, particularly in , guiding urgent interventions. Therapeutically, inotropes such as are employed to augment contractility and elevate CO in low-output states like or shock, improving without excessive . For conditions involving high , such as acute with elevated systemic vascular resistance, vasodilators like nitroprusside reduce impedance to ejection, thereby enhancing CO and alleviating pulmonary congestion.

Historical Development

Early Principles

The foundational understanding of cardiac output emerged from early anatomical and physiological insights into blood circulation. In 1628, published Exercitatio Anatomica de Motu Cordis et Sanguinis in Animalibus, demonstrating through quantitative experiments that blood circulates continuously in a driven by the heart's pumping action, challenging ancient theories of blood generation and consumption. This concept established the heart as the central organ propelling blood flow, laying the groundwork for later quantification of cardiac output as the total volume of blood ejected by the heart per unit time. Building on circulatory principles, Adolf Fick proposed in 1870 a method for indirect measurement of cardiac output using oxygen consumption as a marker substance. The Fick principle states that cardiac output (CO) is equal to the rate of oxygen consumption divided by the arteriovenous oxygen content difference: CO=V˙O2CaO2CvO2CO = \frac{\dot{V}O_2}{C_aO_2 - C_vO_2} where V˙O2\dot{V}O_2 is oxygen uptake, CaO2C_aO_2 is arterial oxygen content, and CvO2C_vO_2 is mixed venous oxygen content. This approach enabled estimation of blood flow through gas exchange analysis without direct volumetric measurement, marking a seminal advance in cardiovascular physiology. The first measurement in humans was performed in 1930 by Baumann and Grollman using right heart catheterization. In the early 20th century, physiologists including Ernest Starling formalized the relationship defining cardiac output as the product of heart rate (HR) and stroke volume (SV), where CO=HR×SVCO = HR \times SV. Starling's 1914 experiments on isolated heart-lung preparations demonstrated how SV varies with preload, influencing overall output while HR provides the temporal component, integrating mechanical and neural regulatory aspects of cardiac function. The , however, relies on key assumptions including steady-state conditions for oxygen consumption and uniform mixing of blood, which limit its applicability during transient physiological states or when ventilation-perfusion mismatches occur. These constraints highlight the method's dependence on equilibrium for accurate indirect assessment.

Evolution of Measurement Techniques

The measurement of cardiac output (CO) has evolved significantly since the late , transitioning from invasive, labor-intensive techniques to more accessible and less risky methods. A key milestone was the proposal of the in 1870 by Adolf Fick, which provided the theoretical foundation for quantifying CO through oxygen consumption and arterio-venous oxygen differences, though initial validation in human subjects occurred in 1930 through experiments. This laid the groundwork for subsequent indicator-based approaches. In the late 1890s, George Neil Stewart pioneered the dye dilution method, injecting indicators like saline into the bloodstream and measuring their concentration downstream to estimate blood flow, marking the first practical application of indicator dilution for CO in experimental settings. This technique gained traction in the 1920s with refinements, such as continuous infusion methods using , enabling more reliable measurements in animal and human studies. By the , the method evolved toward thermodilution, where temperature changes from injected cold solutions served as the indicator, offering improved accuracy over dyes by reducing recirculation errors; this was first demonstrated effectively in 1954 by G. Fegler using hepatic vein injections in animals. The 1970s brought a major clinical advancement with the introduction of the flow-directed pulmonary artery catheter, commonly known as the Swan-Ganz catheter, developed by Harold Swan and William Ganz in 1970. This device facilitated bedside thermodilution CO measurements by allowing rapid injection into the right heart and detection in the pulmonary artery, revolutionizing hemodynamic monitoring in critical care and making CO assessment routine during surgeries and in intensive care units. The marked a shift toward non-invasive techniques, with the integration of Doppler ultrasound into enabling estimation of CO through calculations from blood flow velocities across cardiac valves. This approach, building on earlier Doppler developments from the , provided real-time, radiation-free assessments, particularly via transthoracic and transesophageal probes, and became widely adopted for outpatient and intraoperative use. In the 1990s, (MRI) emerged as a gold-standard non-invasive method for volumetric CO measurement, leveraging phase-contrast techniques to quantify blood flow through great vessels with high precision and without . Concurrently, impedance cardiography, originally developed in the 1960s, advanced for settings by the late 1990s and early 2000s, using thoracic changes to estimate beat-to-beat CO in mobile patients, facilitating long-term monitoring outside clinical environments. These innovations continue to expand CO assessment's accessibility while minimizing patient risk.

References

Add your contribution
Related Hubs
Contribute something
User Avatar
No comments yet.